386
views
0
recommends
+1 Recommend
1 collections
    1
    shares

      2023 Journal Citation Reports Journal Impact Factor is 0.9. Scopus Citescore 0.8. 

      Interested in becoming a CVIA published author?

      • Platinum Open Access with no APCs. 
      • Fast peer review/Fast publication online after article acceptance.

      Submissions should be made electronically at: https://mc04.manuscriptcentral.com/cvia-journal.

      Please refer to the Author Guidelines at https://cvia-journal.org/instructions-to-authors/ before submission.

       

      scite_
       
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Better Than You Think—Appropriate Use of Implantable Cardioverter-Defibrillators at a Single Academic Center: A Retrospective Review

      Published
      review-article
      Bookmark

            Abstract

            Background: Implantable cardioverter-defibrillators (ICDs) can be life-saving devices, although they are expensive and may cause complications. In 2013, several professional societies published joint appropriate use criteria (AUC) assessing indications for ICD implantation. Data evaluating the clinical application of AUC are limited. Previous registry-based studies estimated that 22.5% of primary prevention ICD implantations were “non-evidence-based” implantations. On the basis of AUC, we aimed to determine the prevalence of “rarely appropriate” ICD implantation at our institution for comparison with previous estimates.

            Methods: We reviewed 286 patients who underwent ICD implantation between 2013 and 2016. Appropriateness of each ICD implantation was assessed by independent review and rated on the basis of AUC.

            Results: Of 286 ICD implantations, two independent reviewers found that 89.5% and 89.2%, respectively, were appropriate, 5.6% and 7.3% may be appropriate, and 1.8% and 2.1% were rarely appropriate. No AUC indication was found for 3.5% and 3.4% of ICD implantations, respectively. Secondary prevention ICD implantations were more likely rarely appropriate (2.6% vs. 1.2% and 3.6% vs. 1.1%) or unrated (6.0% vs. 1.2% and 2.7% vs. 0.6%). The reviewers found 3.5% and 3.4% of ICD implantations, respectively, were non-evidence-based implantations. The difference in rates between reviewers was not statistically significant.

            Conclusion: Compared with prior reports, our prevalence of rarely appropriate ICD implantation was very low. The high appropriate use rate could be explained by the fact that AUC are based on current clinical practice. The AUC could benefit from additional secondary prevention indications. Most importantly, clinical judgement and individualized care should determine which patients receive ICDs irrespective of guidelines or criteria.

            Main article text

            Background

            Implantable cardioverter-defibrillators (ICDs) can be life-saving devices. They can be implanted in patients who have experienced sudden cardiac death or those who may be at elevated risk of sudden cardiac death. Despite their life-saving potential, they are expensive and carry the risk of serious complications, such as pneumothorax, lead displacement, infection, and inappropriate device therapies [1].

            In 2013, the American College of Cardiology, the Heart Rhythm Society, and the American Heart Association published joint appropriate use criteria (AUC) outlining clinical scenarios in which ICDs are indicated. The criteria were decided upon by an expert panel who rated clinical scenarios on a scale from 1 to 9 as rarely appropriate (1–3), may be appropriate (4–6), and appropriate (7–9). Clinical scenarios are assigned an appropriateness score (from 1 to 9), which was determined by a consensus of the AUC authors [2].

            Prior investigators have discussed that certain clinical scenarios in the AUC are not currently covered by the Medicare national coverage determination. Consequently, submitting such ICDs for reimbursement may be construed as fraudulent despite their being clinically indicated and appropriate [3]. The inability to receive indicated ICD implantation due to coverage and funding discrepancies could be the difference between life and death.

            Previous registry-based studies used various criteria to assess appropriateness of ICD implantation. One study found that 22.5% of primary prevention ICD implantations were non-evidence-based implantations if they met any of four criteria: New York Heart Association class IV symptoms, myocardial infarction within 40 days, revascularization within 3 months, or newly diagnosed heart failure at the time of implantation [4]. A later study found that 86% of all primary prevention ICD implantations met the inclusion criteria of major ICD trials [5]. Generally, although the criteria were published in 2013, clinical data evaluating the real-world application of AUC are sparse.

