Key Teaching Points
•
Phrenic nerve injury is a well-known and feared complication of catheter ablation,
where the risk is dependent on the exact anatomical course of the nerve, its relationship
to the endocardial border, and its distance to the site of ablation.
•
Ablation using pulsed field energy instead of conventional radiofrequency or cryothermal
energy might overcome the risk of collateral damage to the phrenic nerve, owing to
its nonthermal and relatively myocardium-selective properties. Clinical data from
pulmonary vein isolation using pulsed field energy seem to confirm these preclinical
findings.
•
The advantages of pulsed field energy in pulmonary vein isolation, where it enhances
the safety of the procedure, can be extrapolated for the ablation of focal atrial
tachycardias with a focus near the phrenic nerve. This overcomes the need for complex
procedures where the phrenic nerve had to be protected by balloon inflation in the
epicardial space or by creating an iatrogenic pneumothorax.
Introduction
Catheter ablation has become the standard of care in the treatment of recurrent focal
atrial tachycardia (FAT), since these patients are poor responders to medical therapy,
and since catheter ablation treatment yields high success rates.
1
This procedure can routinely be performed with low complication rates. However, if
the FAT originates close to the phrenic nerve, injury of this structure, with resultant
impairment of diaphragmatic function, is a well-recognized and feared complication
of catheter ablation.
Multiple approaches to prevent phrenic nerve injury during ablation of FAT have been
described, but these approaches are generally more invasive and therefore more prone
to complications, and are not easy to implement in routine daily practice.2, 3, 4,
5 The problem with conventional energy sources used for ablation of FAT, being radiofrequency
or cryothermal energy, is that they are thermal and nonselective to cardiac tissue,
with the risk of affecting neighboring structures. Pulsed field energy ablation (PFA)
is a novel ablation modality that is currently used as an alternative energy source
for pulmonary vein isolation in atrial fibrillation to overcome the potential risks
and harm done by conventional thermal energy sources. Compared to radiofrequency energy
preclinical data, PFA showed similar lesion depth but lower lesion width and a low
vulnerability of the phrenic nerve to this type of energy.
6
,
7
In the MANIFEST-PF trial, only transient phrenic nerve palsy was noted, which resolved
within 1 day of the intervention.
8
We present a patient in whom a FAT originating on the posterolateral wall of the right
atrium could not be ablated safely with radiofrequency energy because of the proximity
to the phrenic nerve. In a second procedure the patient’s FAT was successfully ablated
using PFA. This is the first case report where pulsed field energy was used in this
indication, with good clinical outcome.
Case report
A 33-year-old patient had been complaining of palpitations for years. Tachycardia
was, however, never documented, and a previous electrophysiological study was negative.
Recently, a symptomatic episode of atrial fibrillation was documented, and the patient
was put on a treatment with bisoprolol and flecainide and scheduled for a pulmonary
vein isolation. The electrophysiology study and electroanatomic mapping (CARTO 3 System
version 7.2; Biosense Webster, Irvine, CA) were performed using a QDOT Biosense catheter
delivering high-power short-duration radiofrequency energy according to the CLOSE
protocol.
9
The pulmonary veins were successfully isolated. During the procedure, frequent runs
of atrial tachycardia were noted. Mapping of this tachycardia showed a FAT with earliest
activation on the posterolateral wall of the right atrium. Ablation of this focus
was attempted; however, only 2 cautious applications were delivered, since there was
clear capture of the phrenic nerve during pace mapping of the phrenic nerve in the
region of earliest activation (Figure 1). Antiarrhythmic drugs were continued.
Figure 1
Local activation time (LAT) map of the right atrium in a right posterolateral view
during tachycardia. The focus of the tachycardia is easily seen, with earliest activation
in the area that is marked in red. Black dots mark places where capture of the phrenic
nerve was obvious, which is in the same region as ablation should be attempted.
