      To the Editor— Leftward on left anterior oblique is not always septal!

      letter
      , MD , , MD, FACC, FHRS
      HeartRhythm Case Reports
      Elsevier

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We read with great interest the article by Iribarne and colleagues 1 which describes a case of a pacemaker lead perforation. There are 2 issues that we would like to highlight, which we felt are vital for your readership. Mechanism of perforation: The authors surmise that despite the lead’s being placed on the true septum, perforation occurred through the interventricular septum, across the left ventricle (LV) chamber, through the LV free wall, and finally through the parietal pericardium and into the pleural space. In reality, however, the published computed tomography (CT) images show that the lead had instead perforated through the right ventricle (RV) free wall that is close to the septum (anteroseptal junction) and not through the interventricular septum (Figure 1A: CT image is taken from Iribarne and colleagues 1 Figure 2, with added arrow showing lead trajectory and dotted line showing LV endocardial border). It is critical for every implanting physician to be aware that a pacemaker lead tip pointing leftward on the left anterior oblique view does not indicate with certainty that the tip is on the ventricular septum. The tip may be on the anteroseptal or inferoseptal junction, which are anatomic recesses located even more leftward than the true ventricular septum owing to the crescent shape of the RV. Lead positioning at these leftward locations carries an increased risk of perforation through the thin free wall of the RV, just as was seen in this case (Figure 1B; asterisk denotes the anteroseptal junction). Figure 1 A: Computed tomography image from Figure 2 in Iribarne and colleagues 1 , with added arrow showing lead trajectory and dotted line showing left ventricular endocardial border. B: Cross-sectional representation of the right ventricle (RV) and left ventricle (LV) on a traditional left anterior oblique view, with the asterisk denoting the anteroseptal junction. L = left; R = right. Management: In this report, an open surgical removal of the perforated lead was performed. The implanting electrophysiologist should be aware that acute and subacute lead perforation can often be managed safely with manual traction alone to retrieve a perforated lead. Pericardiocentesis equipment, subxiphoid access, and echocardiography, as well as surgical back-up, should be immediately available in case of pericardial tamponade, but in most patients, up-front sternotomy or thoracotomy is not necessary. 2

          Related collections

          Most cited references2

          • Record: found
          • Abstract: found
          • Article: not found

          Management of subacute and delayed right ventricular perforation with a pacing or an implantable cardioverter-defibrillator lead.

          The development of small-diameter active fixation pacing and implantable cardioverter-defibrillator leads may be associated with increased risk for delayed right ventricular perforation. The management of this unforeseen complication has been poorly described. Eleven successive patients referred for right ventricular subacute or delayed perforation (no evidence of lead perforation at the time of the procedure, perforation of the right ventricle diagnosed > or =5 days after implantation) were reviewed. The perforation was related to a pacing (n = 7) or an implantable cardioverter-defibrillator (n = 4) lead. The main symptoms were major dyspnea with pericardial effusion requiring emergency pericardial drainage (n = 3), inappropriate implantable cardioverter-defibrillator shock (n = 1), syncope (n = 2), abdominal pain (n = 1), mammary hematoma (n = 1), diaphragm stimulation (n = 1), and chest pain (n = 1). One patient was strictly asymptomatic. Signs of lead dysfunction were observed in all 11 patients. The diagnosis of lead perforation was confirmed by chest x-ray, echocardiography, or computed tomography. Surgery was directly performed in 1 patient with suspicion of digestive perforation. In the remaining 10 patients, the leads were removed by simple traction under fluoroscopic guidance in the operating room, with surgical backup support. The need for close monitoring was highlighted by the occurrence in 1 patient of tamponade requiring percutaneous pericardiocentesis and urgent surgical revision. The postoperative course of these patients was unremarkable. In conclusion, subacute ventricular perforation is a rare but potentially life threatening complication of lead implantation. In most patients, the leads can safely be removed under fluoroscopic guidance, with surgical backup support and close monitoring.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            Right ventricular lead perforation through the septum, left ventricle, and pleura, managed by an open surgical approach

