      Multiacquisition Variable-Resonance Image Combination Magnetic Resonance Imaging Used to Study Detailed Bone Apposition and Fixation of an Additively Manufactured Cementless Acetabular Shell

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          The ability to utilize magnetic resonance imaging (MRI) to assess bony fixation in 3 dimensions may allow a better understanding of the implant design and bony integration. We hypothesized that a new 3-dimensionally printed cementless highly porous acetabular component (Stryker Trident II Tritanium TM) would show better fixation than an earlier cup from the same manufacturer as assessed by the noninvasive technique of multispectral MRI.

          Methods

          Multiacquisition variable-resonance image combination selective metal suppression MRI was performed in 19 patients implanted with a new 3-dimensionally printed cup and 20 patients who had received a previous-generation cup from the same manufacturer at 1-year follow-up. Each cup was graded globally as well as by 9 specific zones. Integration grades were performed for each zone: 0, full bone integration; 1, fibrous membrane present; 2, osteolysis; and 3, fluid present. A mixed-effects logistic regression model was used to compare fixation between the 2 groups.

          Results

          All cups in both cohorts showed greater than 90% estimated global bony integration (3-dimensionally printed cups, 99.4%; regular cups 91.6%) with no osteolysis or fluid observed in any cup. The 3-dimensionally printed cup had 1 of 171 zones (0.6%) graded as fibrous membrane present, while the 2-dimensional group had 15 of 180 zones (8.3%) graded as fibrous. Of note, screw hole regions were omitted but may be read as fibrous membrane areas.

          Conclusion

          Using multiacquisition variable-resonance image combination selective MRI, our analysis showed greater osteointegration and less fibrous membrane formation in the 3-dimensionally printed cups than the control group at 1-year follow-up.

          Related collections

          Most cited references27

          • Record: found
          • Abstract: found
          • Article: not found

          Total hip arthroplasties: what are the reasons for revision?

          Primary total hip arthroplasties have reported success rates of greater than 95% in many series with a longer than 10-year follow-up. Revision total hip arthroplasty due to such factors as increased high-activity levels, younger patients undergoing the procedure and increasing life expectancy has become more prevalent. An understanding of the mechanisms and timing of total hip arthroplasty failure can direct efforts aimed at reducing revision rates. This study was conducted to evaluate the indications for revision hip arthroplasty and relate these to the time after the index primary hip arthroplasty. A review of all revision hip arthroplasties at two centres over a 6-year time period identified 225 patients who underwent 237 revisions. The overall mean time to revision was 83 months (range: 0-360 months). The cause of failure was aseptic loosening in 123 hips (51.9%), instability in 40 hips (16.9%) and infection in 37 hips (5.5%). When stratified into two groups (less than 5 years, more than 5 years after the index primary hip arthroplasty), 118 of 237 (50%) revisions occurred in less than 5 years, with 33% due to instability and 24% resulting from infection. The majority of the causes of failure within 5 years in these early revisions were instability and deep infection. The success of hip arthroplasty is likely to be compromized if technical aspects of the surgery for appropriate component positioning and critical protocols to minimise complications such as infection are not given the proper attention.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses.

            Roentgen stereophotogrammetry was used to measure the migration of the centre of the femoral head in 84 cemented Lubinus SP I hip arthroplasties (58 primary operations, 26 revisions). Four to seven years later, seven femoral components had been revised because of painful loosening. These implants showed greater subsidence, medial migration and posterior migration during the first two postoperative years than did the hips which had not been revised. Six months after operation, subsidence of more than 0.33 mm combined with a total migration of more than 0.85 mm predicted an increased risk of subsequent revision; the amount of subsidence at two years was an even better predictor. The probability of revision was greater than 50% if the subsidence at two years was 1.2 mm or more.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Total Hip Arthroplasty - over 100 years of operative history

              Total hip arthroplasty (THA) has completely revolutionized the nature in which the arthritic hip is treated, and is considered to be one of the most successful orthopaedic interventions of its generation. With over 100 years of operative history, this review examines the progression of the operation from its origins, together with highlighting the materials and techniques that have contributed to its development. Knowledge of its history contributes to a greater understanding of THA, such as the reasons behind selection of prosthetic materials in certain patient groups, while demonstrating the importance of critically analyzing research to continually determine best operative practice. Finally, we describe current areas of research being undertaken to further advance techniques and improve outcomes.
                Bookmark

                Author and article information

                Contributors
                Journal
                Arthroplast Today
                Arthroplast Today
                Arthroplasty Today
                Elsevier
                2352-3441
                26 August 2020
                December 2020
                26 August 2020
                : 6
                : 4
                : 694-698
                Affiliations
                [a ]Department of Orthopedic Surgery, Adult Reconstruction and Joint Replacement, Hospital for Special Surgery, New York, NY, USA
                [b ]Department of Radiology and Imaging, Hospital for Special Surgery, New York, NY, USA
                [c ]Department of Biostatistics, Hospital for Special Surgery, New York, NY, USA
                Author notes
                []Corresponding author. Hospital for Special Surgery, 535 East 70th Street, New York, NY 10021, USA. Tel.: +1 212 606 1510. Westrichg@ 123456hss.edu
                Article
                S2352-3441(20)30138-2
                10.1016/j.artd.2020.07.019
                7475047
                db79ea1f-8d3c-4b21-bd51-57554520e9d1
                © 2020 The Authors

                This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

                History
                : 14 April 2020
                : 29 May 2020
                : 11 July 2020
                Categories
                Original Research

                mri,total hip replacement,bone fixation,additively manufactured,acetabular shell,bone integration

                Comments

                Comment on this article