Background
Implantable cardioverter-defibrillators (ICDs) can be life-saving devices. They can be implanted in patients who have experienced sudden cardiac death or those who may be at elevated risk of sudden cardiac death. Despite their life-saving potential, they are expensive and carry the risk of serious complications, such as pneumothorax, lead displacement, infection, and inappropriate device therapies [1].
In 2013, the American College of Cardiology, the Heart Rhythm Society, and the American Heart Association published joint appropriate use criteria (AUC) outlining clinical scenarios in which ICDs are indicated. The criteria were decided upon by an expert panel who rated clinical scenarios on a scale from 1 to 9 as rarely appropriate (1–3), may be appropriate (4–6), and appropriate (7–9). Clinical scenarios are assigned an appropriateness score (from 1 to 9), which was determined by a consensus of the AUC authors [2].
Prior investigators have discussed that certain clinical scenarios in the AUC are not currently covered by the Medicare national coverage determination. Consequently, submitting such ICDs for reimbursement may be construed as fraudulent despite their being clinically indicated and appropriate [3]. The inability to receive indicated ICD implantation due to coverage and funding discrepancies could be the difference between life and death.
Previous registry-based studies used various criteria to assess appropriateness of ICD implantation. One study found that 22.5% of primary prevention ICD implantations were non-evidence-based implantations if they met any of four criteria: New York Heart Association class IV symptoms, myocardial infarction within 40 days, revascularization within 3 months, or newly diagnosed heart failure at the time of implantation [4]. A later study found that 86% of all primary prevention ICD implantations met the inclusion criteria of major ICD trials [5]. Generally, although the criteria were published in 2013, clinical data evaluating the real-world application of AUC are sparse.
We aimed to determine the prevalence of AUC-deemed rarely appropriate ICD implantation at our facility and then compare these data with previously published estimates, such as those by Al-Khatib et al. [4] and Kaiser et al. [5]. We hypothesized that we would have a high rate of appropriate ICD implantations, with very few rarely appropriate ICD implantations. Furthermore, as reimbursement is often based on clinical guidelines, we aimed to assess the validity of the AUC should these be a determination of reimbursement in the future.
Methods
We performed an Institutional Review Board–approved retrospective medical record review of 286 patients (Figure 1) at our institution. Patients were identified through the McKesson Cardiology system (McKesson Corporation, Las Colinas, TX, USA), which is used in the electrophysiology laboratory. Patients aged at least 18 years with single-chamber or dual-chamber ICDs implanted at our institution in the 3 years following publication of the AUC (March 2013 to March 2016) were evaluated. Patients who were identified as having biventricular ICD upgrades in the McKesson Cardiology system and those who only underwent generator change were excluded. Data collection was supervised by a staff electrophysiologist, and data were entered by several resident physicians and a medical student into an institutional research electronic data capture (REDCap) database [6].
Two physicians independently reviewed and assessed the appropriateness of all 286 ICD implantations on the basis of the AUC and were blinded to the other physician’s assessments. Deidentified data were exported into an Excel spreadsheet, and were then sorted by indication and appropriateness of each ICD implantation for tallying.
Statistical analysis was performed by a chi-square analysis to compare the appropriateness rates by each reviewer.
Results
Baseline demographics are reported in Table 1. The patient population included approximately 70% males with a mean age of 58 years. Seventy percent were white, 2.8% had a myocardial infarction within 40 days, 72% had heart failure at implantation, and the average ejection fraction was 31%. Most patients had New York Heart Association class I symptoms (26%) or class II symptoms (34%). The indications for which ICDs were implanted are listed in Table 2.
