      How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging.

      Neuroimage
      Elsevier BV

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Diffusion tensor imaging is often performed by acquiring a series of diffusion-weighted spin-echo echo-planar images with different direction diffusion gradients. A problem of echo-planar images is the geometrical distortions that obtain near junctions between tissues of differing magnetic susceptibility. This results in distorted diffusion-tensor maps. To resolve this we suggest acquiring two images for each diffusion gradient; one with bottom-up and one with top-down traversal of k-space in the phase-encode direction. This achieves the simultaneous goals of providing information on the underlying displacement field and intensity maps with adequate spatial sampling density even in distorted areas. The resulting DT maps exhibit considerably higher geometric fidelity, as assessed by comparison to an image volume acquired using a conventional 3D MR technique.