      Use of a do-it-yourself artificial pancreas system is associated with better glucose management and higher quality of life among adults with type 1 diabetes

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background:

          Previous studies show that the use of do-it-yourself artificial pancreas system (DIYAPS) may be associated with better glycemic control characterized by improved estimated hemoglobin A1c (eHbA1c) and time in range among adults with type 1 diabetes (T1D). However, few studies have demonstrated the changes in laboratory-measured HbA1c, which is a more accepted index for glycemic control, after using a DIYAPS.

          Methods:

          This is a retrospective before-after study approaching patients who reported self-use of AndroidAPS. The main inclusion criteria included: T1D; aged ⩾18 years; having complete record of ⩾3 months of continuous AndroidAPS use; with laboratory-measured HbA1c and quality of life scale data before and after 3 months of AndroidAPS use; and not pregnant. The primary outcome was the change in HbA1c between baseline and 3 months after initiation of AndroidAPS use.

          Results:

          Overall, 15 patients (10 females) were included; the median age was 32.2 years (range: 19.2–69.4), median diabetes duration was 9.7 years (range: 1.8–23.7) and median baseline HbA1c was 7.3% (range: 6.4–10.1). The 3 months of AndroidAPS use was associated with substantial reductions in HbA1c [6.79% (SD: 1.29) versus 7.63% (SD: 1.06), p = 0.002] and glycemic variability when compared with sensor-augmented pump therapy. A lower level of fear of hypoglycemia [22.13 points (SD: 6.87) versus 26.27 points (SD: 5.82), p = 0.010] was also observed after using AndroidAPS.

          Conclusions:

          The 3 months of AndroidAPS use was associated with significant improvements in glucose management and quality of life among adults with T1D.

          Related collections

          Most cited references27

          • Record: found
          • Abstract: found
          • Article: not found

          Associations of fats and carbohydrate intake with cardiovascular disease and mortality in 18 countries from five continents (PURE): a prospective cohort study

          The relationship between macronutrients and cardiovascular disease and mortality is controversial. Most available data are from European and North American populations where nutrition excess is more likely, so their applicability to other populations is unclear.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            Artificial pancreas treatment for outpatients with type 1 diabetes: systematic review and meta-analysis

            Abstract Objective To evaluate the efficacy and safety of artificial pancreas treatment in non-pregnant outpatients with type 1 diabetes. Design Systematic review and meta-analysis of randomised controlled trials. Data sources Medline, Embase, Cochrane Library, and grey literature up to 2 February 2018. Eligibility criteria for selecting studies Randomised controlled trials in non-pregnant outpatients with type 1 diabetes that compared the use of any artificial pancreas system with any type of insulin based treatment. Primary outcome was proportion (%) of time that sensor glucose level was within the near normoglycaemic range (3.9-10 mmol/L). Secondary outcomes included proportion (%) of time that sensor glucose level was above 10 mmol/L or below 3.9 mmol/L, low blood glucose index overnight, mean sensor glucose level, total daily insulin needs, and glycated haemoglobin. The Cochrane Collaboration risk of bias tool was used to assess study quality. Results 40 studies (1027 participants with data for 44 comparisons) were included in the meta-analysis. 35 comparisons assessed a single hormone artificial pancreas system, whereas nine comparisons assessed a dual hormone system. Only nine studies were at low risk of bias. Proportion of time in the near normoglycaemic range (3.9-10.0 mmol/L) was significantly higher with artificial pancreas use, both overnight (weighted mean difference 15.15%, 95% confidence interval 12.21% to 18.09%) and over a 24 hour period (9.62%, 7.54% to 11.7%). Artificial pancreas systems had a favourable effect on the proportion of time with sensor glucose level above 10 mmol/L (−8.52%, −11.14% to −5.9%) or below 3.9 mmol/L (−1.49%, −1.86% to −1.11%) over 24 hours, compared with control treatment. Robustness of findings for the primary outcome was verified in sensitivity analyses, by including only trials at low risk of bias (11.64%, 9.1% to 14.18%) or trials under unsupervised, normal living conditions (10.42%, 8.63% to 12.2%). Results were consistent in a subgroup analysis both for single hormone and dual hormone artificial pancreas systems. Conclusions Artificial pancreas systems are an efficacious and safe approach for treating outpatients with type 1 diabetes. The main limitations of current research evidence on artificial pancreas systems are related to inconsistency in outcome reporting, small sample size, and short follow-up duration of individual trials.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Closed-loop insulin delivery in suboptimally controlled type 1 diabetes: a multicentre, 12-week randomised trial

