Introduction
Bayés syndrome is currently defined as the finding of advanced interatrial block (a-IAB)
on 12-lead ECG associated with atrial fibrillation (AF) (1). Several studies have
demonstrated the high risk of developing atrial fibrillation for patients with a-IAB,
which has practical consequences of anticoagulant administration (2-4).
Patients with tetralogy of Fallot have a high risk of developing supraventricular
arrhythmias during their lifetime. The most common arrhythmias are represented by
an atrial flutter, intra-atrial reentry, and atrial fibrillation (5, 6). Minimal data
exist in the literature reporting the presence of a-IAB on the 12-lead ECG in patients
with corrected tetralogy of Fallot and atrial flutter or atrial fibrillation, and
even less about phenocopies of Bayés syndrome in these patients.
Case Report
A 56-year old male patient with past medical history of corrected tetralogy of Fallot
(an initial palliative intervention at the age of 6 years and a second, complete repair,
at the age of 12 years), percutaneous pulmonary valve implantation for severe pulmonary
regurgitation at the age of 55 years, and one episode of paroxysmal atrial fibrillation
who presented for palpitations and near-syncope. His medications included carvedilol
6.25 mg bid, eplerenone 25 mg od, perindopril 4 mg od, furosemide 20 mg od, Fluindione
20 mg od, and potassium supplements 600 mg tid. Clinical examination revealed a blood
pressure of 107/71 mm Hg, heart rate of 260 bpm, SaO2 of 97% breathing room air, and
a temperature of 37 °C. His ECG recorded at admission and 5 minutes later are presented
in Figures 1 and 2.
Figure 1
Wide QRS complex tachycardia with a heart rate of 260 bpm, right bundle branch block
pattern, and left axis deviation. The differential diagnoses are ventricular tachycardia,
supraventricular tachycardia with aberrancy, and pre-excitation syndrome
Figure 2
Twelve-lead ECG showing atrial flutter with 2:1 and 3:1 atrioventricular conduction,
heart rate of 100 bpm, left axis deviation, and right bundle branch block
His echocardiogram revealed a mildly dilated left ventricle (ESD=55 mm) with severe
LV systolic dysfunction, ejection fraction (EF) of 28% (Simpson), strain −8.6%, dilated
right ventricle (48 mm at the level of the tricuspid annulus) and right atrium, sPAP
of 33 mm Hg, and no pericardial effusion.
His biological workup showed no anemia, thyroid dysfunction, or electrolyte imbalance.
Considering his second ECG (Fig. 2), the wide QRS complex tachycardia was interpreted
as atrial flutter with 1:1 A:V conduction. The patient was treated initially with
IV amiodarone and oral carvedilol. An electrophysiological study was subsequently
performed with the CARTO electro-anatomical mapping system, which demonstrated typical
counterclockwise atrial flutter. Radiofrequency ablation of the cavotricuspid isthmus
was performed, and the patient converted to sinus rhythm. The ECG in sinus rhythm
at the end of the ablation procedure, recorded in the electrophysiology lab, is presented
in Figure 3.
Figure 3
Twelve-lead ECG showing sinus rhythm, “north-west” axis deviation, and right bundle
branch block. The P waves have an increased duration of 159 ms, with an initial short
and flat or positive component followed by a deep negative component in leads II,
III, and aVF, suggesting the presence of a-IAB
This ECG showed the presence of peculiar P waves, with an initial short and flat or
positive component followed by a deep negative component in leads II, III, and aVF
(best visible in lead II); positive in leads I and aVL, with a P-wave duration of
159 ms, suggesting the presence of a-IAB. However, on closer inspection, despite the
presence of positive P wave in leads I and aVL, one could see that the morphology
and the amplitude of the QRS complexes were different from the ECG in Figure 2, suggesting
ECG electrode inversion.
A 12-lead ECG was recorded with the electrodes placed in the correct manner, which
made the aspect of a-IAB disappear (Fig. 4). The patient was discharged 48 hours later.
Figure 4
Twelve-lead ECG recorded with correctly-placed ECG electrodes, revealing sinus rhythm
with a different morphology of the P wave and QRS complexes compared with Figure 3,
with positive P waves in leads II, III, and aVF; positive in leads I and aVL; left
axis deviation; right bundle branch block
Discussion
We present the case of a Bayés syndrome phenocopy explained by a 12-lead ECG recorded
with misplaced electrodes, masquerading as a-IAB post catheter ablation of atrial
flutter with 1:1 A:V conduction in a patient with corrected tetralogy of Fallot.
