152
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Real-Time Telemetry System for Amperometric and Potentiometric Electrochemical Sensors

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          A real-time telemetry system, which consists of readout circuits, an analog-to-digital converter (ADC), a microcontroller unit (MCU), a graphical user interface (GUI), and a radio frequency (RF) transceiver, is proposed for amperometric and potentiometric electrochemical sensors. By integrating the proposed system with the electrochemical sensors, analyte detection can be conveniently performed. The data is displayed in real-time on a GUI and optionally uploaded to a database via the Internet, allowing it to be accessed remotely. An MCU was implemented using a field programmable gate array (FPGA) to filter noise, transmit data, and provide control over peripheral devices to reduce power consumption, which in sleep mode is 70 mW lower than in operating mode. The readout circuits, which were implemented in the TSMC 0.18-μm CMOS process, include a potentiostat and an instrumentation amplifier (IA). The measurement results show that the proposed potentiostat has a detectable current range of 1 nA to 100 μA, and linearity with an R 2 value of 0.99998 in each measured current range. The proposed IA has a common-mode rejection ratio (CMRR) greater than 90 dB. The proposed system was integrated with a potentiometric pH sensor and an amperometric nitrite sensor for in vitro experiments. The proposed system has high linearity (an R 2 value greater than 0.99 was obtained in each experiment), a small size of 5.6 cm × 8.7 cm, high portability, and high integration.

          Related collections

          Most cited references33

          • Record: found
          • Abstract: not found
          • Book: not found

          Electrochemical Method: Fundamentals and Applications

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            A low-power low-noise cmos for amplifier neural recording applications

              Bookmark
              • Record: found
              • Abstract: not found
              • Book: not found

              The Scientist and Engineer’s Guide to Digital Signal Processing

                Bookmark

                Author and article information

                Journal
                Sensors (Basel)
                Sensors (Basel, Switzerland)
                Molecular Diversity Preservation International (MDPI)
                1424-8220
                2011
                2 September 2011
                : 11
                : 9
                : 8593-8610
                Affiliations
                [1 ] Department of Electrical Engineering, National Cheng Kung University, Tainan 70101, Taiwan; E-Mails: n2895145@ 123456ccmail.ncku.edu.tw (W.-S.W.); bookteam@ 123456hotmail.com (S.-C.C.)
                [2 ] Graduate Institute of Electrical Engineering, National Taipei University, Taipei 10617, Taiwan; E-Mail: hyhuang@ 123456mail.ntpu.edu.tw (H.-Y.H.)
                [3 ] Department of Chemical Engineering, National Taiwan University, Taipei 10617, Taiwan; E-Mails: kcho@ 123456ntu.edu.tw (K.-C.H.); r92524040@ 123456ntu.edu.tw (C.-Y.L.)
                [4 ] Department of Chemical Engineering, Tatung University, Taipei 104, Taiwan; E-Mail: tcchou@ 123456ttu.edu.tw (T.-C.C.)
                [5 ] Department of Chemical Engineering, National Cheng Kung University, Tainan 70101, Taiwan; E-Mail: ilovenba@ 123456ms2.url.com.tw (C.-H.H.)
                [6 ] Department of Computer Science and Information Engineering, National Yunlin University of Science & Technology, Yunlin 64002, Taiwan; E-Mails: wwf@ 123456yuntech.edu.tw (W.-F.W.); wcf9263622@ 123456hotmail.com (C.-F.W.)
                Author notes
                [* ]Author to whom correspondence should be addressed; E-Mail: robin@ 123456ee.ncku.edu.tw ; Tel.: +886-6-275-7575 ext. 62375; Fax: +886-6-236-6433.
                Article
                sensors-11-08593
                10.3390/s110908593
                3231501
                22164093
                17882c36-3560-483c-bdee-6c45b822e608
                © 2011 by the authors; licensee MDPI, Basel, Switzerland.

                This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license ( http://creativecommons.org/licenses/by/3.0/).

                History
                : 8 August 2011
                : 29 August 2011
                : 1 September 2011
                Categories
                Article

                Biomedical engineering
                instrumentation amplifier,telemetry system,potentiostat,electrochemical sensor

                Comments