      The accuracy and reliability of an infusion pump (STC-3121; Terumo Inst., Japan) during hyperbaric oxygenation.

      1 , ,
      Anaesthesia and intensive care

      Read this article at

      ScienceOpenPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We evaluated the accuracy of delivery of one type of infusion pump at three nominal flow rates under two environmental pressures (measured in atmospheres absolute, ATA), one of which is used for hyperbaric oxygenation. Overall, we observed significant differences between the two different pressures (1 ATA and 2 ATA). Changes in the rate of delivery of between about -4% and +2% were observed over the time course of hyperbaric oxygenation, and the inter-group difference was significant in the phases in which pressure was changing. The effect produced by 2 ATA was statistically significant but small, and its influence is likely to be negligible during practical hyperbaric oxygenation therapy. Thus, it would seem that this type of pump would deliver reliably at the nominal rate during hyperbaric oxygenation at these levels.

          Related collections

          Author and article information

          Journal
          Anaesth Intensive Care
          Anaesthesia and intensive care
          0310-057X
          0310-057X
          Feb 2000
          : 28
          : 1
          Affiliations
          [1 ] Division of Emergency Medicine, Kagoshima University Hospital, Japan.
          Article
          1999125
          10701041
          f1e026b4-8df1-4740-88a6-4d00142b4000
          History

          Comments

          Comment on this article