A correct 12-lead ECG recording requires standard placement of electrodes on the patient’s
limbs and chest, as follows: lead I is a bipolar recording of the electrical current
traveling from the positive electrode (left arm) to the negative electrode (right
arm). Lead II records the electrical activity between the left leg (positive electrode)
and the right arm (negative electrode), and lead III records the electrical activity
between the left leg (positive electrode) and the left arm (negative electrode). Leads
aVR, aVL, and aVF are “augmented” unipolar recordings from the right arm, left arm,
and left leg, respectively. The fourth limb electrode-the “indifferent electrode”-is
placed on the patient’s right leg. Any reversal between each of these two electrodes
will create a typical change in the morphology of the P wave, QRS, and T wave. Therefore,
reversing the two electrodes of lead I (right arm and left arm) will determine an
inversion of polarity of the P wave and QRS such that the P wave and QRS complex become
negative, a switch occurs between leads II and III, as well as between aVR and aVL,
with the lead aVF staying unchanged. A reversal of the electrodes of lead II (right
arm to left leg) will determine an inversion of the polarity of the P waves and QRS
in lead II, with a switch between leads I and III and also between leads aVR and aVF,
with lead aVL remaining unchanged. A reversal of the electrodes of lead III (left
arm to left leg) will determine an inversion of the polarity of the P waves and QRS
in lead III, a switch between leads I and II, as well as between aVL and aVF, with
lead aVR remaining unchanged. With electrode reversal of either right arm to right
leg, left arm to right leg, or bilateral arms to leg reversal, leads II, III, lead
I, respectively, become isoelectric, which is not the case in our patient. Notably,
the left leg to right leg reversal does not change the ECG aspect.
With these in mind, by examining Figures 3 and 4, one can observe that in Figure 3
lead I looks like lead aVF of Figure 4, lead II looks like the mirror image of lead
I of Figure 4, lead III is like the mirror image of lead II of Figure 4, aVR is like
lead aVL of Figure 4, aVL is like lead aVF of Figure 4, and lead aVF is like aVR of
Figure 4. However, no simple or combined limb lead electrode reversal is compatible
with such an ECG aspect. A closer look at the P wave morphology in V1 and V2 reveals
that the P wave has a slightly different morphology in Figure 4 compared with Figure
3, suggesting that the electrodes V1 and V2 have been misplaced too. The only misplacement
combination of electrodes giving such an aspect on ECG is a switch between the right
arm electrode and V1, placement of the V2 electrode on the left leg, placement of
the left leg electrode on the left arm and placement of the left arm electrode in
the position of V2 (Table 1), which is an unusual, albeit possible, misplacement pattern
of the ECG electrodes, as explained probably by the fact that the reversed electrodes
typically have the same color (right arm electrode and V1: red; left arm electrode
and V2: yellow). Notably, the correct placement of the limb electrodes leads to the
disappearance of the a-IAB pattern (Fig. 4).
Table 1
Explanation of the incorrect placement of the ECG electrodes responsible for the “Bayés
syndrome” phenocopy aspect in Figure 3
Electrode
Normal position
Modified position
Right arm (red color electrode)
Responsible for recording DI (with left arm electrode), DII (with left leg) and aVR
(unipolar recording)
V1 (red color electrode)
Left arm (yellow color electrode)
Responsible for recording DI (with right arm electrode), DIII (with left leg electrode)
and aVL (unipolar recording)
V2 (yellow color electrode)
Left leg
Responsible for recording DII (with left arm electrode), DIII (with right arm electrode)
and aVF (unipolar recording)
Left arm
Right leg
Also called the indifferent electrode
unchanged
V1
4th intercostal space, right parasternal
Right arm
V2 (yellow color electrode)
4th intercostal space, left parasternal
Left leg (yellow color electrode)
V3
Halfway between electrode V2 and V4
unchanged
V4
5th intercostal space, on the mid-clavicular line
unchanged
V5
5th intercostal space, on the anterior axillary line
unchanged
V6
5th intercostal space, on the mid-axillary line
unchanged
Even though incorrectly recorded, we consider the ECG in Figure 3 is valuable because
it mimicked Bayés syndrome and created a false aspect of a-IAB. Therefore, a Bayés
syndrome phenocopy can occur because of a not-so-rare technical error, that is, incorrect
limb electrode placement.
Conclusion
This case emphasizes the fact that a thorough inspection of the 12-lead ECG is warranted
when providing care to patients and is, in our knowledge, one of the first reported
cases of Bayés syndrome phenocopy.