519
views
0
recommends
+1 Recommend
1 collections
    0
    shares

      2023 Journal Citation Reports Journal Impact Factor is 0.9. Scopus Citescore 0.8. 

      Interested in becoming a CVIA published author?

      • Platinum Open Access with no APCs. 
      • Fast peer review/Fast publication online after article acceptance.

      Submissions should be made electronically at: https://mc04.manuscriptcentral.com/cvia-journal.

      Please refer to the Author Guidelines at https://cvia-journal.org/instructions-to-authors/ before submission.

       

      scite_
       
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Implantation of an S-ICD in a Patient with a DDD Pacemaker and Congenitally Corrected Transposition of the Great Arteries

      Published
      case-report
      Bookmark

            Abstract

            Subcutaneous implantable cardioverter-defibrillator (S-ICD) therapy has become a viable alternative to conventional transvenous ICD implantation. Patients with congenitally corrected transposition of the great arteries (ccTGA) have a high risk of sudden cardiac death due to malignant arrhythmia. The interaction between the S-ICD system and the transvenous pacemaker system is not fully understood. We report a case of S-ICD implantation in a patient with ccTGA and a DDD pacemaker. The patient was a 30-year-old man with a previously placed pacemaker with diagnoses of congenital heart disease, ccTGA (SLL), left atrioventricular valve insufficiency, and third-degree atrioventricular block. He presented with an out-of-hospital cardiac arrest, and an S-ICD was implanted to prevent sudden cardiac death. Defibrillation checks were performed successfully. We tested the compatibility of the DDD pacemaker with the S-ICD and found that there was no interference between them. In conclusion, an S-ICD system is a reasonable and safe option in patients with ccTGA.

            Main article text

            Using currently available subcutaneous implantable cardioverter defibrillators (S-ICD) in patients with cardiac malposition remains difficult. The feasibility and efficacy remain unknown. We report a case of S-ICD implantation in a patient with congenitally corrected transposition of the great arteries (ccTGA) and a DDD pacemaker. ccTGA is an uncommon cardiac defect, accounting for less than 0.5% of congenital heart disease (CHD) [1, 2]. Most patients may also have one or more anomalies, such as ventricular septal defect, pulmonary stenosis, and abnormalities of the systemic atrioventricular (tricuspid) valve [3].

            The patient was a 30-year-old man with diagnoses of CHD, ccTGA (SLL), left atrioventricular valve insufficiency, and third-degree atrioventricular block. He had left atrioventricular valve replacement and left clavicular fracture fixation at 16 years of age with warfarin as medication. Because of third-degree atrioventricular block, he underwent placement of a transvenous dual-chamber pacemaker (Biotronik Effecta D) in January 2018. He presented with an out-of-hospital cardiac arrest, and ventricular fibrillation was later documented in October 2018. After external electric defibrillation and cardiac compression by the emergency medical team, he was successfully resuscitated. At a local hospital, 24-hour dynamic electrocardiogram (ECG) demonstrated ventricular tachycardia, frequent premature ventricular contractions, and loss of atrial capture. Pacemaker interrogation verified atrial electrode dislocation, and then the pacemaker was set to VVI mode.

            On the basis of the patient’s history, an ICD system was judged to be appropriate for secondary prevention of sudden cardiac arrest. There were detailed discussions between the family and the care team regarding options for ICD placement, and finally a decision was made to implant an S-ICD. Detailed screening for the S-ICD was performed before device placement. Baseline 12-lead ECG demonstrated paced rhythm (VVI, basic rate 70 bpm) and QRS duration of 236 ms. Initial ECG vector screening revealed that the patient would be an optimal candidate on the basis of the screening result with the generator in the left axilla and the electrode placed on the right of the sternum. He also passed screening at higher heart rates, which was paced at 110 bpm.

