      Clinical utility of Gafchromic film in an MRI-guided linear accelerator

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          The purpose of this study is to comprehensively evaluate the suitability of Gafchromic EBT3 and EBT-XD film for dosimetric quality assurance in 0.35 T MR-guided radiotherapy.

          Methods

          A 0.35 T magnetic field strength was utilized to evaluate magnetic field effects on EBT3 and EBT-XD Gafchromic films by studying the effect of film exposure time within the magnetic field using two timing sequences and film not exposed to MR, the effect of magnetic field exposure on the crystalline structure of the film, and the effect of orientation of the film with respect to the bore within the magnetic field. The orientation of the monomer crystal was qualitatively evaluated using scanning electron microscopy (SEM) compared to unirradiated film. Additionally, dosimetric impact was evaluated through measurements of a series of open field irradiations (0.83 × 0.83-cm 2 to 19.92 × 19.92-cm 2) and patient specific quality assurance measurements. Open fields were compared to planned dose and an independent dosimeter. Film dosimetry was applied to twenty conventional and twenty stereotactic body radiotherapy (SBRT) patient specific quality assurance cases.

          Results

          No visual changes in crystal orientation were observed in any evaluated SEM images nor were any optical density differences observed between films irradiated inside or outside the magnetic field for both EBT3 and EBT-XD film. At small field sizes, the average difference along dose profiles measured in film compared to the same points measured using an independent dosimeter and to predicted treatment planning system values was 1.23% and 1.56%, respectively. For large field sizes, the average differences were 1.91% and 1.21%, respectively. In open field tests, the average gamma pass rates were 99.8% and 97.2%, for 3%/3 mm and 3%/1 mm, respectively. The median (interquartile range) 3%/3 mm gamma pass rates in conventional QA cases were 98.4% (96.3 to 99.2%), and 3%/1 mm in SBRT QA cases were 95.8% (95.0 to 97.3%).

          Conclusions

          MR exposure at 0.35 T had negligible effects on EBT3 and EBT-XD Gafchromic film. Dosimetric film results were comparable to planned dose, ion chamber and diode measurements.

          Related collections

          Most cited references28

          • Record: found
          • Abstract: found
          • Article: not found

          The ViewRay system: magnetic resonance-guided and controlled radiotherapy.

          A description of the first commercially available magnetic resonance imaging (MRI)-guided radiation therapy (RT) system is provided. The system consists of a split 0.35-T MR scanner straddling 3 (60)Co heads mounted on a ring gantry, each head equipped with independent doubly focused multileaf collimators. The MR and RT systems share a common isocenter, enabling simultaneous and continuous MRI during RT delivery. An on-couch adaptive RT treatment-planning system and integrated MRI-guided RT control system allow for rapid adaptive planning and beam delivery control based on the visualization of soft tissues. Treatment of patients with this system commenced at Washington University in January 2014. Copyright © 2014 Elsevier Inc. All rights reserved.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams.

              Dose verification in highly conformal radiation therapy such as IMRT or proton therapy can benefit from the high spatial resolution offered by radio-chromic films such as Gafchromic EBT or EBT2. Recently, a new generation of these films, EBT3, has become available. The composition and thickness of the sensitive layer are the same as for the previous EBT2 films. The most important change is the symmetric layer configuration to eliminate side orientation dependence, which is reported for EBT2 films. The general film characteristics such as sensitivity to read-out orientation and postexposure darkening evolution of the new EBT3 film are evaluated. Film response has been investigated in clinical photon and proton beams and compared to former EBT2 films. Quenching effects in the proton Bragg peak region have been studied for both, EBT2 and EBT3 films. The general performance of EBT3 is comparable to EBT2, and the orientation dependence with respect to film side is completely eliminated in EBT3 films. Response differences of EBT2 and EBT3 films are of the same order of magnitude as batch-to-batch variations observed for EBT2 films. No significant difference has been found for both generations of EBT films between photon and proton exposure. Depth dose measurements of EBT2 and EBT3 show an excellent agreement, though underestimating dose by up to 20% in the Bragg peak region. The symmetric configuration of EBT3 presents a major improvement for film handling. EBT3 has similar dosimetric performance as its precursor EBT2 and can, thus, be applied to dose verification in IMRT in the same way. For dose verification in proton therapy the underresponse in the Bragg peak region has to be taken into account.
                Bookmark

                Author and article information

                Contributors
                Ilma.Xhaferllari@beaumont.org
                jkim8@hfhs.org
                ruchiradm@gmail.com
                krogothliu@gmail.com
                dudongsu@gmail.com
                adoemer1@hfhs.org
                ichetty1@hfhs.org
                nwen1@hfhs.org
                Journal
                Radiat Oncol
                Radiat Oncol
                Radiation Oncology (London, England)
                BioMed Central (London )
                1748-717X
                26 June 2021
                26 June 2021
                2021
                : 16
                : 117
                Affiliations
                [1 ]GRID grid.461921.9, ISNI 0000 0004 0460 1081, Department of Radiation Oncology, , Beaumont Health, ; Troy, MI USA
                [2 ]GRID grid.239864.2, ISNI 0000 0000 8523 7701, Department of Radiation Oncology, , Henry Ford Health System, ; 2799 West Grand Blvd, Detroit, MI USA
                [3 ]GRID grid.254444.7, ISNI 0000 0001 1456 7807, Department of Chemical Engineering and Material Science, , Wayne State University, ; Detroit, MI USA
                Author information
                http://orcid.org/0000-0001-9776-8056
                Article
                1844
                10.1186/s13014-021-01844-z
                8236160
                34174932
                54418e56-5e79-4a4c-a3e7-000497e58551
                © The Author(s) 2021

                Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

                History
                : 13 March 2021
                : 15 June 2021
                Funding
                Funded by: FundRef http://dx.doi.org/10.13039/100000048, American Cancer Society;
                Award ID: RSG-15-137-01-CCE
                Award Recipient :
                Categories
                Research
                Custom metadata
                © The Author(s) 2021

                Oncology & Radiotherapy
                mri-guided linac,gafchromic film,quality assurance
                Oncology & Radiotherapy
                mri-guided linac, gafchromic film, quality assurance

                Comments

                Comment on this article