      Performance of conical abutment (Morse Taper) connection implants: a systematic review.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          In this systematic review, we aimed to compare conical versus nonconical implant-abutment connection systems in terms of their in vitro and in vivo performances. An electronic search was performed using PubMed, Embase, and Medline databases with the logical operators: "dental implant" AND "dental abutment" AND ("conical" OR "taper" OR "cone"). Names of the most common conical implant-abutment connection systems were used as additional key words to detect further data. The search was limited to articles published up to November 2012. Recent publications were also searched manually in order to find any relevant studies that might have been missed using the search criteria noted above. Fifty-two studies met the inclusion criteria and were included in this systematic review. As the data and methods, as well as types of implants used was so heterogeneous, this mitigated against the performance of meta-analysis. In vitro studies indicated that conical and nonconical abutments showed sufficient resistance to maximal bending forces and fatigue loading. However, conical abutments showed superiority in terms of seal performance, microgap formation, torque maintenance, and abutment stability. In vivo studies (human and animal) indicated that conical and nonconical systems are comparable in terms of implant success and survival rates with less marginal bone loss around conical connection implants in most cases. This review indicates that implant systems using a conical implant-abutment connection, provides better results in terms of abutment fit, stability, and seal performance. These design features could lead to improvements over time versus nonconical connection systems.

          Related collections

          Author and article information

          Journal
          J Biomed Mater Res A
          Journal of biomedical materials research. Part A
          1552-4965
          1549-3296
          Feb 2014
          : 102
          : 2
          Affiliations
          [1 ] Department of Oral and Maxillofacial Surgery, University of Erlangen-Nuremberg, Erlangen, Germany.
          Article
          10.1002/jbm.a.34709
          23533139
          f23d8e3c-53d6-446f-9f68-27fd5027c878
          © 2013 Wiley Periodicals, Inc.
          History

          Morse Taper,dental abutment,dental implant,implant-abutment connection,systematic review