      Systemic ventricular implantation of a leadless pacemaker in a patient with a univentricular heart and atrioventricular node calcification

      case-report

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Key Teaching Points • Patients with complex congenital cardiac lesions require special consideration when deciding on the optimal pacemaker system. Implantation of a leadless pacemaker with good thresholds is feasible in patients with univentricular physiology. • The use of multiple imaging modalities helps to guide safe pacemaker implantation and optimization of thresholds. • Anticoagulation prophylaxis is essential in patients with endovascular pacemaker systems and an obligatory right-to-left shunt, although many of these patients may be treated prior to implantation. Introduction The development of a fully encapsulated leadless pacing system with active fixation provides an alternative treatment option for patients requiring pacemaker implantation. To date it has predominantly been used in patients with structurally normal hearts. 1 There is limited experience regarding patients with congenital cardiac disease. Since patients with complex cardiac lesions are surviving into adulthood, incidences of conduction disturbances will continue to increase. The implantation of transvenous endocardial pacemakers in this patient cohort may be complicated by the anatomical substrate, venous access, and thromboembolic considerations. Epicardial systems are frequently employed, but concerns exist regarding their longevity and the risk of lead fracture. The development of a leadless Micra Transcatheter Pacing System (Medtronic, Minneapolis, MN) may overcome some of these challenges. To date, the short-term safety and efficacy of the leadless device has been established in a large adult population with structurally normal hearts. 1 Minimal data exists on the use of the Micra system in patients with congenital heart disease, with only 1 reported case in a patient with single-ventricle physiology. 2 , 3 We present a patient with Ebstein anomaly, palliated with a Blalock-Taussig shunt, who developed late-onset complete heart block secondary to atrioventricular node calcification (Figure 1). He was managed with leadless pacemaker implantation into the systemic ventricle. Figure 1 Transesophageal echocardiography demonstrating nodular calcification. Case report We present the case of a 46-year-old man with Ebstein anomaly, resulting in functional tricuspid atresia. The rudimentary right ventricle (RV) was hypoplastic with severe tricuspid regurgitation (Figure 2). An atrial septal defect was present with right-to-left shunting. He was originally palliated with a classic right Blalock-Taussig shunt at age 3 (1976). Shunt occlusion gradually occurred and, at age 44, a left modified Blalock-Taussig shunt was performed. Figure 2 Representative images from cardiac magnetic resonance imaging highlighting the complex congenital anatomy of the rudimentary right ventricle and left ventricle. A severe stenosis in the right pulmonary artery was stented at the hilum in 2016, through the left modified Blalock-Taussig shunt. During follow-up review, atrial tachycardia was noted in late 2017 and 15 months later he developed complete heart block with junctional rhythm with rates of 50–60 beats per minute. Cardiac magnetic resonance imaging was performed, and severe fibrotic infiltration of the heart was reported. Concerns were raised regarding the risk of thrombus formation from endocardial lead contact and friction with the myocardium and venous system, with a potential high risk of embolization into the systemic circulation through the obligatory right-to-left shunt. With limited published data with regard to the risk of embolization in conventional pacing vs leadless pacing, we could only speculate that the overall risk may be lower with a leadless system, where the risk of lead friction is removed. A surgical epicardial dual-chamber pacemaker was also considered but was declined by the patient and surgeon secondary to extreme erythrocytosis (hemoglobin of 27 g/dL) and resultant bleeding and thrombotic risk. The decision was made to implant the device in the left ventricle (LV) owing to severe fibrosis in the RV. We believed the LV would likely have more viable myocardium for engagement of