      A Rabbit Model Study to Determine the Efficacy of a Prototype Corneal Endothelium Protector during Cataract Surgery

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose. We evaluated the efficacy and safety of a mechanical device, the P-chute, in corneal endothelium preservation during phacoemulsification in a rabbit model. Methods. Twenty-four rabbits were randomly assigned into 2 groups. One eye of each rabbit underwent phacoemulsification that simulated the removal of a dense nucleus, with or without the P-chute. Serial slit-lamp examinations, anterior segment optical coherence tomography (ASOCT) scans, and specular microscopy were performed. Three rabbits from each group were sacrificed on postoperative days (PODs) 1, 5, 7, and 14. Histological analysis of the corneas was performed. Results. There was a trend towards lesser endothelial cell loss for the P-chute group at POD1 (4.9% versus 12.5%, p = 0.53), POD5 (10.4% versus 12.2%, p = 0.77), and POD7 (10.5% versus 17.2%, p = 0.52). There was no significant difference in the corneal thickness ( p = >0.05) between the 2 groups. The insertion of the device was challenging. The use of the P-chute only added an extra 15% to the surgical time. Conclusions. There was a trend towards better endothelium preservation with the P-chute even though the results were not statistically significant. We believe that the device could be useful in certain surgical situations. Further work is needed to improve the device insertion.

          Related collections

          Most cited references17

          • Record: found
          • Abstract: found
          • Article: not found

          Effect of cataract surgery on the corneal endothelium: modern phacoemulsification compared with extracapsular cataract surgery.

          To investigate whether modern phacoemulsification surgery results in more damage to the corneal endothelium than extracapsular cataract extraction (ECCE), and to examine which preoperative, operative, and postoperative factors influence the effect of cataract surgery on the endothelium. Randomized controlled trial. Five hundred patients 40 years or older were randomized into 2 groups (ECCE, 249; phacoemulsification, 251). Central corneal endothelial cell counts, coefficient of variation of cell size, and hexagonality were assessed before surgery and up to 1 year postoperatively. Endothelial cell count. Four hundred thirty-three patients completed the trial. The initial preoperative mean cell count for the entire sample was 2481 (standard error [SE]: 18.6), reduced at 1 year postoperatively to 2239 (SE: 23.5). An average 10% reduction in cell count was recorded by 1 year postoperatively. There was no such change in hexagonality or in the coefficient of variation. There was no significant difference in overall percentage cell loss between the 2 treatment groups. Factors associated with excessive cell loss (> or =15% by 1 year) were a hard cataract (odds ratio [OR]: 2.1, 95% confidence limits: 1.1-4.1; P = 0.036), age (OR: 1.04, P = 0.005), and capsule or vitreous loss at surgery (OR: 2.38, P = 0.106). Phacoemulsification carried a significantly higher risk (OR: 3.7, P = 0.045) of severe cell loss in the 45 patients with hard cataracts relative to ECCE (52.6% vs. 23.1%; chi-square test, P = 0.041), with both procedures achieving similar postoperative visual acuity outcomes. No significant difference in overall corneal endothelial cell loss was found between these 2 operative techniques. The increased risk of severe cell loss with phacoemulsification in patients with hard cataracts suggests that phacoemulsification may not be the optimal procedure in these cases, and that ECCE should be preferred.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Regenerative capacity of the corneal endothelium in rabbit and cat.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Endothelial cell loss after phacoemulsification: relation to preoperative and intraoperative parameters.

              To evaluate the effect of the location of the corneoscleral tunnel incision as well as preoperative and intraoperative parameters on total and localized endothelial cell loss. Department of Ophthalmology, Humboldt-University of Berlin, Berlin, Germany. Fifty consecutive patients scheduled for routine cataract surgery were selected prospectively for this clinical trial. Preoperatively, the axial length, anterior chamber depth, lens thickness, and astigmatism were measured. Phacoemulsification time and relative energy as well as total surgical time were recorded. With a specular microscope, endothelial cell counts were determined centrally, superiorly, and temporally preoperatively and 6 weeks and 6 and 12 months postoperatively. After 12 months, the mean overall central endothelial cell loss in all eyes was 8.5%. The mean endothelial cell loss was 11.9% in the lateral quadrant and 11.4% in the superior quadrant. There were no significant differences between superior and temporal surgical approaches in intraoperative parameters of phacoemulsification time, relative intensity of phacoemulsification, and surgical time. There were no significant differences in central endothelial cell loss or in the area localized in the quadrant of the positions of the corneal surgical site. The only risk factors found significant for higher endothelial cell loss were shorter axial length and longer phacoemulsification time. The location of corneoscleral incisions for phacoemulsification can be chosen according to the preoperative astigmatism without inducing additional adverse effects on the corneal endothelium. Shorter eyes have a significantly higher risk for greater endothelial cell loss.
                Bookmark

                Author and article information

                Journal
                J Ophthalmol
                J Ophthalmol
                JOPH
                Journal of Ophthalmology
                Hindawi
                2090-004X
                2090-0058
                2017
                29 March 2017
                : 2017
                : 6906139
                Affiliations
                1Singapore National Eye Centre, Singapore
                2Singapore Eye Research Institute, Singapore
                3Yong Loo Lin School of Medicine, National University of Singapore, Singapore
                Author notes

                Academic Editor: Tamer A. Macky

                Author information
                http://orcid.org/0000-0002-6639-1249
                http://orcid.org/0000-0001-8602-9450
                http://orcid.org/0000-0002-5387-4209
                http://orcid.org/0000-0001-8922-6177
                http://orcid.org/0000-0002-1404-5096
                Article
                10.1155/2017/6906139
                5390626
                5622a0ef-a3b1-4c20-85d9-3b425415e942
                Copyright © 2017 Annabel C. Y. Chew et al.

                This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 6 September 2016
                : 16 November 2016
                : 23 November 2016
                Categories
                Research Article

                Ophthalmology & Optometry
                Ophthalmology & Optometry

                Comments

                Comment on this article