      Quantification of L-ergothioneine in whole blood by hydrophilic interaction ultra-performance liquid chromatography and UV-detection.

      Journal of Separation Science
      Aged, Aged, 80 and over, Chromatography, High Pressure Liquid, instrumentation, methods, Ergothioneine, blood, Female, Humans, Hydrophobic and Hydrophilic Interactions, Male, Spectrophotometry, Ultraviolet

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          A new hydrophilic interaction ultra-performance LC method was established for the whole blood measurement of L-ergothioneine. Chromatographic separation was achieved in a fairly short time, less than 4 min, on a 100 × 2.1 mm Acquity UPLC BEH HILIC 1.7 μm column with a mobile phase consisting of a mixture of 100 mmol/L ammonium acetate/ACN/water (5:85:10, v/v/v) that flowed isocratically at 0.250 mL/min. The LOD and the limit of quantification were 3.85 and 11.67 μmol/L, respectively. The method exhibited linearity in a concentration range of 15.63-1000 μmol/L (R(2) > 0.999). Mean recovery was 96.34% whereas intraassay and interassay precision were 1.52 and 1.82% RSD, respectively. On the whole, the developed method is simple, fast, precise, accurate, and sensitive and may be useful for routine analyses. © 2013 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.