      Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Accurate measurements of anterior chamber depth (ACD) and regional corneal thickness are especially important for the diagnosis and treatment of many ocular disease. This study aimed to evaluate the repeatability and reliability of a new swept source anterior segment optical coherence tomography (AS-OCT) and its agreement with Oculus Pentacam for measurements of ACD and corneal thickness (CT).

          Methods

          The central corneal thickness (CCT), superior corneal thickness (SCT), inferior corneal thickness (ICT), nasal corneal thickness (NCT), temporal corneal thickness (TCT) and ACD of the right eye from forty-nine adults aged 18 to 36 years (24.78±4.36 years old) were measured with Pentacam and AS-OCT (CASIA2). All subjects were measured twice with each device. The repeatability was determined using the coefficient of repeatability (COR), the relative COR and the limits of agreement (LOA). Bland-Altman plot was also used for evaluating the agreement between parameters from the two devices.

          Results

          For the repeatability of CASIA2, the COR of the two measurements of ACD, CCT, SCT, ICT, NCT and TCT was 0.31 mm, 18.58, 25.83, 28.32, 26.71 and 22.09 µm respectively. There were no statistically significant differences between the CT and ACD measurements by CASIA2 (P>0.05). For measurements with Pentacam and CASIA2, the COR of ACD, CCT, SCT, NCT, ICT and TCT was 0.294 mm, 13.10, 51.57, 48.06, 56.21 and 47.69 µm respectively. No significant differences were found between the values measured by CAISA2 and Pentacam for CT and ACD (P>0.05). The Bland–Altman analysis also suggested high consistency between measurements obtained by Pentacam and CASIA2.

          Conclusions

          Our results suggest that Pentacam and CASIA2 have good agreement in CT and ACD measurements. The two devices can be considered interchangeable for these parameters’ measurements in healthy subjects when monitoring corneal conditions or planning ocular surgery. However, subtle differences between CASIA2 and Pentacam should also be kept in mind for certain specific clinical or research purposes.

          Related collections

          Author and article information

          Journal
          Ann Transl Med
          Ann Transl Med
          ATM
          Annals of Translational Medicine
          AME Publishing Company
          2305-5839
          2305-5847
          July 2020
          July 2020
          : 8
          : 14
          : 857
          Affiliations
          [1 ]State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University , Guangzhou, China;
          [2 ]Department of Ophthalmology, Third Affiliated Hospital, Nanchang University , Nanchang, China;
          [3 ] The Fifth Affiliated Hospital of Sun Yat-sen University , Zhuhai, China
          Author notes

          Contributions: (I) Conception and design: G Jin, D Zheng; (II) Administrative support: G Jin, D Zheng; (III) Provision of study materials or patients: None; (IV) Collection and assembly of data: X Li, Y Zhou; (V) Data analysis and interpretation: G Jin, Y Zhou, A Chen; (VI) Manuscript writing: All authors; (VII) Final approval of manuscript: All authors.

          Correspondence to: Professor Danying Zheng. State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, China. Email: zhengdyy@ 123456163.com ; Guangming Jin, MD. State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, China. Email: jingm@ 123456mail2.sysu.edu.cn .
          Article
          PMC7396784 PMC7396784 7396784 atm-08-14-857
          10.21037/atm-20-187
          7396784
          32793701
          64303dbc-f74c-4d3b-9b5e-209046bdd795
          2020 Annals of Translational Medicine. All rights reserved.
          History
          : 04 January 2020
          : 25 June 2020
          Categories
          Original Article

          Oculus Pentacam,swept source anterior segment optical coherence tomography,Anterior chamber depth (ACD),corneal thickness

          Comments

          Comment on this article