Self-expanding nitinol stent (SENS) implantation is commonly oversized in the superficial femoral artery (SFA), and leads to chronic outward force (COF) and in-stent restenosis (ISR). This study aimed to investigate the impact of COF of oversizing SENS on ISR of SFA.
In patients with implanted SENS in SFA, intimal hyperplasia especially between proximal segment and distal segment was evaluated by quantitative angiography, and the impact of COF on mid-term angiographic outcomes was investigated. In addition, porcine model with implanted SENS was used to evaluate the impact of COF on angiographic and histopathologic outcomes at 1 month. Excised stented arteries were evaluated by histopathologic analysis.
We analyzed 65 SENS in 61 patients with follow-up angiography at 6 months to 1 year. The baseline diameter was 6.8 ± 0.71 mm and length were 97.0 ± 33.8 mm for the SENS. The ratio of the diameter of the stent to the reference vessel was 1.3 ± 0.24 at the proximal portion and 1.53 ± 0.27 at the distal portion ( P < 0.001). In the long SFA stent, stent-to-vessel ratio was significantly higher in the distal stent than in the proximal stent (1.3 ± 0.2 vs. 1.55 ± 0.25, P = 0.001). ISR incidence was higher at the distal stent (37.3% vs 52.6%, P = 0.029). All 11 pigs survived for 4 weeks after SENS implantation. The vessel diameter was 4.04 ± 0.40 mm (control group) vs 4.45 ± 0.63 mm (oversized group), and the implanted stent diameter was 5.27 ± 0.46 mm vs. 7.18 ± 0.4 mm ( P = 0.001). The stent-to-vessel diameter ratio was 1.31 ± 0.12 versus 1.63 ± 0.20 ( P < 0.001). After 4 weeks, restenosis % was 29.5 ± 12.9% versus 46.8 ± 21.5% ( P = 0.016). The neointimal area was 5.37 ± 1.15 mm 2 vs. 8.53 ± 5.18 mm 2 ( P = 0.05). The restenosis % was 39.34 ± 8.53% versus 63.97 ± 17.1% ( P = 0.001).