One month after the ablation procedure, the patient was seen for follow-up and appeared
to be in atrial tachycardia at 123 beats per minute being highly symptomatic despite
medical therapy. The p-wave morphology was suspect of an FAT originating in the lateral
right atrium. A redo ablation of the FAT was discussed, but this time we decided to
use pulsed field energy to overcome and minimize the risk of potential phrenic nerve
injury.
The procedure was performed using the same mapping system. A Biosense SmartTouch catheter
(Biosense Webster) was used, which was connected to a pulsed field energy generator
(Galaxy Centauri; Galaxy Medical, San Carlos, CA), enabling a focal contact force–guided
ablation using pulsed field energy. Isolation of the 4 pulmonary veins was first confirmed.
Voltage mapping of the right and left atrium did not show any regions of low potentials.
Burst pacing could easily induce an atrial tachycardia with a cycle length of 550
ms. Mapping of the tachycardia confirmed again an FAT in the right atrium with earliest
activation in a region where local phrenic nerve capture was obvious. A relatively
large region of early fragmented potentials was noted at the location of earliest
activation. Ablation was performed with QRS-triggered pulses at an energy output of
25 amperes in a monopolar and biphasic mode. A setting of 25 amperes is the highest
energy output setting available, which allows for safe and contiguous lesions.
8
Contact force of at least 5 grams was deemed sufficient to allow for adequate electrode-tissue
proximity, resulting in efficient energy delivery to the myocardial tissue with broad
and contiguous lesions.
10
The amount of irrigation was kept constant at 2 mL/min, since cooling of the catheter
tip is not required when delivering nonthermal pulsed field energy.
During application of the lesions, the phrenic nerve was often stimulated by the pulse
trains of ablation energy. There was frequent transient focal firing of the atrial
tachycardia, necessitating a relatively broad region of ablation before the tachycardia
eventually stopped firing. Some extra lesions were applied to consolidate the region
of fragmented potentials. A total number of 14 applications were delivered (Figure 2).
Figure 2
Local activation time (LAT) map of the same tachycardia during the second procedure,
showing the same focus of the atrial tachycardia. At the site of earliest activation
a fragmented signal was noted preceding the surface p-wave with 35 ms. The green dots
mark the 14 places where ablation with pulsed field energy was performed. Lesion tag
size was set at 6 mm, corresponding with lesion diameters of 7.2 ± 1.6 mm seen on
histologic examination with the same energy settings.
8
At the end of the procedure, phrenic nerve capture was confirmed at the ablated area
where the ablated atrial tissue was not entrainable. The tachycardia was noninducible
despite aggressive burst pacing at a cycle length of 220 ms in baseline and during
infusion of isoprenaline at 2 μg/min. The patient was discharged on the same day of
the intervention. Antiarrhythmic drugs were discontinued. During 3 months of follow-up,
she experienced no recurrence of palpitations and she remained in sinus rhythm confirmed
on 24-hour Holter recording.
Discussion
Pulsed field energy has emerged as a novel energy form for the ablation of atrial
fibrillation and will probably gain importance in the near future. Recent clinical
data seem to confirm the preclinical data, where less collateral damage is done, especially
with regard to nearby structures such as the esophagus or the phrenic nerves.
8
,
11
The use of pulsed field energy for the treatment of FAT, however, is still new, with
only 1 case report available describing ablation of a focus within the right atrial
appendage.
12
This is the first case report where the myocardial tissue–selective properties of
pulsed field ablation were directly applicable in a case of FAT with a focus close
to the right phrenic nerve. We were able to deliver pulsed field energy with the CE-labeled
Galaxy system through a conventional focal contact force catheter.
Our approach seems to be an efficient and safe solution that overcomes the known difficulties
of ablating foci within proximity of the phrenic nerve. Up until now, the only way
to ablate safely in this region was by protecting the phrenic nerve during the ablation
procedure, which has been done by balloon inflation in the epicardial space or by
creating an iatrogenic pneumothorax.2, 3, 4, 5 We believe that these kinds of complex
procedures will become obsolete once PFA with a focal catheter becomes more available.