9
views
0
recommends
+1 Recommend
1 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found

      Performance of cardiac cadmium-zinc-telluride gamma camera imaging in coronary artery disease: a review from the cardiovascular committee of the European Association of Nuclear Medicine (EANM)

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The trade-off between resolution and count sensitivity dominates the performance of standard gamma cameras and dictates the need for relatively high doses of radioactivity of the used radiopharmaceuticals in order to limit image acquisition duration. The introduction of cadmium-zinc-telluride (CZT)-based cameras may overcome some of the limitations against conventional gamma cameras. CZT cameras used for the evaluation of myocardial perfusion have been shown to have a higher count sensitivity compared to conventional single photon emission computed tomography (SPECT) techniques. CZT image quality is further improved by the development of a dedicated three-dimensional iterative reconstruction algorithm, based on maximum likelihood expectation maximization (MLEM), which corrects for the loss in spatial resolution due to line response function of the collimator. All these innovations significantly reduce imaging time and result in a lower patient's radiation exposure compared with standard SPECT. To guide current and possible future users of the CZT technique for myocardial perfusion imaging, the Cardiovascular Committee of the European Association of Nuclear Medicine, starting from the experience of its members, has decided to examine the current literature regarding procedures and clinical data on CZT cameras. The committee hereby aims 1) to identify the main acquisitions protocols; 2) to evaluate the diagnostic and prognostic value of CZT derived myocardial perfusion, and finally 3) to determine the impact of CZT on radiation exposure.

          Related collections

          Most cited references59

          • Record: found
          • Abstract: found
          • Article: not found

          A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera.

          This study described and validated a new solid-state single-photon gamma-camera and compared it with a conventional-SPECT Anger camera. The compact new camera uses a unique method for localizing gamma-photon information with a bank of 9 solid-state detector columns with tungsten collimators that rotate independently. Several phantom studies were performed comparing the new technology with conventional-SPECT technology. These included measurements of line sources and single- and dual-radionuclide studies of a torso phantom. Simulations were also performed using a cardiothoracic phantom. Furthermore, 18 patients were scanned with both the new camera and a conventional-SPECT camera. The new camera had a count sensitivity that was 10 times higher than that of the conventional camera and a compensated spatial resolution that was moderately better. Dual-radionuclide studies using a phantom show the further potential of the new camera for a 2-tracer simultaneous acquisition. Two-minute clinical studies with the new camera and 11-min studies with the conventional camera qualitatively showed good-to-excellent image quality and improved myocardial edge definition for the new camera. These initial performance characteristics of a new solid-state single-photon gamma-camera offer great promise for clinical dynamic SPECT protocols, with important implications for applications in nuclear cardiology and molecular imaging.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            EANM procedural guidelines for radionuclide myocardial perfusion imaging with SPECT and SPECT/CT: 2015 revision

            Since the publication of the European Association of Nuclear Medicine (EANM) procedural guidelines for radionuclide myocardial perfusion imaging (MPI) in 2005, many small and some larger steps of progress have been made, improving MPI procedures. In this paper, the major changes from the updated 2015 procedural guidelines are highlighted, focusing on the important changes related to new instrumentation with improved image information and the possibility to reduce radiation exposure, which is further discussed in relation to the recent developments of new International Commission on Radiological Protection (ICRP) models. Introduction of the selective coronary vasodilator regadenoson and the use of coronary CT-contrast agents for hybrid imaging with SPECT/CT angiography are other important areas for nuclear cardiology that were not included in the previous guidelines. A large number of minor changes have been described in more detail in the fully revised version available at the EANM home page: http://eanm.org/publications/guidelines/2015_07_EANM_FINAL_myocardial_perfusion_guideline.pdf.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Compared performance of high-sensitivity cameras dedicated to myocardial perfusion SPECT: a comprehensive analysis of phantom and human images.

              Differences in the performance of cadmium-zinc-telluride (CZT) cameras or collimation systems that have recently been commercialized for myocardial SPECT remain unclear. In the present study, the performance of 3 of these systems was compared by a comprehensive analysis of phantom and human SPECT images.
                Bookmark

                Author and article information

                Journal
                European Journal of Nuclear Medicine and Molecular Imaging
                Eur J Nucl Med Mol Imaging
                Springer Science and Business Media LLC
                1619-7070
                1619-7089
                December 2016
                August 19 2016
                December 2016
                : 43
                : 13
                : 2423-2432
                Article
                10.1007/s00259-016-3467-5
                27542010
                3936c614-9f03-4835-bdc8-bac6c7255365
                © 2016

                http://www.springer.com/tdm

                History

                Comments