      Rapid characterisation of the inherent dispersibility of respirable powders using dry dispersion laser diffraction

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Graphical abstract

          Abstract

          Understanding and controlling powder de-agglomeration is of great importance in the development of dry powder inhaler (DPI) products. Dry dispersion laser diffraction measures particle size readily under controlled dispersing conditions, but has not been exploited fully to characterise inherent powder dispersibility. The aim of the study was to utilise particle size-dispersing pressure titration curves to characterise powder cohesivity and ease of de-agglomeration. Seven inhaled drug/excipient powders (beclometasone dipropionate, budesonide, fluticasone propionate, lactohale 300, salbutamol base, salmeterol xinafoate and tofimilast) were subjected to a range of dispersing pressures (0.2–4.5 Bar) in the Sympatec HELOS/RODOS laser diffractometer and particle size measurements were recorded. Particle size-primary pressure data were used to determine the pressures required for complete de-agglomeration. The latter were employed as an index of the cohesive strength of the powder (critical primary pressure; CPP), and the curves were modelled empirically to derive the pressure required for 50% de-agglomeration (DA 50). The powders presented a range of CPP (1.0–3.5 Bar) and DA 50 (0.23–1.45 Bar) which appeared to be characteristic for different mechanisms of powder de-agglomeration. This approach has utility as a rapid pre-formulation tool to measure inherent powder dispersibility, in order to direct the development strategy of DPI products.

          Related collections

          Most cited references52

          • Record: found
          • Abstract: found
          • Article: not found

          Regional lung deposition and bronchodilator response as a function of beta2-agonist particle size.

          Aerosol particle size influences the extent, distribution, and site of inhaled drug deposition within the airways. We hypothesized that targeting albuterol to regional airways by altering aerosol particle size could optimize inhaled bronchodilator delivery. In a randomized, double-blind, placebo-controlled study, 12 subjects with asthma (FEV1, 76.8 +/- 11.4% predicted) inhaled technetium-99m-labeled monodisperse albuterol aerosols (30-microg dose) of 1.5-, 3-, and 6-microm mass median aerodynamic diameter, at slow (30-60 L/min) and fast (> 60 L/min) inspiratory flows. Lung and extrathoracic radioaerosol deposition were quantified using planar gamma-scintigraphy. Pulmonary function and tolerability measurements were simultaneously assessed. Clinical efficacy was also compared with unlabeled monodisperse albuterol (15-microg dose) and 200 microg metered-dose inhaler (MDI) albuterol. Smaller particles achieved greater total lung deposition (1.5 microm [56%], 3 microm [50%], and 6 microm [46%]), farther distal airways penetration (0.79, 0.60, and 0.36, respective penetration index), and more peripheral lung deposition (25, 17, and 10%, respectively). However, larger particles (30-microg dose) were more efficacious and achieved greater bronchodilation than 200 microg MDI albuterol (deltaFEV1 [ml]: 6 microm [551], 3 microm [457], 1.5 microm [347], MDI [494]). Small particles were exhaled more (1.5 microm [22%], 3 microm [8%], 6 microm [2%]), whereas greater oropharyngeal deposition occurred with large particles (15, 31, and 43%, respectively). Faster inspiratory flows decreased total lung deposition and increased oropharyngeal deposition for the larger particles, with less bronchodilation. A shift in aerosol distribution to the proximal airways was observed for all particles. Regional targeting of inhaled beta2-agonist to the proximal airways is more important than distal alveolar deposition for bronchodilation. Altering intrapulmonary deposition through aerosol particle size can appreciably enhance inhaled drug therapy and may have implications for developing future inhaled treatments.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Dry powder inhaler formulation.

            A drug product combines pharmacologic activity with pharmaceutical properties. Desirable performance characteristics are physical and chemical stability, ease of processing, accurate and reproducible delivery to the target organ, and availability at the site of action. For the dry powder inhaler (DPI), these goals can be met with a suitable powder formulation, an efficient metering system, and a carefully selected device. This review focuses on the DPI formulation and development process. Most DPI formulations consist of micronized drug blended with larger carrier particles, which enhance flow, reduce aggregation, and aid in dispersion. A combination of intrinsic physicochemical properties, particle size, shape, surface area, and morphology affects the forces of interaction and aerodynamic properties, which in turn determine fluidization, dispersion, delivery to the lungs, and deposition in the peripheral airways. When a DPI is actuated, the formulation is fluidized and enters the patient's airways. Under the influence of inspiratory airflow, the drug particles separate from the carrier particles and are carried deep into the lungs, while the larger carrier particles impact on the oropharyngeal surfaces and are cleared. If the cohesive forces acting on the powder are too strong, the shear of the airflow may not be sufficient to separate the drug from the carrier particles, which results in low deposition efficiency. Advances in understanding of aerosol and solid state physics and interfacial chemistry are moving formulation development from an empirical activity to a fundamental scientific foundation.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              The cohesive-adhesive balances in dry powder inhaler formulations I: Direct quantification by atomic force microscopy.

              To obtain a quantitative assessment of the cohesive and adhesive force balance within dry powder inhaler formulations. The atomic force microscope (AFM) colloid probe technique was used to measure the adhesive and cohesive force characteristics of dry powder systems containing an active component (budesonide, salbutamol sulphate) and alpha-lactose monohydrate. To minimize the variations in contact area between colloid probe and substrates, nanometer smooth crystal surfaces of the drugs and the excipient were prepared. The uniformity in contact area allowed accurate and reproducible force measurements. Cohesive-adhesive balance (CAB) graphs were developed to allow direct comparison of the interaction forces occurring in model carrier-based formulations. A salbutamol sulphate-lactose system revealed a significant tendency for the two materials to adhere, suggesting a propensity for the powder to form a homogenous blend. In contrast, the budesonide-lactose system exhibited strong cohesive properties suggesting that the formulation may exhibit poor blend homogeneity and potential for segregation upon processing and handling. The novel approach provides a fundamental insight into the cohesive-adhesive balances in dry powder formulations and further understanding of powder behavior.
                Bookmark

                Author and article information

                Journal
                Int J Pharm
                Int J Pharm
                International Journal of Pharmaceutics
                Elsevier/North-Holland Biomedical Press
                0378-5173
                1873-3476
                15 April 2013
                15 April 2013
                : 447
                : 1-2
                : 124-131
                Affiliations
                [a ]King's College London, Institute of Pharmaceutical Science, Stamford Street, London SE1 9NH, UK
                [b ]Pfizer Global R&D, Ramsgate Road, Sandwich, Kent CT13 9NJ, UK
                [c ]University of Hertfordshire, Department of Pharmacy, College Lane, Hatfield AL10 9AB, UK
                Author notes
                [* ]Corresponding author. Tel.: +44 01707 285904; fax: +44 01707 284506. d.murnane@ 123456herts.ac.uk
                Article
                IJP13157
                10.1016/j.ijpharm.2013.02.034
                3636537
                23434542
                f1e084c3-2077-4954-b328-c11b8e5ce1d2
                © 2013 Elsevier B.V.

                This document may be redistributed and reused, subject to certain conditions.

                History
                : 12 November 2012
                : 7 February 2013
                : 8 February 2013
                Categories
                Article

                Pharmacology & Pharmaceutical medicine
                dry powder inhaler (dpi),laser diffraction analysis,sympatec helos and rodos,cohesion,inhalation,de-agglomeration

                Comments

                Comment on this article