            We aimed to determine the prevalence of AUC-deemed rarely appropriate ICD implantation at our facility and then compare these data with previously published estimates, such as those by Al-Khatib et al. [4] and Kaiser et al. [5]. We hypothesized that we would have a high rate of appropriate ICD implantations, with very few rarely appropriate ICD implantations. Furthermore, as reimbursement is often based on clinical guidelines, we aimed to assess the validity of the AUC should these be a determination of reimbursement in the future.

            Methods

            We performed an Institutional Review Board–approved retrospective medical record review of 286 patients (Figure 1) at our institution. Patients were identified through the McKesson Cardiology system (McKesson Corporation, Las Colinas, TX, USA), which is used in the electrophysiology laboratory. Patients aged at least 18 years with single-chamber or dual-chamber ICDs implanted at our institution in the 3 years following publication of the AUC (March 2013 to March 2016) were evaluated. Patients who were identified as having biventricular ICD upgrades in the McKesson Cardiology system and those who only underwent generator change were excluded. Data collection was supervised by a staff electrophysiologist, and data were entered by several resident physicians and a medical student into an institutional research electronic data capture (REDCap) database [6].

            Figure 1:

            Search and Exclusion Criteria.

            ERI, elective replacement indicator; ICD, implantable cardioverter-defibrillator.

            Two physicians independently reviewed and assessed the appropriateness of all 286 ICD implantations on the basis of the AUC and were blinded to the other physician’s assessments. Deidentified data were exported into an Excel spreadsheet, and were then sorted by indication and appropriateness of each ICD implantation for tallying.

            Statistical analysis was performed by a chi-square analysis to compare the appropriateness rates by each reviewer.

            Results

            Baseline demographics are reported in Table 1. The patient population included approximately 70% males with a mean age of 58 years. Seventy percent were white, 2.8% had a myocardial infarction within 40 days, 72% had heart failure at implantation, and the average ejection fraction was 31%. Most patients had New York Heart Association class I symptoms (26%) or class II symptoms (34%). The indications for which ICDs were implanted are listed in Table 2.

            Table 1

            Baseline Characteristics of Patients.

            CharacteristicValue
            Mean age (years)58
            Male201 (70.28%)
            White201 (70.28%)
            Alcohol abuse19 (6.64%)
            Beta blocker247 (86.35%)
            ACE or ARB219 (76.57%)
            Nitrates17 (5.94%)
            Inotropes14 (4.90%)
            Hypertension197 (68.88%)
            Hyperlipidemia151 (52.80%)
            Diabetes mellitus93 (32.52%)
            CKD stage 3 or higher65 (22.73%)
            Coronary artery disease168 (58.74%)
            MI within past 40 days8 (2.80%)
            Heart failure at implantation216 (72.52%)
            Systolic heart failure198 (69.23%)
            Mean ejection fraction31%
            GDMT for at least 3 months177 (61.89%)
            NYHA class I disease75 (26.22%)
            NYHA class II disease97 (33.92%)
            NYHA class III disease5 (1.75%)
            NYHA class IV disease23 (8.04%)

            ACE, angiotensin-converting enzyme; ARB, angiotensin-receptor blocker; CKD, chronic kidney disease: GDMT, guideline-directed medical therapy; MI, myocardial infarction; NYHA, New York Heart Association.

            Table 2

            Implantable Cardioverter-Defibrillator (ICD) Indications.

