      Evaluation of Different Dialyzers and the Impact of Predialysis Albumin Priming in Intermittent Hemodialysis With Reduced Anticoagulation

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction

          Systemic anticoagulation is widely used in routine clinical hemodialysis, but can be contraindicated in specific settings. Anticoagulant-free treatment regimens are prone to failure even in chronic intermittent hemodialysis. We quantified fiber blocking in settings of reduced anticoagulation to assess performance of different dialyzers and the potential benefit of albumin priming.

          Methods

          This crossover study included 10 patients performing 4 hours of hemodialysis at midweek in 7 different settings: that is, using Solacea 19H and FX800, both with regular and half dose of anticoagulation, Evodial 1.3 without systemic anticoagulation, and FX800 (half dose) and Evodial (no anticoagulation) when primed with a human albumin solution. Dialyzer fiber blocking was visualized in the dialyzer outlet potting using a 3-dimensional computed tomography (CT) scanning technique on micrometer resolution.

          Results

          No sessions had to be prematurely interrupted because of circuit clotting. The relative number of open fibers post dialysis was not influenced by the reduction of anticoagulation in the Solacea making this dialyzer superior in fiber patency in this setting above both the FX800 with reduced anticoagulation and the Evodial with no anticoagulation. Furthermore, no differences in relative number of open fibers were found in the FX800 and Evodial dialyzers with versus without albumin priming.

          Conclusion

          In situations in which reduced anticoagulation is indicated, the asymmetric triacetate ATA Solacea dialyzer outperforms a dialyzer with a conventional polysulfone membrane (FX800) or with the heparin-coated polyacrylonitrile membrane (Evodial). The use of human albumin to prime the dialysis circuit did not improve dialyzer patency.

          Graphical abstract

          Related collections

          Most cited references25

          • Record: found
          • Abstract: found
          • Article: not found

          High shear stress can initiate both platelet aggregation and shedding of procoagulant containing microparticles.

          Previous studies have demonstrated that a high level of shear stress can produce platelet aggregation without the addition of any agonist. We investigated whether high shear stress could cause both platelet aggregation and shedding of microparticles from the platelet plasma membrane. A coneplate viscometer was used to apply shear stress and microparticle formation was measured by flow cytometry. It was found that microparticle formation increased as the duration of shear stress increased. Both microparticles and the remnant platelets showed the exposure of procoagulant activity on their surfaces. Investigation of the mechanisms involved in shear-dependent microparticle generation showed that binding of von Willebrand factor (vWF) to platelet glycoprotein lb, influx of extracellular calcium, and activation of platelet calpain were required to generate microparticles under high shear stress conditions. Activation of protein kinase C (PKC) promoted shear-dependent microparticle formation. Epinephrine did not influence microparticle formation, although it enhanced platelet aggregation by high shear stress. These findings suggest the possibility that local generation of microparticles in atherosclerotic arteries, the site that pathologically high shear stress could occur, may contribute to arterial thrombosis by providing and expanding a catalytic surface for the coagulation cascade.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Albumin versus crystalloid for pump priming in cardiac surgery: Meta-analysis of controlled trials

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Results of the HepZero study comparing heparin-grafted membrane and standard care show that heparin-grafted dialyzer is safe and easy to use for heparin-free dialysis.

              Heparin is used to prevent clotting during hemodialysis, but heparin-free hemodialysis is sometimes needed to decrease the risk of bleeding. The HepZero study is a randomized, multicenter international controlled open-label trial comparing no-heparin hemodialysis strategies designed to assess non-inferiority of a heparin grafted dialyzer (NCT01318486). A total of 251 maintenance hemodialysis patients at increased risk of hemorrhage were randomly allocated for up to three heparin-free hemodialysis sessions using a heparin-grafted dialyzer or the center standard-of-care consisting of regular saline flushes or pre-dilution. The first heparin-free hemodialysis session was considered successful when there was neither complete occlusion of air traps or dialyzer, nor additional saline flushes, changes of dialyzer or bloodlines, or premature termination. The current standard-of-care resulted in high failure rates (50%). The success rate in the heparin-grafted membrane arm was significantly higher than in the control group (68.5% versus 50.4%), which was consistent for both standard-of-care modalities. The absolute difference between the heparin-grafted membrane and the controls was 18.2%, with a lower bound of the 90% confidence interval equal to plus 7.9%. The hypothesis of the non-inferiority at the minus 15% level was accepted, although superiority at the plus 15% level was not reached. Thus, use of a heparin-grafted membrane is a safe, helpful, and easy-to-use method for heparin-free hemodialysis in patients at increased risk of hemorrhage.
                Bookmark

                Author and article information

                Contributors
                Journal
                Kidney Int Rep
                Kidney Int Rep
                Kidney International Reports
                Elsevier
                2468-0249
                24 July 2019
                November 2019
                24 July 2019
                : 4
                : 11
                : 1538-1545
                Affiliations
                [1 ]Department of Nephrology, Ghent University Hospital, Gent, Belgium
                [2 ]Department of Cardiac Surgery, Ghent University Hospital, Gent, Belgium
                [3 ]Centre for X-ray Tomography, Physics and Astronomy, Ghent University, Belgium
                Author notes
                [] Correspondence: Sunny Eloot, Hemodialysis Unit 1P8, Ghent University Hospital, Corneel Heymanslaan 10, 9000 Gent, Belgium. sunny.eloot@ 123456ugent.be
                Article
                S2468-0249(19)31427-5
                10.1016/j.ekir.2019.07.010
                6933477
                31890995
                6572fec1-1812-4fa6-8fbf-f6fbb8c2bc68
                © 2019 International Society of Nephrology. Published by Elsevier Inc.

                This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

                History
                : 9 April 2019
                : 8 July 2019
                : 15 July 2019
                Categories
                Clinical Research

                albumin priming,clotting,coagulation,dialyzer,fiber patency,zero anticoagulation