          Abstract

          Background

          The aim of this paper was to conduct a systematic review of the published literatures comparing the use of mechanical chest compression device and manual chest compression during cardiac arrest (CA) with respect to short-term survival outcomes and neurological function.

          Methods

          Databases including MEDLINE, EMBASE, Web of Science and the ClinicalTrials.gov registry were systematically searched. Further references were gathered from cross-references from articles by handsearch. The inclusion criteria for this review must be human prospective controlled studies of adult CA. Random effects models were used to assess the risk ratios and 95 % confidence intervals for return of spontaneous circulation (ROSC), survival to admission and discharge, and neurological function.

          Results

          Twelve trials (9 out-of-hospital and 3 in-hospital studies), involving 11,162 participants, were included in the review. The results of this meta-analysis indicated no differences were found in Cerebral Performance Category (CPC) scores, survival to hospital admission and survival to discharge between manual cardiopulmonary resuscitation (CPR) and mechanical CPR for out-of-hospital CA (OHCA) patients. The data on achieving ROSC in both of in-hospital and out-of-hospital setting suggested poor application of the mechanical device (RR 0.71, [95 % CI, 0.53, 0.97] and 0.87 [95 % CI, 0.81, 0.94], respectively). OHCA patients receiving manual resuscitation were more likely to attain ROSC compared with load-distributing bands chest compression device (RR 0.88, [95 % CI, 0.80, 0.96]). The in-hospital studies suggested increased relative harm with mechanical compressions for ratio of survival to hospital discharge (RR 0.54, [95 % CI 0.29, 0.98]). However, the results were not statistically significant between different kinds of mechanical chest compression devices and manual resuscitation in survival to admission, discharge and CPC scores for OHCA patients and survival to discharge for in-hospital CA patients.

          Conclusions

          The ability to achieve ROSC with mechanical devise was inferior to manual chest compression during resuscitation. The use of mechanical chest compression cannot be recommended as a replacement for manual CPR, but rather a supplemental treatment in an overall strategy for treating CA patients.

          Related collections

          Most cited references42

          • Record: found
          • Abstract: found
          • Article: not found

          Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest.

          Cardiopulmonary resuscitation (CPR) guidelines recommend target values for compressions, ventilations, and CPR-free intervals allowed for rhythm analysis and defibrillation. There is little information on adherence to these guidelines during advanced cardiac life support in the field. To measure the quality of out-of-hospital CPR performed by ambulance personnel, as measured by adherence to CPR guidelines. Case series of 176 adult patients with out-of-hospital cardiac arrest treated by paramedics and nurse anesthetists in Stockholm, Sweden, London, England, and Akershus, Norway, between March 2002 and October 2003. The defibrillators recorded chest compressions via a sternal pad fitted with an accelerometer and ventilations by changes in thoracic impedance between the defibrillator pads, in addition to standard event and electrocardiographic recordings. Adherence to international guidelines for CPR. Chest compressions were not given 48% (95% CI, 45%-51%) of the time without spontaneous circulation; this percentage was 38% (95% CI, 36%-41%) when subtracting the time necessary for electrocardiographic analysis and defibrillation. Combining these data with a mean compression rate of 121/min (95% CI, 118-124/min) when compressions were given resulted in a mean compression rate of 64/min (95% CI, 61-67/min). Mean compression depth was 34 mm (95% CI, 33-35 mm), 28% (95% CI, 24%-32%) of the compressions had a depth of 38 mm to 51 mm (guidelines recommendation), and the compression part of the duty cycle was 42% (95% CI, 41%-42%). A mean of 11 (95% CI, 11-12) ventilations were given per minute. Sixty-one patients (35%) had return of spontaneous circulation, and 5 of 6 patients discharged alive from the hospital had normal neurological outcomes. In this study of CPR during out-of-hospital cardiac arrest, chest compressions were not delivered half of the time, and most compressions were too shallow. Electrocardiographic analysis and defibrillation accounted for only small parts of intervals without chest compressions.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest.

            The survival benefit of well-performed cardiopulmonary resuscitation (CPR) is well-documented, but little objective data exist regarding actual CPR quality during cardiac arrest. Recent studies have challenged the notion that CPR is uniformly performed according to established international guidelines. To measure multiple parameters of in-hospital CPR quality and to determine compliance with published American Heart Association and international guidelines. A prospective observational study of 67 patients who experienced in-hospital cardiac arrest at the University of Chicago Hospitals, Chicago, Ill, between December 11, 2002, and April 5, 2004. Using a monitor/defibrillator with novel additional sensing capabilities, the parameters of CPR quality including chest compression rate, compression depth, ventilation rate, and the fraction of arrest time without chest compressions (no-flow fraction) were recorded. Adherence to American Heart Association and international CPR guidelines. Analysis of the first 5 minutes of each resuscitation by 30-second segments revealed that chest compression rates were less than 90/min in 28.1% of segments. Compression depth was too shallow (defined as <38 mm) for 37.4% of compressions. Ventilation rates were high, with 60.9% of segments containing a rate of more than 20/min. Additionally, the mean (SD) no-flow fraction was 0.24 (0.18). A 10-second pause each minute of arrest would yield a no-flow fraction of 0.17. A total of 27 patients (40.3%) achieved return of spontaneous circulation and 7 (10.4%) were discharged from the hospital. In this study of in-hospital cardiac arrest, the quality of multiple parameters of CPR was inconsistent and often did not meet published guideline recommendations, even when performed by well-trained hospital staff. The importance of high-quality CPR suggests the need for rescuer feedback and monitoring of CPR quality during resuscitation efforts.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found
              Is Open Access

              Mechanical versus manual chest compression for out-of-hospital cardiac arrest (PARAMEDIC): a pragmatic, cluster randomised controlled trial.

              Mechanical chest compression devices have the potential to help maintain high-quality cardiopulmonary resuscitation (CPR), but despite their increasing use, little evidence exists for their effectiveness. We aimed to study whether the introduction of LUCAS-2 mechanical CPR into front-line emergency response vehicles would improve survival from out-of-hospital cardiac arrest.
                Bookmark

                Author and article information

                Contributors
                86 20 87755766-8500 , jing-xl@163.com
                Journal
                Scand J Trauma Resusc Emerg Med
                Scand J Trauma Resusc Emerg Med
                Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine
                BioMed Central (London )
                1757-7241
                1 February 2016
                1 February 2016
                2016
                : 24
                : 10
                Affiliations
                [ ]Department of Emergency, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhongshan 2nd Road, Guangzhou, Guangdong 510080 China
                [ ]Department of Organ Transplantation, The First Affiliated Hospital of Sun Yat-Sen University, Guangzhou, 510080 China
                Article
                202
                10.1186/s13049-016-0202-y
                4736253
                26830837
                efaac17d-b061-4964-9a64-283dc75a8858
                © Li et al. 2016

                Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 12 November 2015
                : 21 January 2016
                Funding
                Funded by: NSFC
                Award ID: 81372022
                Award Recipient :
                Funded by: Medical Scientific Research Foundation of Guangdong Province, China
                Award ID: B2014112
                Award Recipient :
                Categories
                Original Research
                Custom metadata
                © The Author(s) 2016

                Emergency medicine & Trauma
                cardiac arrest,cardiopulmonary resuscitation,device,meta-analysis
                Emergency medicine & Trauma
                cardiac arrest, cardiopulmonary resuscitation, device, meta-analysis

                Comments

                Comment on this article