            Primary preventionSecondary prevention
            Cardiac allograft vasculopathyLV dysfunctionNonsustained VT
            Syncope
            Ischemic cardiomyopathy/coronary artery disease<40 days after MI with EF <40% and pacing indicationVF
            >40 days after MI with EF <35%Hemodynamically unstable VT
            EF <40% despite optimal GDMTSustained hemodynamically stable VT
            Unrevascularizable disease with VT/VF
            Acute MI without revascularization and VT/VF
            Acute MI with revascularization and VT/VF
            Syncope with inducible sustained VT/VF
            EF <40% with syncope
            Sustained VT after VT ablation
            Nonischemic cardiomyopathyEF <35% <3 months since diagnosisVT/VF associated with cocaine abuse
            EF <35% despite optimal GDMTSustained hemodynamically stable monomorphic VT
            Syncope
            Idiopathic arrhythmiasVF
            Hemodynamically unstable VT
            Outflow tract tachycardia with normal LV function and unexplained syncope
            Generator changePersistently reduced LV functionClinically relevant ventricular arrhythmias since implantation
            Normalized LV functionImproved LV function but clinically relevant ventricular arrhythmias since implantation
            CRT-D system with improved LV functionNormalized LV function but clinically relevant ventricular arrhythmias since implantation
            Initial secondary prevention ICD with no ventricular arrhythmia since implantation
            Inherited disorders
            Cardiac sarcoidosisPrimary prevention
            Myotonic dystrophyPrimary prevention
            Hypertrophic cardiomyopathyRisk factors meeting criteria for primary preventionSustained VT/VF
            Syncope
            LV noncompactionPrimary prevention with EF >35%EF <40% and syncope
            Brugada syndromeSustained VT/VF
            Arrhythmogenic right ventricular cardiomyopathySustained VT/VF
            Sustained VT/VF after VT ablation
            Congenital disorders
            Tetralogy of FallotEF <50% and nonsustained VT during exercise testing
            Congenital long QT syndromeSustained VT/VF
            Unexplained syncope
            Catecholaminergic polymorphic VTSustained VT/VF
            Mitochondrial myopathySyncope and nonsustained VT

            CRT-D, cardiac resynchronization therapy—defibrillator; EF, ejection fraction; GDMT, guideline-directed medical therapy; LV, left ventricular; MI, myocardial infarction; VF, ventricular fibrillation; VT, ventricular tachycardia.

            The reviewers (reviewer 1 and reviewer 2, respectively) determined that 89.5% (n = 256) and 89.2% (n = 255) of implantations were appropriate, 5.6% (n = 16) and 7.3% (n = 21) may be appropriate, and 1.8% (n = 5) and 2.1% (n = 6) were rarely appropriate. For the remaining 3.2% (n = 9) and 1.4% (n = 4), respectively, there was no ICD indication listed in the AUC (Figure 2, Table 3) (P = 0.44 by chi-square analysis for all comparisons).

            Figure 2:

            Appropriateness of All Implantable Cardioverter-Defibrillator Implantations (n = 286).

            Table 3

            Appropriateness of All Implantable Cardioverter-Defibrillator Implantations.

            Reviewer 1Reviewer 2
            Appropriate256 (89.5%)255 (89.2%)
            May be appropriate16 (5.6%)21 (7.3%)
            Rarely appropriate5 (1.8%)6 (2.1%)
            No indication listed9 (3.2%)4 (1.4%)

            Of the total 286 ICD implantations, the reviewers (reviewer 1 and reviewer 2, respectively) deemed 170 and 176 ICDs to be implanted for primary prevention. They found that 89.4% (n = 152) and 88.6% (n = 88.6%) of implantations were appropriate, 8.2% (n = 14) and 9.7% (n = 17) may be appropriate, 1.2% (n = 2) and 1.1% (n = 2) were rarely appropriate. For 1.2% (n = 2) and 0.6% (n = 1) of ICD implantations, the reviewers were unable to find any ICD indication in the AUC document (Figure 3, Table 4). Per the criteria of Al-Khatib et al. [4] for “non-evidence-based” ICD implantation, the reviewers found that only 3.5% (n = 6) and 3.4% (n = 6) were “non-evidenced-based” implantations (P = 0.90 by chi-square analysis for all comparisons).

            Figure 3:

            Appropriateness of Primary Prevention Implantable Cardioverter-Defibrillator Implantations.

            Table 4

            Appropriateness of Primary Prevention Implantable Cardioverter-Defibrillator Implantations.