Characteristic | Value |
---|---|
Mean age (years) | 58 |
Male | 201 (70.28%) |
White | 201 (70.28%) |
Alcohol abuse | 19 (6.64%) |
Beta blocker | 247 (86.35%) |
ACE or ARB | 219 (76.57%) |
Nitrates | 17 (5.94%) |
Inotropes | 14 (4.90%) |
Hypertension | 197 (68.88%) |
Hyperlipidemia | 151 (52.80%) |
Diabetes mellitus | 93 (32.52%) |
CKD stage 3 or higher | 65 (22.73%) |
Coronary artery disease | 168 (58.74%) |
MI within past 40 days | 8 (2.80%) |
Heart failure at implantation | 216 (72.52%) |
Systolic heart failure | 198 (69.23%) |
Mean ejection fraction | 31% |
GDMT for at least 3 months | 177 (61.89%) |
NYHA class I disease | 75 (26.22%) |
NYHA class II disease | 97 (33.92%) |
NYHA class III disease | 5 (1.75%) |
NYHA class IV disease | 23 (8.04%) |
ACE, angiotensin-converting enzyme; ARB, angiotensin-receptor blocker; CKD, chronic kidney disease: GDMT, guideline-directed medical therapy; MI, myocardial infarction; NYHA, New York Heart Association.
Primary prevention | Secondary prevention | |
---|---|---|
Cardiac allograft vasculopathy | LV dysfunction | Nonsustained VT Syncope |
Ischemic cardiomyopathy/coronary artery disease | <40 days after MI with EF <40% and pacing indication | VF |
>40 days after MI with EF <35% | Hemodynamically unstable VT | |
EF <40% despite optimal GDMT | Sustained hemodynamically stable VT | |
Unrevascularizable disease with VT/VF | ||
Acute MI without revascularization and VT/VF | ||
Acute MI with revascularization and VT/VF | ||
Syncope with inducible sustained VT/VF | ||
EF <40% with syncope | ||
Sustained VT after VT ablation | ||
Nonischemic cardiomyopathy | EF <35% <3 months since diagnosis | VT/VF associated with cocaine abuse |
EF <35% despite optimal GDMT | Sustained hemodynamically stable monomorphic VT | |
Syncope | ||
Idiopathic arrhythmias | VF | |
Hemodynamically unstable VT | ||
Outflow tract tachycardia with normal LV function and unexplained syncope | ||
Generator change | Persistently reduced LV function | Clinically relevant ventricular arrhythmias since implantation |
Normalized LV function | Improved LV function but clinically relevant ventricular arrhythmias since implantation | |
CRT-D system with improved LV function | Normalized LV function but clinically relevant ventricular arrhythmias since implantation | |
Initial secondary prevention ICD with no ventricular arrhythmia since implantation | ||
Inherited disorders | ||
Cardiac sarcoidosis | Primary prevention | |
Myotonic dystrophy | Primary prevention | |
Hypertrophic cardiomyopathy | Risk factors meeting criteria for primary prevention | Sustained VT/VF |
Syncope | ||
LV noncompaction | Primary prevention with EF >35% | EF <40% and syncope |
Brugada syndrome | Sustained VT/VF | |
Arrhythmogenic right ventricular cardiomyopathy | Sustained VT/VF Sustained VT/VF after VT ablation | |
Congenital disorders | ||
Tetralogy of Fallot | EF <50% and nonsustained VT during exercise testing | |
Congenital long QT syndrome | Sustained VT/VF Unexplained syncope | |
Catecholaminergic polymorphic VT | Sustained VT/VF | |
Mitochondrial myopathy | Syncope and nonsustained VT |
CRT-D, cardiac resynchronization therapy—defibrillator; EF, ejection fraction; GDMT, guideline-directed medical therapy; LV, left ventricular; MI, myocardial infarction; VF, ventricular fibrillation; VT, ventricular tachycardia.
The reviewers (reviewer 1 and reviewer 2, respectively) determined that 89.5% (n = 256) and 89.2% (n = 255) of implantations were appropriate, 5.6% (n = 16) and 7.3% (n = 21) may be appropriate, and 1.8% (n = 5) and 2.1% (n = 6) were rarely appropriate. For the remaining 3.2% (n = 9) and 1.4% (n = 4), respectively, there was no ICD indication listed in the AUC (Figure 2, Table 3) (P = 0.44 by chi-square analysis for all comparisons).