            Key Teaching Points • Even with confirmed right ventricular septal pacemaker lead placement, right ventricular perforation can still occur. • Evaluation for a ventricular perforation should include a computed tomography scan. • Subsequent management is likely to include an open surgical procedure for lead removal. Introduction Cardiac perforation of the right ventricle is an uncommon but potentially serious complication of transvenous pacemaker implantation. When it occurs, right ventricular (RV) perforation is typically seen through the apex. Here, we describe a case of a pacemaker lead that perforated through the RV septum and then the left ventricle, discuss the management of this complication, and provide a brief review of the literature. Case report A 69-year-old woman with a history of asthma, hypothyroidism, anxiety, and oral steroid therapy for reactive airways disease was admitted for evaluation of syncope. Upon admission she was found to have recurrent episodes of sinus bradycardia and, subsequently, a 10-second sinus pause that reproduced syncope. Her baseline electrocardiogram was unremarkable. A transthoracic echocardiogram showed a left ventricular ejection fraction of 73% with no segmental wall motion abnormalities. There was mild-to-moderate aortic stenosis with a peak aortic valve gradient of 25 mm Hg. She was diagnosed with symptomatic sick sinus syndrome and a dual-chamber pacemaker was implanted without immediate complication. The active fixation pacemaker leads (Medtronic “CapSurefix” Model #5076, 52 cm and 45 cm) were placed on the RV septum and right atrial appendage, respectively. The following day, a chest radiograph confirmed satisfactory lead position and device interrogation confirmed excellent pacing parameters. She was discharged home. At a routine 2-week follow-up appointment, the pacemaker was interrogated. Loss of ventricular lead capture at maximal output was seen and it was noted that the pacemaker had automatically switched from bipolar to unipolar pacing in the ventricular channel, just 1 day prior to her follow-up appointment. A wide variation in lead impedances was also observed. A chest radiograph suggested likely RV lead perforation (Figure 1), subsequently confirmed with computed tomography (CT) imaging (Figure 2A–C). The patient was hemodynamically stable and a transthoracic echocardiogram showed no pericardial effusion. Given the location of the lead, the decision was made to proceed to the operating room for surgical pacemaker lead removal. In the operating room, the previous pacemaker incision was opened, the generator exposed, and the ventricular lead detached from the pulse generator. Concurrently, a median sternotomy was performed and the pericardium opened. There was no pericardial effusion. The ventricular lead was found to pass through the ventricular septum, within 1 cm of the left anterior descending artery (Figure 3A), exiting the left ventricle and through the left pleura (Figure 3B). A pledgeted horizontal mattress suture was placed and the lead removed, with good hemostasis. The lead was abutting but not entering the lung parenchyma. A bipolar epicardial ventricular lead was placed and attached to the previous generator, and the patient was discharged home 4 days later, without complication. In follow-up, she has been doing well and has had no further syncopal episodes. Figure 1 Postimplant chest radiograph showing septal right ventricular lead placement (A: posteroanterior view; B: lateral view) on day 0, and subsequent image on day 14 (C: posteroanterior view; D: lateral view) showing overt perforation with extension of lead into the left pleural space. Figure 2 Computed tomography in sagittal section demonstrating ventricular lead (A), exiting the myocardium (B) with tip near the left lingula parenchyma (C). Figure 3 A: Intraoperative image of ventricular lead exiting the left ventricle in close proximity to the left anterior descending artery and entering the left pleura. B: The pleura is opened and retracted with forceps, and the ventricular lead is seen abutting the lingula of the left lung. Discussion RV perforation is a known, although rare and serious, complication of pacemaker lead implantation. 1 Estimates of the frequency of pacemaker lead perforation suggest an incidence of <0.8% 2 and that when these complications occur, they often manifest within 48 hours. Clinically manifest subacute or late perforation (variously described as from greater than 48 hours to greater than 5 days after implantation) 3 is more unusual and might be underdiagnosed. Hirschl and colleagues showed that asymptomatic lead perforation may be present in as many as 15% of patients with pacemakers or defibrillators, with atrial lead perforation being more common than ventricular lead perforation. 4 Interestingly, in this patient, there were no lead pacing characteristics that were suggestive of lead dysfunction with asymptomatic perforation. In various retrospective studies, the risk of perforation has inconsistently appeared to be related to lead type (defibrillator versus pacemaker), fixation mechanism (active versus passive fixation), lead caliber, female sex, age, body habitus, and steroid use. CT scan is considered the gold standard for evaluating for perforation, 5 but there can be difficulties with interpretation secondary to beam-hardening artifact from the lead. There is also no clear consensus on appropriate management of lead perforation in stable patients, as individual characteristics can affect the risks of transvenous lead extraction versus an open surgical procedure.6, 7 It is assumed that placement of a ventricular lead on the RV septum would avoid this complication entirely, as it seems unlikely that a lead would perforate through the thick ventricular septal tissue. In a study of 8 perforations in 1419 patients, RV lead perforation was only seen with apical lead placement. 6 In another study where the majority of RV lead perforations were confirmed to be apical, the RV wall thickness was measured at 1.5 ± 0.5 mm at the apex, with a typical helix length of 1.8 mm for the active fixation leads used, 8 suggestive of a mechanism that requires lead placement in a thin-walled structure for perforation to occur. In our case, the ventricular lead appeared to be well sited on the septum on a postoperative chest radiograph, without excessive redundancy, and yet at surgery the lead was confirmed to have passed through the ventricular septum, below the left anterior descending artery, through the thick free wall of the left ventricle and then through the pericardium. On echocardiogram, there were no unusual features of myocardial tissue that would have suggested an increased tendency to perforation. Conclusion This case illustrates that even with confirmed RV septal pacemaker lead placement, RV perforation can still occur. If the clinical presentation is suggestive, the patient should be evaluated with consideration for a CT scan and possible open surgical procedure for lead removal.
              Bookmark

              Author and article information

              Contributors
              Journal
              HeartRhythm Case Rep
              HeartRhythm Case Rep
              HeartRhythm Case Reports
              Elsevier
              2214-0271
              18 September 2018
              December 2018
              18 September 2018
              : 4
              : 12
              : 604
              Affiliations
              []Section of Cardiac Electrophysiology, Loma Linda University Medical Center, Loma Linda, California
              []Section of Cardiac Electrophysiology, Temple University Hospital, Philadelphia, Pennsylvania
              Article
              S2214-0271(18)30257-4
              10.1016/j.hrcr.2018.06.015
              6301883
              de9bc45f-c798-4dc7-a786-943969a6364c
              © 2018 Heart Rhythm Society. Published by Elsevier Inc.

              This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

              History
              Categories
              Letters to the Editor

              Comments

              Comment on this article