            Reviewer 1 (n = 170)Reviewer 2 (n = 176)
            Appropriate152 (89.4%)156 (88.6%)
            May be appropriate14 (8.2%)17 (9.7%)
            Rarely appropriate2 (1.2%)2 (1.1%)
            No indication listed2 (1.2%)1 (0.6%)

            The reviewers (reviewer 1 and reviewer 2, respectively) deemed 116 and 110 ICDs to be implanted for secondary prevention. They found 89.7% (n = 104) and 90.0% (n = 99) of implantations to be appropriate, 1.7% (n = 2) and 3.6% (n = 4) may be appropriate, and 2.6% (n = 3) and 3.6% (n = 4) were rarely appropriate. Six percent (n = 7) and 2.7% (n = 3) had no listed indication in the AUC document (Figure 4, Table 5) (P = 0.50 by chi-square analysis for all comparisons).

            Figure 4:

            Appropriateness of Secondary Prevention Implantable Cardioverter-Defibrillator Implantations.

            Table 5

            Appropriateness of Secondary Prevention Implantable Cardioverter-Defibrillator Implantations.

            Reviewer 1 (n = 116)Reviewer 2 (n = 110)
            Appropriate104 (89.7%)99 (90.0%)
            May be appropriate2 (1.7%)4 (3.6%)
            Rarely appropriate3 (2.6%)4 (3.6%)
            No indication listed7 (6.0%)3 (2.7%)

            There was some interreviewer variability in appropriateness. In 5.24% of cases (n = 15), the reviewers disagreed on the level of appropriateness of the ICD indication. In 2.80% of cases (n = 8), one reviewer deemed an ICD implantation appropriate and the other deemed it rarely appropriate.

            Discussion

            There was a high proportion of appropriately implanted ICDs and a very low proportion of rarely appropriate ICD implantations for both primary and secondary prevention indications. Our high appropriate implantation rate suggests that the AUC accurately reflect current clinical practice.

            Comparing these data with data from prior studies (Table 6), we found an average of 89.4% of ICD implantations in this study were appropriate compared with 86% from a study that assessed ICD implantation on the basis of major ICD trial inclusion criteria [5]. Per the criteria of Al-Khatib et al., only 3–4% of ICD implantations for primary prevention in our study were “non-evidence-based” implantations compared with 22.5% in their study [4]. One drastic difference is that the previous studies were registry analyses with more than 100,000 patients in all practice settings, compared with 286 patients in this study. Furthermore, all ICDs in our study were implanted by electrophysiologists in a university academic center, and thus the familiarity of the implanters with current evidence and guidelines likely contributed to the high appropriateness rate.

            Table 6

            Comparison with Prior Registry-Based Studies.

            Al-Khatib et al. [4]Kaiser et al. [5]This study
            Appropriate primary prevention ICD implantationsNA86%Reviewer 1: 89.5%
            Reviewer 2: 89.2%
            Non-evidence-based primary prevention ICD implantations22.5%NAReviewer 1: 3.4%
            Reviewer 2: 3.5%
            Study size (n)111,707150,264286
            Hospital settingUniversity
            Government
            Private/community
            University
            Government
            Private/community
            University
            Implanter specialtyElectrophysiology
            Cardiology
            Thoracic surgery
            Other
            Electrophysiology
            Cardiology
            Surgery
            Other
            Electrophysiology

            ICD, implantable cardioverter-defibrillator, NA, not applicable.

            The low rarely appropriate implantation rate (1.2% and 1.1%) of primary preventions is likely due to clear primary prevention indications. Secondary prevention ICDs may have had a higher rate of rarely appropriate implantations (2.6% and 3.6%) as the AUC may not fully reflect complex clinical scenarios often seen at university centers, such as patients with transplant vasculopathy, rare genetic disorders, or congenital heart disease.

            The relatively high number of unrated secondary prevention implantations (6.0% and 2.7%) indicates that there are gaps in the AUC. Specific examples include patients with hemodynamically unstable ventricular tachycardia without syncope in a structurally normal heart, as well as cardiac arrest in patients too unstable to undergo ischemic evaluation. In addition, the AUC do not address whether patients with a newly discovered cardiomyopathy who require permanent pacing should also receive an ICD at the time of device implantation, although for these clinical scenarios most clinicians would likely consider ICD implantation as the benefit likely outweighs the risk in this circumstance.

