            Abstract

            This poster reports the design, implementation and testing of a portable and inexpensive bio-impedance measurement system intended for electrical impedance tomography (EIT) in cell cultures. The system is based on the AD5933 impedance analyser integrated circuit with additional circuitry to enable four-terminal measurement. Initial results of impedance measurements are reported along with an EIT image reconstructed using the open source EIDORS package.

            Content

            Author and article information

            Journal
            ScienceOpen Posters
            ScienceOpen
            16 November 2015
            Article
            10.14293/P2199-8442.1.SOP-ENG.PSMGXT.v1
            26d585c5-7c75-448c-9a03-61787a399aee

            This work has been published open access under Creative Commons Attribution License CC BY 4.0 , which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Conditions, terms of use and publishing policy can be found at www.scienceopen.com .