      Leaflet Mechanical Stress in Different Designs and Generations of Transcatheter Aortic Valves: An in Vitro Study

      research-article
      , Dipl-Ing, MSc, PhD a , b , , , Dipl-Ing, PhD a , , PhD a , , MD, PhD b , , MD b , , PhD b , , Dipl-Ing, PhD c , , DVM, PhD b
      Structural Heart
      Elsevier
      Durability, In silico, In vitro, Mechanical stress, Transcatheter heart valves

      Read this article at

      ScienceOpenPublisherPMC
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          It is unknown whether bioprostheses used for transcatheter aortic valve implantation will have similar long-term durability as those used for surgical aortic valve replacement. Repetitive mechanical stress applied to the valve leaflets, particularly during diastole, is the main determinant of structural valve deterioration. Leaflet mechanical stress cannot be measured in vivo. The objective of this in vitro/in silico study was thus to compare the magnitude and regional distribution of leaflet mechanical stress in old vs new generations of self-expanding (SE) vs balloon expandable (BE) transcatheter heart valves (THVs).

          Methods

          A double activation simulator was used for in vitro testing of two generations of SE THV (Medtronic CoreValve 26 mm and EVOLUT PRO 26 mm) and two generations of BE THV (Edwards SAPIEN 23 mm vs SAPIEN-3 23 mm). These THVs were implanted within a 21-mm aortic annulus. A noncontact system based on stereophotogammetry and digital image correlation with high spatial and temporal resolution (2000 img/sec) was used to visualize the valve leaflet motion and perform the three-dimensional analysis. A finite element model of the valve was developed, and the leaflet deformation obtained from the digital image correlation analysis was applied to the finite element model to calculate local leaflet mechanical stress during diastole.

          Results

          The maximum von Mises leaflet stress was higher in early vs new THV generation ( p < 0.05) and in BE vs SE THV ( p < 0.05): early generation BE: 2.48 vs SE: 1.40 MPa; new generation BE: 1.68 vs SE: 1.07 MPa. For both types of THV, the highest values of leaflet stress were primarily observed in the upper leaflet edge near the commissures and to a lesser extent in the mid-portion of the leaflet body, which is the area where structural leaflet deterioration most often occurs in vivo.

          Conclusions

          The results of this in vitro/in silico study suggest that: i) Newer generations of THVs have ∼30% lower leaflet mechanical stress than the early generations; ii) For a given generation, SE THVs have lower leaflet mechanical stress than BE THVs. Further studies are needed to determine if these differences between new vs early THV generations and between SE vs BE THVs will translate into significant differences in long-term valve durability in vivo.

          Related collections

          Most cited references40

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Global Burden of Cardiovascular Diseases and Risk Factors, 1990–2019

          Cardiovascular diseases (CVDs), principally ischemic heart disease (IHD) and stroke, are the leading cause of global mortality and a major contributor to disability. This paper reviews the magnitude of total CVD burden, including 13 underlying causes of cardiovascular death and 9 related risk factors, using estimates from the Global Burden of Disease (GBD) Study 2019. GBD, an ongoing multinational collaboration to provide comparable and consistent estimates of population health over time, used all available population-level data sources on incidence, prevalence, case fatality, mortality, and health risks to produce estimates for 204 countries and territories from 1990 to 2019. Prevalent cases of total CVD nearly doubled from 271 million (95% uncertainty interval [UI]: 257 to 285 million) in 1990 to 523 million (95% UI: 497 to 550 million) in 2019, and the number of CVD deaths steadily increased from 12.1 million (95% UI:11.4 to 12.6 million) in 1990, reaching 18.6 million (95% UI: 17.1 to 19.7 million) in 2019. The global trends for disability-adjusted life years (DALYs) and years of life lost also increased significantly, and years lived with disability doubled from 17.7 million (95% UI: 12.9 to 22.5 million) to 34.4 million (95% UI:24.9 to 43.6 million) over that period. The total number of DALYs due to IHD has risen steadily since 1990, reaching 182 million (95% UI: 170 to 194 million) DALYs, 9.14 million (95% UI: 8.40 to 9.74 million) deaths in the year 2019, and 197 million (95% UI: 178 to 220 million) prevalent cases of IHD in 2019. The total number of DALYs due to stroke has risen steadily since 1990, reaching 143 million (95% UI: 133 to 153 million) DALYs, 6.55 million (95% UI: 6.00 to 7.02 million) deaths in the year 2019, and 101 million (95% UI: 93.2 to 111 million) prevalent cases of stroke in 2019. Cardiovascular diseases remain the leading cause of disease burden in the world. CVD burden continues its decades-long rise for almost all countries outside high-income countries, and alarmingly, the age-standardized rate of CVD has begun to rise in some locations where it was previously declining in high-income countries. There is an urgent need to focus on implementing existing cost-effective policies and interventions if the world is to meet the targets for Sustainable Development Goal 3 and achieve a 30% reduction in premature mortality due to noncommunicable diseases.
            • Record: found
            • Abstract: not found
            • Article: not found

            2017 ESC/EACTS Guidelines for the management of valvular heart disease.

              • Record: found
              • Abstract: found
              • Article: not found

              Standardized Definition of Structural Valve Degeneration for Surgical and Transcatheter Bioprosthetic Aortic Valves

              Bioprostheses are prone to structural valve degeneration, resulting in limited long-term durability. A significant challenge when comparing the durability of different types of bioprostheses is the lack of a standardized terminology for the definition of a degenerated valve. This issue becomes especially important when we try to compare the degeneration rate of surgically inserted and transcatheter bioprosthetic valves. This document, by the VIVID (Valve-in-Valve International Data), proposes practical and standardized definitions of valve degeneration and provides recommendations for the timing of clinical and imaging follow-up assessments accordingly. Its goal is to improve the quality of research and clinical care for patients with deteriorated bioprostheses by providing objective and strict criteria that can be utilized in future clinical trials. We hope that the adoption of these criteria by both the cardiological and surgical communities will lead to improved comparability and interpretation of durability analyses.

                Author and article information

                Contributors
                Journal
                Struct Heart
                Struct Heart
                Structural Heart
                Elsevier
                2474-8706
                2474-8714
                19 December 2023
                March 2024
                19 December 2023
                : 8
                : 2
                : 100262
                Affiliations
                [a ]Aix-Marseille University/Gustave Eiffel University, LBA-UMRT24, Marseille, France
                [b ]Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval, Québec, Canada
                [c ]Laboratory for Biomaterials and Bioengineering, Laval University, Quebec, Canada
                Author notes
                []Address correspondence to: Viktória Stanová, Dipl-Ing, MSc, PhD, Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval, Québec, Canada. stano.viktoria@ 123456gmail.com
                Article
                S2474-8706(23)00198-7 100262
                10.1016/j.shj.2023.100262
                10927450
                d6353dae-88aa-49fc-afee-9b0abdcd5c77
                © 2023 The Authors

                This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

                History
                : 28 December 2022
                : 14 September 2023
                : 14 October 2023
                Categories
                Original Research

                durability,in silico,in vitro,mechanical stress,transcatheter heart valves

                Comments

                Comment on this article

                Related Documents Log