the tines and better threshold measurements. We recognize the lack of data on the technicalities of implanting a leadless device in the less trabeculated LV. There is also the potential risk of embolization within the systemic ventricle. The patient was carefully counseled on all options available with a full and open discussion. The patient declined a surgical approach and was keen to proceed with leadless pacemaker implantation. The procedure was performed under general anesthetic to provide for transesophageal echo (TOE) guidance. The procedure was carried out using fluoroscopy and TOE. TOE showed an atrial communication with right-to-left shunt and hypoplastic RV. A 12-mm-diameter calcified mass was noted close to the atrioventricular node (Figure 1). The right groin was then prepared and punctured, and two purse string sutures placed. The femoral vein was dilated to 23 French and the Micra introducer inserted. The Micra delivery system catheter was easily passed through the septal defect into the left atrium and further to the LV. During manipulation he developed prolonged asystole, and a temporary wire was introduced into the rudimentary RV via the left femoral vein. The device was positioned in the left ventricular apex with initial high threshold noted. The device was captured and redeployed into the rudimentary RV with adequate separation to the tricuspid valve, but high thresholds were again present. The device was passed across the atrial septum a second time with implantation into the left ventricular apex. Pull test showed good engagement of the tines and satisfactory parameters were achieved. Pacemaker check revealed an R wave of 7.1 mV, threshold of 0.75 V @ 0.24 ms, and impedance of 980 ohms. The device was released and rechecked for stability of the parameters. The delivery system was removed with purse-string closure and effective hemostasis. The patient was recommenced on rivaroxaban anticoagulation following the procedure and remains well. Follow-up threshold of the Micra pacer was 1.13 V @ 0.24 ms (Figure 3). Figure 3 Micra (Medtronic, Minneapolis, MN) implant position in left ventricle with temporary wire in the rudimentary right ventricle. Discussion We have demonstrated that implantation of a leadless pacemaker with good thresholds is feasible in patients with univentricular physiology. It is possible that the use of this device may increase as patients with congenital heart disease survive longer, with an increased incidence of conduction disturbance secondary to disease progression and multiple interventions. 4 Patients with congenital heart disease and rhythm disturbances are a challenging patient group owing to their complex anatomy and embolic risk. Conventional pacing with either epicardial or endocardial systems for patients with univentricular physiology may be suboptimal owing to endovascular thrombosis risk and need for multiple re-do procedures. The leadless device avoids the risk of lead fracture, lead displacement, and endovascular thrombus formation with venous occlusion. Owing to the short pulse width and battery technology, the longevity of the device may exceed conventional devices, resulting in fewer revisions. 1 , 5 Intracardiac thrombus risk is not eliminated with this device, and concern still exists regarding the potential for systemic emboli and stroke in patients with univentricular physiology. Anticoagulation prophylaxis is essential in all patients with endovascular pacemaker systems (including leadless pacemakers) and an obligatory right-to-left shunt in order to minimize this risk. There are no data supporting the use of one anticoagulant over another. Our patient was already managed with rivaroxaban, which we continued post implantation. The use of multiple imaging modalities may guide safe pacemaker implantation and was recently advised in a case report by Ferrero and colleagues. 2 In our patient, TOE provided visual and spatial awareness, which was crucial to the success of the implant. Conclusion Implantation of a leadless device within the systemic ventricle is feasible in patients with complex cardiac anatomy. Long-term thrombotic risks in this context have not been described and these patients will require close follow-up. Leadless pacing offers an alternative approach and may have advantages in specific patients.

