      A comprehensive dosimetric evaluation of using RapidArc volumetric‐modulated arc therapy for the treatment of early‐stage nasopharyngeal carcinoma

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The purpose of this study was to investigate the potential benefits of using triple‐arc volumetric‐intensity modulated arc radiotherapy (RapidArc (RA)) for the treatment of early‐stage nasopharyngeal carcinoma (NPC). A comprehensive evaluation was performed including plan quality, integral doses, and peripheral doses. Twenty cases of stage I or II NPC were selected for this study. Nine‐field sliding window IMRT, double‐arc, and triple‐arc RA treatment plans were compared with respect to target coverage, dose conformity, critical organ sparing, and integral doses. Measurement of peripheral doses was performed using thermoluminescent dosimeters in an anthropomorphic phantom. While similar conformity and target coverage were achieved by the three types of plans, triple‐arc RA produced better sparing of parotid glands and spinal cord than double‐arc RA or IMRT. Double‐arc RA plans produced slightly inferior parotid sparing and dose homogeneity than the other two delivery methods. The monitor units (MU) required for triple‐arc were about 50% less than those of IMRT plans, while there was no significant difference in the required MUs between triple‐arc and double‐arc RA plans. The peripheral dose in triple‐arc RA was found to be 50% less compared to IMRT near abdominal and pelvic region. Triple‐arc RA improves both the plan quality and treatment efficiency compared with IMRT for the treatment of early stage NPC. It has become the preferred choice of treatment delivery method for early stage NPC at our center.

          PACS numbers: 87.53.Bn, 87.55.D, 87.55.de, 87.55.dk, 87.56.ng

          Related collections

          Most cited references17

          • Record: found
          • Abstract: found
          • Article: not found

          Volumetric modulated arc therapy: IMRT in a single gantry arc.

          In this work a novel plan optimization platform is presented where treatment is delivered efficiently and accurately in a single dynamically modulated arc. Improvements in patient care achieved through image-guided positioning and plan adaptation have resulted in an increase in overall treatment times. Intensity-modulated radiation therapy (IMRT) has also increased treatment time by requiring a larger number of beam directions, increased monitor units (MU), and, in the case of tomotherapy, a slice-by-slice delivery. In order to maintain a similar level of patient throughput it will be necessary to increase the efficiency of treatment delivery. The solution proposed here is a novel aperture-based algorithm for treatment plan optimization where dose is delivered during a single gantry arc of up to 360 deg. The technique is similar to tomotherapy in that a full 360 deg of beam directions are available for optimization but is fundamentally different in that the entire dose volume is delivered in a single source rotation. The new technique is referred to as volumetric modulated arc therapy (VMAT). Multileaf collimator (MLC) leaf motion and number of MU per degree of gantry rotation is restricted during the optimization so that gantry rotation speed, leaf translation speed, and dose rate maxima do not excessively limit the delivery efficiency. During planning, investigators model continuous gantry motion by a coarse sampling of static gantry positions and fluence maps or MLC aperture shapes. The technique presented here is unique in that gantry and MLC position sampling is progressively increased throughout the optimization. Using the full gantry range will theoretically provide increased flexibility in generating highly conformal treatment plans. In practice, the additional flexibility is somewhat negated by the additional constraints placed on the amount of MLC leaf motion between gantry samples. A series of studies are performed that characterize the relationship between gantry and MLC sampling, dose modeling accuracy, and optimization time. Results show that gantry angle and MLC sample spacing as low as 1 deg and 0.5 cm, respectively, is desirable for accurate dose modeling. It is also shown that reducing the sample spacing dramatically reduces the ability of the optimization to arrive at a solution. The competing benefits of having small and large sample spacing are mutually realized using the progressive sampling technique described here. Preliminary results show that plans generated with VMAT optimization exhibit dose distributions equivalent or superior to static gantry IMRT. Timing studies have shown that the VMAT technique is well suited for on-line verification and adaptation with delivery times that are reduced to approximately 1.5-3 min for a 200 cGy fraction.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Intensity-modulated radiation therapy, protons, and the risk of second cancers.