Reviewer 1 | Reviewer 2 | |
---|---|---|
Appropriate | 256 (89.5%) | 255 (89.2%) |
May be appropriate | 16 (5.6%) | 21 (7.3%) |
Rarely appropriate | 5 (1.8%) | 6 (2.1%) |
No indication listed | 9 (3.2%) | 4 (1.4%) |
Of the total 286 ICD implantations, the reviewers (reviewer 1 and reviewer 2, respectively) deemed 170 and 176 ICDs to be implanted for primary prevention. They found that 89.4% (n = 152) and 88.6% (n = 88.6%) of implantations were appropriate, 8.2% (n = 14) and 9.7% (n = 17) may be appropriate, 1.2% (n = 2) and 1.1% (n = 2) were rarely appropriate. For 1.2% (n = 2) and 0.6% (n = 1) of ICD implantations, the reviewers were unable to find any ICD indication in the AUC document (Figure 3, Table 4). Per the criteria of Al-Khatib et al. [4] for “non-evidence-based” ICD implantation, the reviewers found that only 3.5% (n = 6) and 3.4% (n = 6) were “non-evidenced-based” implantations (P = 0.90 by chi-square analysis for all comparisons).
Reviewer 1 (n = 170) | Reviewer 2 (n = 176) | |
---|---|---|
Appropriate | 152 (89.4%) | 156 (88.6%) |
May be appropriate | 14 (8.2%) | 17 (9.7%) |
Rarely appropriate | 2 (1.2%) | 2 (1.1%) |
No indication listed | 2 (1.2%) | 1 (0.6%) |
The reviewers (reviewer 1 and reviewer 2, respectively) deemed 116 and 110 ICDs to be implanted for secondary prevention. They found 89.7% (n = 104) and 90.0% (n = 99) of implantations to be appropriate, 1.7% (n = 2) and 3.6% (n = 4) may be appropriate, and 2.6% (n = 3) and 3.6% (n = 4) were rarely appropriate. Six percent (n = 7) and 2.7% (n = 3) had no listed indication in the AUC document (Figure 4, Table 5) (P = 0.50 by chi-square analysis for all comparisons).
Reviewer 1 (n = 116) | Reviewer 2 (n = 110) | |
---|---|---|
Appropriate | 104 (89.7%) | 99 (90.0%) |
May be appropriate | 2 (1.7%) | 4 (3.6%) |
Rarely appropriate | 3 (2.6%) | 4 (3.6%) |
No indication listed | 7 (6.0%) | 3 (2.7%) |
There was some interreviewer variability in appropriateness. In 5.24% of cases (n = 15), the reviewers disagreed on the level of appropriateness of the ICD indication. In 2.80% of cases (n = 8), one reviewer deemed an ICD implantation appropriate and the other deemed it rarely appropriate.
Discussion
There was a high proportion of appropriately implanted ICDs and a very low proportion of rarely appropriate ICD implantations for both primary and secondary prevention indications. Our high appropriate implantation rate suggests that the AUC accurately reflect current clinical practice.
Comparing these data with data from prior studies (Table 6), we found an average of 89.4% of ICD implantations in this study were appropriate compared with 86% from a study that assessed ICD implantation on the basis of major ICD trial inclusion criteria [5]. Per the criteria of Al-Khatib et al., only 3–4% of ICD implantations for primary prevention in our study were “non-evidence-based” implantations compared with 22.5% in their study [4]. One drastic difference is that the previous studies were registry analyses with more than 100,000 patients in all practice settings, compared with 286 patients in this study. Furthermore, all ICDs in our study were implanted by electrophysiologists in a university academic center, and thus the familiarity of the implanters with current evidence and guidelines likely contributed to the high appropriateness rate.
Al-Khatib et al. [4] | Kaiser et al. [5] | This study | |
---|---|---|---|
Appropriate primary prevention ICD implantations | NA | 86% | Reviewer 1: 89.5% Reviewer 2: 89.2% |
Non-evidence-based primary prevention ICD implantations | 22.5% | NA | Reviewer 1: 3.4% Reviewer 2: 3.5% |
Study size (n) | 111,707 | 150,264 | 286 |
Hospital setting | University Government Private/community | University Government Private/community | University |
Implanter specialty | Electrophysiology Cardiology Thoracic surgery Other | Electrophysiology Cardiology Surgery Other | Electrophysiology |
ICD, implantable cardioverter-defibrillator, NA, not applicable.
