      Evaluation of the BIOMIC video reader system for routine use in the clinical microbiology laboratory

      ,
      Diagnostic Microbiology and Infectious Disease
      Elsevier BV

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The BIOMIC Video Reader System (Giles Scientific, New York, NY, USA) is a semi-automated AST method that combines disk diffusion testing with automated reading and data interpretation. We evaluated this system with 497 strains comprising a total of 5821 drug organism combinations (DOC) from our routine diagnostic laboratory. Additionally, we compared the time required of the manual and the automated method. The overall agreement of interpretative categories of all DOC was 96.1%. However, comparing complete tests the agreement was only 70.8%. The average time required of the BIOMIC system to complete a test was more than twice as long as that of the manual method. Our data suggest that the tested version of the BIOMIC system cannot be recommended for routine use in diagnostic laboratories.

          Related collections

          Author and article information

          Journal
          Diagnostic Microbiology and Infectious Disease
          Diagnostic Microbiology and Infectious Disease
          Elsevier BV
          07328893
          July 2000
          July 2000
          : 37
          : 3
          : 151-155
          Article
          10.1016/S0732-8893(00)00132-2
          10904186
          e59694fd-0262-461c-a13c-b9b3899b29b9
          © 2000

          https://www.elsevier.com/tdm/userlicense/1.0/

          History

          Comments

          Comment on this article