            An Emblem S-ICD (Boston Scientific, Marlborough, MA, USA) was placed with the generator in the left axilla, the inferior portion of the electrode to the right of the xyphoid, and the tip of the electrode to the immediate right of the sternum. Sensing through the device was noted to be best in the primary vector. The S-ICD was implanted by a three-incision technique with the patient under general anesthesia. Defibrillation testing was performed at the time of implantation. Ventricular fibrillation was induced by a 50-Hz current and was adequately sensed, but was not terminated by the first 65-J shock. Sixteen seconds later, the device successfully converted the arrhythmia back to paced rhythm with an 80-J shock; the charge time was less than 10s and the shock impedance was 70 Ω (Figure 1). A postoperative chest X-ray showed proper position of the lead and S-ICD (Figure 2). After successful implantation, the interaction between the S-ICD and the pacemaker was investigated. When the ventricular lead was programmed to bipolar 3.0 V, unipolar 3.0 V, unipolar 7.5 V, and bipolar 7.5 V, at 0.4 ms, the morphology of the QRS complex recorded by the S-ICD remained quite stable (Figure 3). There was a slight change of the morphology of the QRS complex in the secondary vector, and the S-ICD sensed the QRS complex as noise in the primary vector when paced at unipolar 7.5V. Neither bipolar nor unipolar pacing led to oversensing in any of the vectors in the supine and standing positions. Otherwise, the pacemaker functioned normally after two high-output shocks from the S-ICD. In the postoperative interrogation, the ventricular electrode of the pacemaker was programmed to bipolar pacing. The S-ICD was programmed with a shock zone of 200 bpm and a conditional shock zone of 170 bpm with postshock pacing off. At the 3-month follow-up visit in April 2019, the patient did not report syncope or other discomfort, and had not received any S-ICD shocks.

            Figure 1:

            Induction of Ventricular Fibrillation by a 50-Hz Current and Successful Defibrillation after an 80-J Shock with an Impedance of 70 Ω.

            The first defibrillation with a 65-J shock failed.

            Figure 2:

            Chest Radiographs of the Posteroanterior (A) and Lateral (B) Chest after Subcutaneous Implantable Cardioverter-Defibrillator Placement.

            Figure 3:

            Real-Time Electrogram Recordings from the Subcutaneous Implantable Cardioverter-Defibrillator System During Different Pacing Configurations at all Three Vectors.

            (A) Primary vector when the pacemaker was programmed to VVI mode, bipolar 3.0 V, unipolar 3.0 V, unipolar 7.5 V, and bipolar 7.5 V (from top to bottom) at 0.4 ms. (B) Alternate vector. (C) Secondary vector.

            Discussion

            Because of a history of recurrent ventricular arrhythmias, the patient met the criteria for ICD implantation. However, there might have been technical difficulties and a high risk of postoperative complications for a traditional transvenous dual-chamber ICD with his complicated status and anatomical challenges. Besides, atrial electrode dislocation was another tricky problem. One solution could be transvenous lead extraction and implantation of a new transvenous ICD, which may cause major damage to the heart. Implantation of a new ICD without lead extraction could be another solution, thus leaving four leads in the vein, which could potentially cause venous occlusion and make it difficult to replace or upgrade the device in the future. Since an S-ICD avoids the need for transvenous leads, use of this technique may provide an excellent alternative to epicardial or transvenous systems in young patients who require ICD placement.

            The S-ICD is compatible with a pacemaker (Boston Scientific, Shanghai, China, pacemakers), which will avoid unipolar pacing in safe mode according to the technical manual for the S-ICD. Moreover, patients requiring permanent pacing and those with a previously implanted functional unipolar pacing system were excluded from S-ICD-related clinical studies [4, 5]. However, several case reports examining the interaction of pacemaker and S-ICD systems have been published. Steinberg et al. [6] demonstrated the feasibility and safety of the addition of an S-ICD to a bipolar epicardial pacemaker system. A case report by Kossidas et al. [7] found that a unipolar epicardial pacing system can be used in combination with an S-ICD, with certain assessment and programming steps. In our case, we found the S-ICD sensed the QRS complex as noise in the primary vector when paced at unipolar 7.5 V at 0.4 ms. One way to overcome this situation is to avoid unipolar pacing at high amplitude. In our postoperative interrogation, the pacemaker was programmed to bipolar pacing. Also, it is recommended to program the pacemaker to bipolar pacing to prevent oversensing.