The low rarely appropriate implantation rate (1.2% and 1.1%) of primary preventions is likely due to clear primary prevention indications. Secondary prevention ICDs may have had a higher rate of rarely appropriate implantations (2.6% and 3.6%) as the AUC may not fully reflect complex clinical scenarios often seen at university centers, such as patients with transplant vasculopathy, rare genetic disorders, or congenital heart disease.
The relatively high number of unrated secondary prevention implantations (6.0% and 2.7%) indicates that there are gaps in the AUC. Specific examples include patients with hemodynamically unstable ventricular tachycardia without syncope in a structurally normal heart, as well as cardiac arrest in patients too unstable to undergo ischemic evaluation. In addition, the AUC do not address whether patients with a newly discovered cardiomyopathy who require permanent pacing should also receive an ICD at the time of device implantation, although for these clinical scenarios most clinicians would likely consider ICD implantation as the benefit likely outweighs the risk in this circumstance.
The greatest limitation of this study is the small sample size (n = 286) and the subsequent limitation in power, which may limit generalizability. Baseline demographics were similar to those of other major ICD trials [7–9]. Furthermore, as this study examined ICD implantation in a university setting in which ICDs are implanted only by electrophysiologists, it may not represent clinical practice in settings in which ICDs may be implanted by nonelectrophysiologists, such as general cardiologists and thoracic surgeons [4, 5].
Some patients had ICDs implanted for secondary prevention, but if there were no clear or appropriate secondary prevention indications, the reviewers may have instead listed a primary prevention indication, which would falsely decrease the rate of rarely appropriate or unrated secondary prevention ICD implantation. Similarly, the differences in appropriateness rates between reviewers may have been due to differences in how strictly each reviewer interpreted the AUC. For example, the reviewers disagreed on whether class IV symptoms were a contraindication, regardless of whether the patient was listed for transplantation. Of note, the most recent ICD guidelines [10], which were published in 2018 after the reviewers had assessed appropriateness for this study, recommend ICD implantation in patients who are candidates for transplantation or ventricular assist device implantation; however, this recommendation was not included in prior guidelines [11]. The updated guidelines also expand recommendations for ICD implantation in patients with genetic, neuromuscular, and congenital heart disorders [10], some of which were not included in the AUC [2].
Lastly, given that the reviewers were physicians at the academic center studied, there may have been a degree of observer bias involved in rating appropriateness. The differences in appropriateness rates between reviewers were not statistically significant, with P = 0.44 for all ICD implantations, P = 0.90 for primary prevention ICD implantations, and P = 0.50 for secondary prevention ICD implantations.
The Medicare national coverage determination determines reimbursement for ICDs based on indication. It is based on major ICD trial criteria from MADIT, MADIT II, MUSTT, and SCD-HeFT [3], although indications were last added in 2005 [12]. Should the Medicare national coverage determination be based on AUC in the future, any gaps or inadequacies of the AUC would negatively impact patients, clinicians, and hospitals if clinically indicated ICDs are not reimbursable.
When one is considering the application of these findings to clinical practice, it is important that each patient receive individualized care, irrespective of the guidelines. Although the AUC may aid clinical practice and decision-making, they cannot provide guidance in every scenario for every patient. If an indication is not listed or rarely appropriate, clinical judgment must determine whether a specific patient would benefit from ICD implantation.
Conclusion
Compared with prior registry data reports, the prevalence of rarely appropriate ICD implantation at our facility was very low. Our high appropriate use rate could be explained by appropriate clinical practice or by the AUC being evidence based and reflecting clinical practice. The AUC have gaps and could benefit from additional indications regarding secondary prevention. Most importantly, clinical judgement and individualized care should determine which patients receive ICDs irrespective of guidelines or criteria.