Blog
About

  • Record: found
  • Abstract: found
  • Poster: not found
Is Open Access

Low Cost Instrumentation for Electrical Impedance Tomography in Biomedical Applications

ScienceOpen Posters

ScienceOpen

This work has been published open access under Creative Commons Attribution License CC BY 4.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Conditions, terms of use and publishing policy can be found at www.scienceopen.com.

Electrical Impedance Tomography, Bioimpedance, Instrumentation

Read this article at

ScienceOpen
Bookmark
      There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

      Abstract

      This poster reports the design, implementation and testing of a portable and inexpensive bio-impedance measurement system intended for electrical impedance tomography (EIT) in cell cultures. The system is based on the AD5933 impedance analyser integrated circuit with additional circuitry to enable four-terminal measurement. Initial results of impedance measurements are reported along with an EIT image reconstructed using the open source EIDORS package.

      Related collections

      Author and article information

      Journal
      10.14293/P2199-8442.1.SOP-ENG.PSMGXT.v1

      Comments

      Comment on this article