              Summary Background The achievement of glycaemic control remains challenging for patients with type 1 diabetes. We assessed the effectiveness of day-and-night hybrid closed-loop insulin delivery compared with sensor-augmented pump therapy in people with suboptimally controlled type 1 diabetes aged 6 years and older. Methods In this open-label, multicentre, multinational, single-period, parallel randomised controlled trial, participants were recruited from diabetes outpatient clinics at four hospitals in the UK and two centres in the USA. We randomly assigned participants with type 1 diabetes aged 6 years and older treated with insulin pump and with suboptimal glycaemic control (glycated haemoglobin [HbA1c] 7·5–10·0%) to receive either hybrid closed-loop therapy or sensor-augmented pump therapy over 12 weeks of free living. Training on study insulin pump and continuous glucose monitoring took place over a 4-week run-in period. Eligible subjects were randomly assigned using central randomisation software. Allocation to the two study groups was unblinded, and randomisation was stratified within centre by low (<8·5%) or high (≥8·5%) HbA1c. The primary endpoint was the proportion of time that glucose concentration was within the target range of 3·9–10·0 mmol/L at 12 weeks post randomisation. Analyses of primary outcome and safety measures were done in all randomised patients. The trial is registered with ClinicalTrials.gov, number NCT02523131, and is closed to accrual. Findings From May 12, 2016, to Nov 17, 2017, 114 individuals were screened, and 86 eligible patients were randomly assigned to receive hybrid closed-loop therapy (n=46) or sensor-augmented pump therapy (n=40; control group). The proportion of time that glucose concentration was within the target range was significantly higher in the closed-loop group (65%, SD 8) compared with the control group (54%, SD 9; mean difference in change 10·8 percentage points, 95% CI 8·2 to 13·5; p<0·0001). In the closed-loop group, HbA1c was reduced from a screening value of 8·3% (SD 0·6) to 8·0% (SD 0·6) after the 4-week run-in, and to 7·4% (SD 0·6) after the 12-week intervention period. In the control group, the HbA1c values were 8·2% (SD 0·5) at screening, 7·8% (SD 0·6) after run-in, and 7·7% (SD 0·5) after intervention; reductions in HbA1c percentages were significantly greater in the closed-loop group compared with the control group (mean difference in change 0·36%, 95% CI 0·19 to 0·53; p<0·0001). The time spent with glucose concentrations below 3·9 mmol/L (mean difference in change −0·83 percentage points, −1·40 to −0·16; p=0·0013) and above 10·0 mmol/L (mean difference in change −10·3 percentage points, −13·2 to −7·5; p<0·0001) was shorter in the closed-loop group than the control group. The coefficient of variation of sensor-measured glucose was not different between interventions (mean difference in change −0·4%, 95% CI −1·4% to 0·7%; p=0·50). Similarly, total daily insulin dose was not different (mean difference in change 0·031 U/kg per day, 95% CI −0·005 to 0·067; p=0·09) and bodyweight did not differ (mean difference in change 0·68 kg, 95% CI −0·34 to 1·69; p=0·19). No severe hypoglycaemia occurred. One diabetic ketoacidosis occurred in the closed-loop group due to infusion set failure. Two participants in each study group had significant hyperglycaemia, and there were 13 other adverse events in the closed-loop group and three in the control group. Interpretation Hybrid closed-loop insulin delivery improves glucose control while reducing the risk of hypoglycaemia across a wide age range in patients with suboptimally controlled type 1 diabetes. Funding JDRF, NIHR, and Wellcome Trust.
                Bookmark

