      Stent aórtico pulsátil: uma nova alternativa na assistência ventricular crônica Translated title: Pulsatile stent graft: a new alternative in chronic ventricular assistance

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          OBJETIVO: A insuficiência cardíaca é uma das causas mais comuns de internação. Dispositivos para assistência circulatória crônica foram testados e, em sua maioria, são de alta complexidade. O objetivo deste estudo é a descrição de uma endoprótese contrátil com capacidade de pulsação crônica no interior da aorta descendente, de maneira semelhante à produzida pelo balão intra-aórtico. MÉTODOS: Endopróteses pulsáteis compostas de níquel-titânio foram posicionadas de forma a envolver tubos de látex, simulando a aorta. Diferentes correntes elétricas foram aplicadas a unidades ligadas em série, de modo a causar contração da estrutura e deslocamento de uma coluna líquida. Foram realizadas duas sequências de testes: a primeira com duas gaiolas metálicas e a segunda com cinco gaiolas. Na primeira sequência de testes, aplicou-se tensão de 16,3 volts e corrente de 5 amperes e, na segunda sequência, tensão de 15 volts e corrente de 7 amperes. RESULTADOS: Na primeira sequência de testes, obteve-se o efeito pulsátil dos 2 stents, havendo contração do tubo e deslocamento da coluna d'água suficientes para validar o efeito pulsátil da endoprótese. As duas estruturas ejetaram um volume de 2,6 mL por ciclo, com uma variação de 29 mm na altura da coluna de água, equivalente a 8% de contração durante a pulsação. Na segunda sequência, conseguiu-se uma variação de 7,4 mL por ciclo. CONCLUSÃO: Os resultados obtidos comprovam a contratilidade da endoprótese pulsátil ativada pela aplicação de corrente elétrica. Continuidade do estudo e aperfeiçoamento do material se fazem necessários para obtenção de modelo mais eficiente do ponto de vista energético e com maior pulsação, para permitir volumes de ejeção comparáveis aos de balões intra-aórticos.

          Translated abstract

          OBJECTIVE: Heart failure is currently one of the most common hospitalization causes. Several chronic circulatory assist devices have been tested and are highly complex. The objective is the description of a pulsatile endoprosthesis capable of applying a chronic pulse within the descending aorta, similar to that produced by intra-aortic balloon. METHODS: Pulsatile stents composed of nickel-titanium were built and positioned to engage latex tubes simulating the aorta. Different electric currents were applied to units connected in series in order to cause structure contraction and displacement of a liquid column. There were two sequence tests: first composed of two metallic cages and the second composed of five cages. At first sequence tests was applied a voltage of 16.3 volts and a current of 5 amperes. In the second, voltage of 15 volts and current of 07 amperes. RESULTS: In the first sequence was obtained the pulsatile effect of stent, with contraction of the tube and displacement of the water column sufficient to validate the pulsating effect of the endoprosthesis. The two structures ejected a volume of 2.6 ml per cycle, with a range of 29 mm in height of the column of water equivalent to 8% shrinkage during the pulse. In the second sequence, it reachead a variation of 7.4 mL per cycle. CONCLUSION: The results obtained confirm the stent pulsatile contractility activated by electrical current. The continuity of the study and material improvement are necessary to obtain more efficient model from the point of view of energy and pulse, to allow ejection volumes comparable with the intra-aortic balloons

          Related collections

          Most cited references16

          • Record: found
          • Abstract: not found
          • Article: not found

          Polymers in Biomedicine and Electronics

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Biocompatibility of Nitinol Stapes Prosthesis

              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              Smart and hybrid materials: perspectives for their use in textile structures for better health care

                Bookmark

                Author and article information

                Contributors
                Role: ND
                Role: ND
                Role: ND
                Role: ND
                Role: ND
                Role: ND
                Role: ND
                Journal
                rbccv
                Brazilian Journal of Cardiovascular Surgery
                Braz. J. Cardiovasc. Surg.
                Sociedade Brasileira de Cirurgia Cardiovascular (São José do Rio Preto )
                1678-9741
                June 2013
                : 28
                : 2
                : 217-223
                Affiliations
                [1 ] Universidade Federal de São Paulo Brazil
                [2 ] Universidade Federal de São Paulo Brazil
                [3 ] Universidade Federal de São Paulo Brazil
                [4 ] Braile Biomédica Brazil
                [5 ] Faculdade de Medicina de São José do Rio Preto Brazil
                [6 ] Universidade Estadual de Campinas Brazil
                Article
                S0102-76382013000200010
                10.5935/1678-9741.20130031
                d90b5bc3-8803-4368-9582-c6b201e99769

                http://creativecommons.org/licenses/by/4.0/

                History
                Product

                SciELO Brazil

                Self URI (journal page): http://www.scielo.br/scielo.php?script=sci_serial&pid=0102-7638&lng=en
                Categories
                CARDIAC & CARDIOVASCULAR SYSTEMS
                SURGERY

                Surgery,Cardiovascular Medicine
                Heart failure,Aorta,Stents,Insuficiência cardíaca
                Surgery, Cardiovascular Medicine
                Heart failure, Aorta, Stents, Insuficiência cardíaca

                Comments

                Comment on this article