            Eric Hall (2006)
            Intensity-modulated radiation therapy (IMRT) allows dose to be concentrated in the tumor volume while sparing normal tissues. However, the downside to IMRT is the potential to increase the number of radiation-induced second cancers. The reasons for this potential are more monitor units and, therefore, a larger total-body dose because of leakage radiation and, because IMRT involves more fields, a bigger volume of normal tissue is exposed to lower radiation doses. Intensity-modulated radiation therapy may double the incidence of solid cancers in long-term survivors. This outcome may be acceptable in older patients if balanced by an improvement in local tumor control and reduced acute toxicity. On the other hand, the incidence of second cancers is much higher in children, so that doubling it may not be acceptable. IMRT represents a special case for children for three reasons. First, children are more sensitive to radiation-induced cancer than are adults. Second, radiation scattered from the treatment volume is more important in the small body of the child. Third, the question of genetic susceptibility arises because many childhood cancers involve a germline mutation. The levels of leakage radiation in current Linacs are not inevitable. Leakage can be reduced but at substantial cost. An alternative strategy is to replace X-rays with protons. However, this change is only an advantage if the proton machine employs a pencil scanning beam. Many proton facilities use passive modulation to produce a field of sufficient size, but the use of a scattering foil produces neutrons, which results in an effective dose to the patient higher than that characteristic of IMRT. The benefit of protons is only achieved if a scanning beam is used in which the doses are 10 times lower than with IMRT.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              A conformation number to quantify the degree of conformality in brachytherapy and external beam irradiation: application to the prostate.

              This article presents a method of quantitative assessment of the degree of conformality and its designation by a single numerical value. A conformation number is introduced to evaluate objectively the degree of conformality. A comparison is made between the conformation number as found for external beam treatment plans and ultrasonically guided 125I seed implants for localized prostate cancer. The conformation number in case of a planning target volume irradiated with two opposed open beams, three open beams, and three beams with customized blocks amounted to 0.17, 0.39, and 0.65, respectively. The conformation number as found for ultrasonically guided permanent prostate implants using 125I seeds averaged 0.72. The conformation number is a convenient instrument for indicating the degree of conformality by a single numerical value. Treatments with a conformation number greater than 0.60 might be termed conformal radiotherapy.
                Bookmark

                Author and article information

                Contributors
                kanwkm@ha.org.hk
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                November 2012
                08 November 2012
                : 13
                : 6 ( doiID: 10.1002/acm2.2012.13.issue-6 )
                : 189-202
                Affiliations
                [ 1 ] Department of Oncology Princess Margaret Hospital Hong Kong SAR
                [ 2 ] Department of Physics and Materials Science City University of Hong Kong Hong Kong SAR
                Author notes
                [*] [* ]Corresponding author: Monica W.K. Kan, Rm. 605, 6/F., Block H, Department of Oncology, Princess Margaret Hospital, Kowloon, Hong Kong SAR; phone: (852) 29902776; fax: (852) 29902775; email: kanwkm@ 123456ha.org.hk
                Article
                ACM20189
                10.1120/jacmp.v13i6.3887
                5718553
                23149781
                df1fc70b-1448-4145-9bf6-7a977a1e84aa
                © 2012 The Authors.

                This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 12 January 2012
                : 08 July 2012
                Page count
                Figures: 3, Tables: 4, References: 19, Pages: 14, Words: 6249
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                acm20189
                November 2012
                Converter:WILEY_ML3GV2_TO_NLMPMC version:5.2.5 mode:remove_FC converted:16.11.2017

                rapidarc,triple‐arc,peripheral doses,sliding window imrt,early‐stage nasopharyngeal carcinoma

                Comments

                Comment on this article

                scite_

                Similar content357

                Cited by5

                Most referenced authors150