      Interlaboratory round robin on cantilever calibration for AFM force spectroscopy.

      Ultramicroscopy
      Algorithms, Calibration, Ligands, Microscopy, Atomic Force, instrumentation, methods, Models, Theoretical, Spectrum Analysis, Static Electricity

      Read this article at

      ScienceOpenPublisherPubMed
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Single-molecule force spectroscopy studies performed by Atomic Force Microscopes (AFMs) strongly rely on accurately determined cantilever spring constants. Hence, to calibrate cantilevers, a reliable calibration protocol is essential. Although the thermal noise method and the direct Sader method are frequently used for cantilever calibration, there is no consensus on the optimal calibration of soft and V-shaped cantilevers, especially those used in force spectroscopy. Therefore, in this study we aimed at establishing a commonly accepted approach to accurately calibrate compliant and V-shaped cantilevers. In a round robin experiment involving eight different laboratories we compared the thermal noise and the Sader method on ten commercial and custom-built AFMs. We found that spring constants of both rectangular and V-shaped cantilevers can accurately be determined with both methods, although the Sader method proved to be superior. Furthermore, we observed that simultaneous application of both methods on an AFM proved an accurate consistency check of the instrument and thus provides optimal and highly reproducible calibration. To illustrate the importance of optimal calibration, we show that for biological force spectroscopy studies, an erroneously calibrated cantilever can significantly affect the derived (bio)physical parameters. Taken together, our findings demonstrated that with the pre-established protocol described reliable spring constants can be obtained for different types of cantilevers. Copyright © 2011. Published by Elsevier B.V.

          Related collections

          Author and article information

          Journal
          22094372
          10.1016/j.ultramic.2011.09.012

          Chemistry
          Algorithms,Calibration,Ligands,Microscopy, Atomic Force,instrumentation,methods,Models, Theoretical,Spectrum Analysis,Static Electricity

          Comments

          Comment on this article