            The greatest limitation of this study is the small sample size (n = 286) and the subsequent limitation in power, which may limit generalizability. Baseline demographics were similar to those of other major ICD trials [79]. Furthermore, as this study examined ICD implantation in a university setting in which ICDs are implanted only by electrophysiologists, it may not represent clinical practice in settings in which ICDs may be implanted by nonelectrophysiologists, such as general cardiologists and thoracic surgeons [4, 5].

            Some patients had ICDs implanted for secondary prevention, but if there were no clear or appropriate secondary prevention indications, the reviewers may have instead listed a primary prevention indication, which would falsely decrease the rate of rarely appropriate or unrated secondary prevention ICD implantation. Similarly, the differences in appropriateness rates between reviewers may have been due to differences in how strictly each reviewer interpreted the AUC. For example, the reviewers disagreed on whether class IV symptoms were a contraindication, regardless of whether the patient was listed for transplantation. Of note, the most recent ICD guidelines [10], which were published in 2018 after the reviewers had assessed appropriateness for this study, recommend ICD implantation in patients who are candidates for transplantation or ventricular assist device implantation; however, this recommendation was not included in prior guidelines [11]. The updated guidelines also expand recommendations for ICD implantation in patients with genetic, neuromuscular, and congenital heart disorders [10], some of which were not included in the AUC [2].

            Lastly, given that the reviewers were physicians at the academic center studied, there may have been a degree of observer bias involved in rating appropriateness. The differences in appropriateness rates between reviewers were not statistically significant, with P = 0.44 for all ICD implantations, P = 0.90 for primary prevention ICD implantations, and P = 0.50 for secondary prevention ICD implantations.

            The Medicare national coverage determination determines reimbursement for ICDs based on indication. It is based on major ICD trial criteria from MADIT, MADIT II, MUSTT, and SCD-HeFT [3], although indications were last added in 2005 [12]. Should the Medicare national coverage determination be based on AUC in the future, any gaps or inadequacies of the AUC would negatively impact patients, clinicians, and hospitals if clinically indicated ICDs are not reimbursable.

            When one is considering the application of these findings to clinical practice, it is important that each patient receive individualized care, irrespective of the guidelines. Although the AUC may aid clinical practice and decision-making, they cannot provide guidance in every scenario for every patient. If an indication is not listed or rarely appropriate, clinical judgment must determine whether a specific patient would benefit from ICD implantation.

            Conclusion

            Compared with prior registry data reports, the prevalence of rarely appropriate ICD implantation at our facility was very low. Our high appropriate use rate could be explained by appropriate clinical practice or by the AUC being evidence based and reflecting clinical practice. The AUC have gaps and could benefit from additional indications regarding secondary prevention. Most importantly, clinical judgement and individualized care should determine which patients receive ICDs irrespective of guidelines or criteria.

            Acknowledgements

            None.

            Ethics Approval and Consent to Participate

            This study was approved by the University of Florida Institutional Review Board (ID no. IRB201600095). Given the deidentified data used in this study, patient consent was not required.

            Consent for Publication

            Not applicable.

            Availability of Data and Material

            The datasets generated and/or analyzed during the current study are not publicly available as they were extracted from University of Florida Health electronic medical records and entered into a REDCap database. Deidentified data were used for subsequent analysis but are available from the corresponding author on reasonable request.

            Conflicts of Interest

            The authors declare that they have no conflicts of interest.

            Funding

            None.

            Author Contributions

            The implantable cardioverter-defibrillators studied were implanted by MK, TAB, WMM, and MM. NHS, DEW, and MM were responsible for study design. NHS, SJR, SANN, JM, and AK were responsible for data collection, which was supervised by MM. NHS analyzed the final dataset. DEW was responsible for statistical analysis. NHS was responsible for drafting the manuscript, and SJR, DEW, TAB, and MM were involved in editing the manuscript. All authors read and approved the final manuscript.

            References

            1. , , , , , , et al. A systematic review of ICD complications in randomised controlled trials versus registries: is our ‘real-world’ data an underestimation? Open Heart 2015;2(1):e000198.

