      The reproducibility of urinalysis using multiple reagent test strips.

      European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies
      Esterases, analysis, Glycosuria, urine, Humans, Hydrogen-Ion Concentration, Leukocytes, enzymology, Observer Variation, Proteinuria, Reagent Strips, Reproducibility of Results, Spectrophotometry, Urinalysis, methods

      Read this article at

      ScienceOpenPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Ninety urine samples were examined twice by 3 "observers" (two persons, using only visual observation, and one person using a spectrophotometric analyser) using multiple reagent teststrips. To determine reproducibility, inter- and intra-observer agreement were calculated and expressed as Cohen's kappa and as weighted kappa. The results show negligible intra-observer differences between the visual and spectrophotometric observation. The lack of agreement between inter- and intra-observer urinalysis results, using multiple reagent test strips was disappointing, considering the simplicity of the test procedure. Further improvement of reproducibility, e.g. by enhancing the discoloration of the test pads, is necessary. Reproducibility is not improved by using a spectrophotometric analyser instead of visual reading of the test strips.