          Related collections

          Most cited references4

          • Record: found
          • Abstract: found
          • Article: not found

          Transvenous pacing leads and systemic thromboemboli in patients with intracardiac shunts: a multicenter study.

          The risk of systemic thromboemboli associated with transvenous leads in the presence of an intracardiac shunt is currently unknown. To define this risk, we conducted a multicenter, retrospective cohort study of 202 patients with intracardiac shunts: Sixty-four had transvenous leads (group 1), 56 had epicardial leads (group 2), and 82 had right-to-left shunts but no pacemaker or implantable cardioverter defibrillator leads (group 3). Patient-years were accrued until the occurrence of systemic thromboemboli or study termination. Censoring occurred in the event of complete shunt closure, death, or loss to follow-up. Mean ages for groups 1, 2, and 3 were 33.9+/-18.0, 22.2+/-12.6, and 22.9+/-15.0 years, respectively. Respective oxygen saturations were 91.2+/-9.1%, 88.1+/-8.1%, and 79.7+/-6.7%. During respective median follow-ups of 7.3, 9.3, and 17.0 years, 24 patients had at least 1 systemic thromboembolus: 10 (15.6%), 5 (8.9%), and 9 (11.0%) in groups 1, 2, and 3, respectively. Univariate risk factors were older age (hazard ratio [HR], 1.05; P=0.0001), ongoing phlebotomy (HR, 3.1; P=0.0415), and an transvenous lead (HR, 2.4; P=0.0421). In multivariate, stepwise regression analyses, transvenous leads remained an independent predictor of systemic thromboemboli (HR, 2.6; P=0.0265). In patients with transvenous leads, independent risk factors were older age (HR, 1.05; P=0.0080), atrial fibrillation or flutter (HR, 6.7; P=0.0214), and ongoing phlebotomy (HR, 14.4; P=0.0349). Having had aspirin or warfarin prescribed was not protective. Epicardial leads were, however, associated with higher atrial (P=0.0407) and ventricular (P=0.0270) thresholds and shorter generator longevity (HR, 1.9; P=0.0176). Transvenous leads incur a >2-fold increased risk of systemic thromboemboli in patients with intracardiac shunts.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Update on interventional electrophysiology in congenital heart disease: evolving solutions for complex hearts.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Cardiac Arrhythmias In Congenital Heart Diseases

              Arrhythmias figure prominently among the complications encountered in the varied and diverse population of patients with congenital heart disease, and are the leading cause of morbidity and mortality. The incidence generally increases as the patient ages, with multifactorial predisposing features that may include congenitally malformed or displaced conduction systems, altered hemodynamics, mechanical or hypoxic stress, and residual or postoperative sequelae. The safe and effective management of arrhythmias in congenital heart disease requires a thorough appreciation for conduction system variants, arrhythmia mechanisms, underlying anatomy, and associated physiology. We, therefore, begin this review by presenting the scope of the problem, outlining therapeutic options, and summarizing congenital heart disease-related conduction system anomalies associated with disorders of the sinus node and AV conduction system. Arrhythmias encountered in common forms of congenital heart disease are subsequently discussed. In so doing, we touch upon issues related to risk stratification for sudden death, implantable cardiac devices, catheter ablation, and adjuvant surgical therapy.
                Bookmark

                Author and article information

                Contributors
                Journal
                HeartRhythm Case Rep
                HeartRhythm Case Rep
                HeartRhythm Case Reports
                Elsevier
                2214-0271
                28 January 2020
                May 2020
                28 January 2020
                : 6
                : 5
                : 265-267
                Affiliations
                []Medtronic plc, Dublin, Ireland
                []Children’s Health Ireland (CHI) Crumlin, Dublin, Ireland
                []Beacon Hospital, Dublin, Ireland
                [§ ]The Mater Misericordiae University Hospital, Dublin, Ireland
                Author notes
                [] Address reprint requests and correspondence: Dr Colm Breatnach, Cardiology Fellow, Children’s Health Ireland (CHI) at Crumlin, Cooley Road, Crumlin, Dublin, Ireland, D12N512. Colm.breatnach@ 123456gmail.com
                Article
                S2214-0271(20)30008-7
                10.1016/j.hrcr.2020.01.008
                7244624
                e6b2ebf7-3bdf-4d3a-a416-8b724645d18b
                © 2020 Heart Rhythm Society. Published by Elsevier Inc.

                This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

                History
                Categories
                Case Report

                leadless pacemaker,univentricular physiology,atrioventricular node calcification,congenital heart defect,tricuspid atresia

                Comments

                Comment on this article