A photometric flow titration based on the redox reaction between KMnO4 and minoxidil is described. The best titration results were observed at 3.20 x 10-4 mol L-1 KMnO4 and 1.00 x 10-3 mol L-1 minoxidil, using the minoxidil solutions as titrant. The flow rate was fixed at 17 mL min-1 and the titrant was added to the system in aliquots of 500 µL, the color changes were monitored at 550 nm. The method was applied to commercial samples and compared with the results from a chromatographic procedure. Recoveries from 97.6 to 102.8 % were observed depending on the sample. Comparison with the chromatographic procedure reveled relative errors of 3.5 - 4.0 %.
Um procedimento fotométrico em fluxo contínuo, baseado na reação de óxido-redução entre KMnO4 e minoxidil é proposto. Os melhores resultados na titulação foram obtidos usando concentrações de permanganato de 3,20x10-4 mol L-1 e minoxidil, usado como titulante, da ordem de 1,00x10-3 mol L-1. Vazão de 17 mL min-1 foi fixada para as leituras fotométricas contínuas e o titulante foi adicionado em alíquotas de 500 µL, enquanto as mudanças de cor foram monitoradas em 550 nm. Recuperações da ordem de 97,6 a 102,8% foram observadas, dependo da amostra analisada. O método foi aplicado a amostras comerciais contendo minoxidil e comparado com resultados obtidos a partir de procedimentos cromatográficos com erros relativos da ordem de 3,5 a 4,0 %.