                Author and article information

                Contributors
                Role: ConceptualizationRole: Data curationRole: Formal analysisRole: MethodologyRole: SoftwareRole: Writing original draftRole: Writing review editing
                Role: ConceptualizationRole: MethodologyRole: Writing review editing
                Role: ConceptualizationRole: Funding acquisitionRole: MethodologyRole: Writing review editing
                Role: InvestigationRole: Writing review editing
                Role: InvestigationRole: Writing review editing
                Role: InvestigationRole: Writing review editing
                Role: InvestigationRole: Writing review editing
                Role: ConceptualizationRole: Funding acquisitionRole: Project administrationRole: ResourcesRole: Writing review editing
                Journal
                Ther Adv Endocrinol Metab
                Ther Adv Endocrinol Metab
                TAE
                sptae
                Therapeutic Advances in Endocrinology and Metabolism
                SAGE Publications (Sage UK: London, England )
                2042-0188
                2042-0196
                25 August 2020
                2020
                : 11
                : 2042018820950146
                Affiliations
                [1-2042018820950146]Department of Endocrinology and Metabolism, The Third Affiliated Hospital of Sun Yat-sen University, Guangdong Provincial Key Laboratory of Diabetology, Guangzhou, China
                [2-2042018820950146]Department of Endocrinology and Metabolism, The First Affiliated Hospital of University of Science and Technology of China, Division of Life Sciences of Medicine, University of Science and Technology of China, Hefei, China
                [3-2042018820950146]Department of Endocrinology and Metabolism, The First Affiliated Hospital of University of Science and Technology of China, Division of Life Sciences of Medicine, University of Science and Technology of China, Hefei, China
                [4-2042018820950146]Department of Endocrinology, Drum Tower Hospital Affiliated to Nanjing University Medical School, Nanjing, China
                [5-2042018820950146]Department of Endocrinology and Metabolism, The Third Affiliated Hospital of Sun Yat-sen University, Guangdong Provincial Key Laboratory of Diabetology, Guangzhou, China
                [6-2042018820950146]Department of Endocrinology and Metabolism, The Third Affiliated Hospital of Sun Yat-sen University, Guangdong Provincial Key Laboratory of Diabetology, Guangzhou, China
                [7-2042018820950146]Department of Endocrinology and Metabolism, The Third Affiliated Hospital of Sun Yat-sen University, Guangdong Provincial Key Laboratory of Diabetology, Guangzhou, China
                [8-2042018820950146]Department of Endocrinology and Metabolism, The Third Affiliated Hospital of Sun Yat-sen University, Guangdong Provincial Key Laboratory of Diabetology, Guangzhou 510630, China
                [9-2042018820950146]Department of Endocrinology and Metabolism, The First Affiliated Hospital of USTC, Division of Life Sciences of Medicine, University of Science and Technology of China, 17 Lujiang Road, Hefei 230001, People’s Republic of China
                Author notes
                [*]

                These authors contributed equally to this work.

                Author information
                https://orcid.org/0000-0002-0671-1888
                Article
                10.1177_2042018820950146
                10.1177/2042018820950146
                7453453
                32922721
                e633a671-2505-4062-b7f9-e8037c83d029
                © The Author(s), 2020

                This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License ( https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages ( https://us.sagepub.com/en-us/nam/open-access-at-sage).

                History
                : 11 April 2020
                : 23 July 2020
                Funding
                Funded by: fundamental research funds for the central universities, FundRef https://doi.org/10.13039/501100012226;
                Award ID: WK9110000137
                Funded by: National Natural Science Foundation of China, FundRef https://doi.org/10.13039/501100001809;
                Award ID: 81941022
                Funded by: National Natural Science Foundation of China, FundRef https://doi.org/10.13039/501100001809;
                Award ID: Key Programme 81530025
                Funded by: National Key R&D Program of China, ;
                Award ID: 2017YFC1309600
                Categories
                Original Research
                Custom metadata
                January-December 2020
                ts1

                do-it-yourself artificial pancreas system,glycemic control,glycemic variability,quality of life,type 1 diabetes

                Comments

                Comment on this article