      AAPM Medical Physics Practice Guideline 5.a.: Commissioning and QA of Treatment Planning Dose Calculations — Megavoltage Photon and Electron Beams

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education and professional practice of medical physics. The AAPM has more than 8,000 members and is the principal organization of medical physicists in the United States.

          The AAPM will periodically define new practice guidelines for medical physics practice to help advance the science of medical physics and to improve the quality of service to patients throughout the United States. Existing medical physics practice guidelines will be reviewed for the purpose of revision or renewal, as appropriate, on their fifth anniversary or sooner.

          Each medical physics practice guideline represents a policy statement by the AAPM, has undergone a thorough consensus process in which it has been subjected to extensive review, and requires the approval of the Professional Council. The medical physics practice guidelines recognize that the safe and effective use of diagnostic and therapeutic radiology requires specific training, skills, and techniques, as described in each document. Reproduction or modification of the published practice guidelines and technical standards by those entities not providing these services is not authorized.

          The following terms are used in the AAPM practice guidelines:

          • Must and Must Not: Used to indicate that adherence to the recommendation is considered necessary to conform to this practice guideline.

          • Should and Should Not: Used to indicate a prudent practice to which exceptions may occasionally be made in appropriate circumstances.

          Related collections

          Most cited references63

          • Record: found
          • Abstract: found
          • Article: not found

          IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119.

          AAPM Task Group 119 has produced quantitative confidence limits as baseline expectation values for IMRT commissioning. A set of test cases was developed to assess the overall accuracy of planning and delivery of IMRT treatments. Each test uses contours of targets and avoidance structures drawn within rectangular phantoms. These tests were planned, delivered, measured, and analyzed by nine facilities using a variety of IMRT planning and delivery systems. Each facility had passed the Radiological Physics Center credentialing tests for IMRT. The agreement between the planned and measured doses was determined using ion chamber dosimetry in high and low dose regions, film dosimetry on coronal planes in the phantom with all fields delivered, and planar dosimetry for each field measured perpendicular to the central axis. The planar dose distributions were assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop confidence limits for the test results using the concept confidence limit = /mean/ + 1.96sigma. Other facilities can use the test protocol and results as a basis for comparison to this group. Locally derived confidence limits that substantially exceed these baseline values may indicate the need for improved IMRT commissioning.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning.

            In recent years, the sophistication and complexity of clinical treatment planning and treatment planning systems has increased significantly, particularly including three-dimensional (3D) treatment planning systems, and the use of conformal treatment planning and delivery techniques. This has led to the need for a comprehensive set of quality assurance (QA) guidelines that can be applied to clinical treatment planning. This document is the report of Task Group 53 of the Radiation Therapy Committee of the American Association of Physicists in Medicine. The purpose of this report is to guide and assist the clinical medical physicist in developing and implementing a comprehensive but viable program of quality assurance for modern radiotherapy treatment planning. The scope of the QA needs for treatment planning is quite broad, encompassing image-based definition of patient anatomy, 3D beam descriptions for complex beams including multileaf collimator apertures, 3D dose calculation algorithms, and complex plan evaluation tools including dose volume histograms. The Task Group recommends an organizational framework for the task of creating a QA program which is individualized to the needs of each institution and addresses the issues of acceptance testing, commissioning the planning system and planning process, routine quality assurance, and ongoing QA of the planning process. This report, while not prescribing specific QA tests, provides the framework and guidance to allow radiation oncology physicists to design comprehensive and practical treatment planning QA programs for their clinics.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Variation in external beam treatment plan quality: An inter-institutional study of planners and planning systems.

              This study quantifies variation in radiation treatment plan quality for plans generated by a population of treatment planners given very specific plan objectives. A "Plan Quality Metric" (PQM) with 14 submetrics, each with a unique value function, was defined for a prostate treatment plan, serving as specific goals of a hypothetical "virtual physician." The exact PQM logic was distributed to a population of treatment planners (to remove ambiguity of plan goals or plan assessment methodology) as was a predefined computed tomographic image set and anatomic structure set (to remove anatomy delineation as a variable). Treatment planners used their clinical treatment planning system (TPS) to generate their best plan based on the specified goals and submitted their results for analysis. One hundred forty datasets were received and 125 plans accepted and analyzed. There was wide variability in treatment plan quality (defined as the ability of the planners and plans to meet the specified goals) quantified by the PQM. Despite the variability, the resulting PQM distributions showed no statistically significant difference between TPS employed, modality (intensity modulated radiation therapy versus arc), or education and certification status of the planner. The PQM results showed negligible correlation to number of beam angles, total monitor units, years of experience of the planner, or planner confidence. The ability of the treatment planners to meet the specified plan objectives (as quantified by the PQM) exhibited no statistical dependence on technologic parameters (TPS, modality, plan complexity), nor was the plan quality statistically different based on planner demographics (years of experience, confidence, certification, and education). Therefore, the wide variation in plan quality could be attributed to a general "planner skill" category that would lend itself to processes of continual improvement where best practices could be derived and disseminated to improve the mean quality and minimize the variation in any population of treatment planners. Copyright © 2012 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.
                Bookmark

                Author and article information

                Contributors
                Role: Medical Physics Practice Guideline, Chairsmilowitz@humonc.wisc.edu
                Role: AAPM Staff
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                08 September 2015
                September 2015
                : 16
                : 5 ( doiID: 10.1002/acm2.2015.16.issue-5 )
                : 14-34
                Author notes
                [*] [* ] a Corresponding author: Jennifer B. Smilowitz, Medical Physicist, Medical Physics and Human Oncology, University of Wisconsin, Rm B1149, WIMR 1111 Highland Avenue Madison, WI 53705‐2275, USA; phone: (608) 263‐0657, email: smilowitz@ 123456humonc.wisc.edu

                Article
                ACM20014
                10.1120/jacmp.v16i5.5768
                5690154
                26699330
                eb80ad63-f1ba-48bf-831d-0cd9345e22c7
                © 2015 The Authors.

                This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 05 September 2014
                Page count
                Figures: 1, Tables: 10, References: 58, Pages: 21, Words: 10515
                Categories
                AAPM Reports & Documents
                AAPM Reports & Documents
                Custom metadata
                2.0
                acm20014
                September 2015
                Converter:WILEY_ML3GV2_TO_NLMPMC version:5.2.5 mode:remove_FC converted:16.11.2017

                Comments

                Comment on this article