            2. , , , , , , et al. ACCF/HRS/AHA/ASE/HFSA/SCAI/SCCT/SCMR 2013 appropriate use criteria for implantable cardioverter-defibrillators and cardiac resynchronization therapy: a report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, Heart Rhythm Society, American Heart Association, American Society of Echocardiography, Heart Failure Society of America, Society for Cardiovascular Angiography and Interventions, Society of Cardiovascular Computed Tomography, and Society for Cardiovascular Magnetic Resonance. J Am Coll Cardiol 2013;61(12):1318–68.

            3. , , , , , , et al. The disconnect between the guidelines, the appropriate use criteria, and reimbursement coverage decisions: the ultimate dilemma. J Am Coll Cardiol 2014;63(1):12–4.

            4. , , , , , , et al. Non-evidence-based ICD implantations in the United States. J Am Med Assoc 2011;305(1):43–9.

            5. , , , , , , et al. Defibrillator implantations for primary prevention in the United States: inappropriate care or inadequate documentation: insights from the National Cardiovascular Data ICD Registry. Heart Rhythm 2015;12(10):2086–93.

            6. , , , , , . Research electronic data capture (REDCap)—a metadata-driven methodology and workflow process for providing translational research informatics support. J Biomed Inform 2009;42(2):377–81.

            7. , , , , , , et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352(3):225–37.

            8. , , , , , , et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002;346(12):877–83.

            9. , , , , , . A randomized study of the prevention of sudden death in patients with coronary artery disease. N Engl J Med 1999;341(25):1882–90.

            10. , , , , , , et al. 2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Heart Rhythm 2018;15(10):e73–189.

            11. , , , , , , et al. ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death). J Am Coll Cardiol 2006;48(5):e247–346.

            12. US Centers for Medicare & Medicaid Services. National Coverage Determination (NCD) for Implantable Automatic Defibrillators (20.4). Baltimore: US Centers for Medicare & Medicaid Services [cited 2017 August 30]. Available from: https://www.cms.gov/medicare-coverage-database/details/ncd-details.aspx?NCDId---equalsym---110&ncdver---equalsym---3&bc---equalsym---AAAAgAAAAAAAAA%3d%3d&amp .

            Author and article information

            Journal
            CVIA
            Cardiovascular Innovations and Applications
            CVIA
            Compuscript (Ireland )
            2009-8782
            2009-8618
            May 2021
            May 2021
            : 5
            : 4
            : 235-243
            Affiliations
            [1] 1UF Division of Cardiovascular Medicine, 1600 SW Archer Rd, PO Box 100277, Gainesville, FL 32610, USA
            [2] 2UVA Division of Cardiovascular Medicine, PO Box 800158 1215 Lee St. Charlottesville, VA 22908-0158, USA
            [3] 3The Johns Hopkins Hospital Department of Radiology, 601 N Caroline St, Baltimore, MD 21287, USA
            [4] 4The Heart Center, 930 Franklin Street SE, Huntsville, AL, 358015, USA
            [5] 5Intermountain Medical Center, 1380 E Medical Center Dr, Ste 1500, St. George, UT 847906, USA
            [6] 6Carolina Cardiology Consultants, Prisma Health, 1005 Grove Road, Greenville, SC 29605, USA
            Author notes
            Correspondence: Nikhil H. Shah, MD, Division of Cardiovascular Medicine, University of Florida, PO Box 100277, 1600 SW Archer Rd, Gainesville, FL 32610, USA, E-mail: nikhil.shah@ 123456medicine.ufl.edu
            Author information
            https://orcid.org/0000-0003-2717-1427
            https://orcid.org/0000-0002-4169-4876
            https://orcid.org/0000-0003-4805-7375
            https://orcid.org/0000-0002-5224-2891
            Article
            cvia.2021.0005
            10.15212/CVIA.2021.0005
            e01e461b-314a-4f59-895b-1e606efff847
            Copyright © 2021 Cardiovascular Innovations and Applications

            This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 Unported License (CC BY-NC 4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc/4.0/.

            History
            : 25 November 2020
            : 18 January 2021
            : 20 January 2021
            Categories
            Reviews

            General medicine,Medicine,Geriatric medicine,Transplantation,Cardiovascular Medicine,Anesthesiology & Pain management
            implantable cardioverter-defibrillator,Appropriate use,appropriate use criteria

            Comments

            Comment on this article