      The Theory and Fundamentals of Bioimpedance Analysis in Clinical Status Monitoring and Diagnosis of Diseases

      review-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Bioimpedance analysis is a noninvasive, low cost and a commonly used approach for body composition measurements and assessment of clinical condition. There are a variety of methods applied for interpretation of measured bioimpedance data and a wide range of utilizations of bioimpedance in body composition estimation and evaluation of clinical status. This paper reviews the main concepts of bioimpedance measurement techniques including the frequency based, the allocation based, bioimpedance vector analysis and the real time bioimpedance analysis systems. Commonly used prediction equations for body composition assessment and influence of anthropometric measurements, gender, ethnic groups, postures, measurements protocols and electrode artifacts in estimated values are also discussed. In addition, this paper also contributes to the deliberations of bioimpedance analysis assessment of abnormal loss in lean body mass and unbalanced shift in body fluids and to the summary of diagnostic usage in different kinds of conditions such as cardiac, pulmonary, renal, and neural and infection diseases.

          Related collections

          Most cited references137

          • Record: found
          • Abstract: found
          • Article: not found

          Bioelectrical phase angle and impedance vector analysis--clinical relevance and applicability of impedance parameters.

          The use of phase angle (PhA) and raw parameters of bioelectrical impedance analysis (BIA) has gained attention as alternative to conventional error-prone calculation of body composition in disease. This review investigates the clinical relevance and applicability of PhA and Bioelectrical Impedance Vector Analysis (BIVA) which uses the plot of resistance and reactance normalized per height. A comprehensive literature search was conducted using Medline identifying studies relevant to this review until March 2011. We included studies on the use of PhA or BIVA derived from tetrapolar BIA in out- and in-patient settings or institutionalized elderly. Numerous studies have proven the prognostic impact of PhA regarding mortality or postoperative complications in different clinical settings. BIVA has been shown to provide information about hydration and body cell mass and therefore allows assessment of patients in whom calculation of body composition fails due to altered hydration. Reference values exist for PhA and BIVA facilitating interpretation of data. PhA, a superior prognostic marker, should be considered as a screening tool for the identification of risk patients with impaired nutritional and functional status, BIVA is recommended for further nutritional assessment and monitoring, in particular when calculation of body composition is not feasible. Copyright © 2012 Elsevier Ltd and European Society for Clinical Nutrition and Metabolism. All rights reserved.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            A new method for monitoring body fluid variation by bioimpedance analysis: the RXc graph.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Body fluid volumes measurements by impedance: A review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods.

              This paper reviews various bioimpedance methods permitting to measure non-invasively, extracellular, intracellular and total body water (TBW) and compares BIA methods based on empirical equations of the wrist-ankle resistance or impedance at 50 kHz, height and weight with BIS methods which rely on an electrical model of tissues and resistances measured at zero and infinite frequencies. In order to compare these methods, impedance measurements were made with a multifrequency Xitron 4200 impedance meter on 57 healthy subjects which had undergone simultaneously a Dual X-ray absorptiometry examination (DXA), in order to estimate their TBW from their fat-free-mass. Extracellular (ECW) and TBW volumes were calculated for these subjects using the original BIS method and modifications of Matthie[Matthie JR. Second generation mixture theory equation for estimating intracellular water using bioimpedance spectroscopy. J Appl Physiol 2005;99:780-1], Jaffrin et al. [Jaffrin MY, Fenech M, Moreno MV, Kieffer R. Total body water measurement by a modification of the bioimpédance spectroscopy method. Med Bio Eng Comput 2006;44:873-82], Moissl et al. [Moissl UM, Wabel P, Chamney PW, Bosaeus I, Levin NW, et al. Body fluid volume determination via body composition spectroscopy in health and disease. Physiol Meas 2006;27:921-33] and their TBW resistivities were compared and discussed. ECW volumes were calculated by BIA methods of Sergi et al. [Sergi G, Bussolotto M, Perini P, Calliari I, et al. Accuracy of bioelectrical bioimpedance analysis for the assessment of extracellular space in healthy subjects and in fluid retention states. Ann Nutr Metab 1994;38(3):158-65] and Hannan et al. [Hannan WJ, Cowen SJ, Fearon KC, Plester CE, Falconer JS, Richardson RA. Evaluation of multi-frequency bio-impedance analysis for the assessment of extracellular and total body water in surgical patients. Clin Sci 1994;86:479-85] and TBW volumes by BIA methods of Kushner and Schoeller [Kushner RF, Schoeller DA. Estimation of total body water by bioelectrical impedance analysis. Am J Clin Nutr 1986;44(3):417-24], Lukaski et al. [Lukaski HC, Bolonchuk WW. Estimation of body fluid volumes using tetrapolar bioelectrical impedance measurements. Aviat Space Environ Med 1988;59:1163-9], Hannan et al. [Hannan WJ, Cowen SJ, Fearon KC, Plester CE, Falconer JS, Richardson RA. Evaluation of multi-frequency bio-impedance analysis for the assessment of extracellular and total body water in surgical patients. Clinical Science 1994;86:479-85], Deurenberg et al. [Deurenberg P, van der Koy K, Leenen R, Westrate JA, Seidell JC. Sex and age specific prediction formulas for estimating body composition from bioelectric impedance: a cross validation study. Int J Obesity 1991;15:17-25] These volumes were compared against those given by BIS method and, in the case of TBW, with those by DXA. For ECW, a good agreement was found between various BIS methods and that of Sergi while Hannan's values were higher. Both Matthie's and Moissl's methods gave mean TBW resistivities and volumes lower than those of Jaffrin's and DXA methods. Kushner et al. method gave values of TBW not significantly different from those of Jaffrin et al. and DXA, as Hannan's method in men, but Lukaski and Deurenberg methods led to an underestimation.
                Bookmark

                Author and article information

                Journal
                Sensors (Basel)
                Sensors (Basel)
                Sensors (Basel, Switzerland)
                MDPI
                1424-8220
                June 2014
                19 June 2014
                : 14
                : 6
                : 10895-10928
                Affiliations
                [1 ] Department of Biomedical Engineering, Faculty of Engineering, University of Malaya, 50603 Kuala Lumpur, Malaysia; E-Mails: samifathi@ 123456siswa.um.edu.my (S.F.K.); mas_dayana@ 123456um.edu.my (M.S.M.)
                [2 ] Centre for Innovation in Medical Engineering (CIME), Faculty of Engineering, University of Malaya, 50603 Kuala Lumpur, Malaysia
                [3 ] Department of Biomedical Engineering, College of Engineering, Sudan University of Science and Technology, 407, Khartoum, Sudan
                Author notes
                [* ] Author to whom correspondence should be addressed; E-Mail: fatimah@ 123456um.edu.my ; Tel.: +60-37-967-4486; Fax: +60-37-697-6878.
                Article
                sensors-14-10895
                10.3390/s140610895
                4118362
                24949644
                59f824dc-a1d7-448c-b696-dfa072c58142
                © 2014 by the authors; licensee MDPI, Basel, Switzerland.

                This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license ( http://creativecommons.org/licenses/by/3.0/).

                History
                : 08 April 2014
                : 03 June 2014
                : 04 June 2014
                Categories
                Review

                Biomedical engineering
                bioimpedance analysis,body composition,clinical status monitoring,diseases diagnostic and prediction

                Comments

                Comment on this article