            In our patient, defibrillation testing revealed excellent sensing during ventricular fibrillation, while ventricular fibrillation was not terminated by the first 65-J shock, although the second defibrillation with an 80-J shock succeeded. In a meta-analysis of the cohort of CHD patients in the IDE study and the EFFORTLESS registry, both patients who failed defibrillation testing at 65 J had d-transposition of the great arteries (d-TGA). The total number of patients with d-TGA was six. However, the sample size of the congenital cohort was small, which complicates any assessment of the efficacy of the S-ICD specifically in d-TGA [8]. After atrial switch surgery, the anatomical right ventricle becomes the systemic ventricle, which is similar to the case for our patient. A right parasternal electrode position may not sufficiently include the dilated and shifting to right systemic (right) ventricle in the shock path (Figure 4), which might explain the unsuccessful shock at 65J. Besides, noncompaction of right ventricular myocardium could also be part of the reason (Figure 4C). Further study of defibrillation testing in CHD patients, especially those with ccTGA or d-TGA, will be necessary in the future.

            Figure 4:

            Postoperative Computed Tomography Scans (A–C) and Reconstruction (D) Showing the Pacemaker Leads, the Subcutaneous Implantable Cardioverter-Defibrillator (S-ICD) Lead, and the Structure of the Heart.

            Noncompaction of right ventricular myocardium (functional left ventricle) was observed at the apex (C, arrow). *, S-ICD electrode; **, pacemaker leads; ***, left atrioventricular valve.

            In this case we reported that an S-ICD system was used in conjunction with a previously placed transvenous DDD pacemaker that provides chronic pacing. Because the S-ICD system avoids the need for transvenous leads, use of this system in young patients with ccTGA and other CHD who require ICD implantation is quite appealing.

            Conflict of Interest

            The authors confirm that there is no conflict of interest.

            References

            1. , , . Congenitally corrected transposition. Orphanet J Rare Dis 2011;6:22.

            2. , , , , , , et al. Birth prevalence of congenital heart disease worldwide: a systematic review and meta-analysis. J Am Coll Cardiol 2011;58:2241–7.

            3. . Transposition of the great arteries. Circulation 2006;114:2699–709.

            4. , , , , , , et al. Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry. Eur Heart J 2014;35:1657–65.

            5. , , , , , , et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation 2013;128:944–53.

            6. , , , . Nothing inside the heart-combining epicardial pacing with the S-ICD. HeartRhythm Case Rep 2015;1:419–23.

            7. , , , . Managing cross talk between a subcutaneous implantable cardioverter-defibrillator and a dual-chamber unipolar pacemaker system. HeartRhythm Case Rep 2017;3(12):579–83.

            8. , , , , , , et al. Outcomes in patients with congenital heart disease receiving the subcutaneous implantable-cardioverter defibrillator. JACC Clin Electrophysiol 2016;2:615–22.

            Author and article information

            Journal
            CVIA
            Cardiovascular Innovations and Applications
            CVIA
            Compuscript (Ireland )
            2009-8782
            2009-8618
            February 2021
            February 2021
            : 5
            : 3
            : 207-212
            Affiliations
            [1] 1Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China
            Author notes
            Correspondence: Cheng-Jun Guo, Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China. E-mail: guochengjunbj@ 123456qq.com
            Article
            cvia.2019.0597
            10.15212/CVIA.2019.0597
            8c0531db-847a-44d7-b9ef-d1fe6434e150
            Copyright © 2021 Cardiovascular Innovations and Applications

            This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 Unported License (CC BY-NC 4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc/4.0/.

            History
            : 7 July 2020
            : 24 September 2020
            : 12 October 2020
            Categories
            Case Reports

            General medicine,Medicine,Geriatric medicine,Transplantation,Cardiovascular Medicine,Anesthesiology & Pain management
            congenitally corrected transposition of the great arteries,pacemaker,Subcutaneous implantable cardioverter-defibrillator

            Comments

            Comment on this article