      Application of volumetric modulated arc therapy (VMAT) in a dual-vendor environment

      research-article
      1 , , 1 , 1
      Radiation Oncology (London, England)
      BioMed Central

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background and Purpose

          The purpose of this study was to assess plan quality and treatment time achievable with the new VMAT optimization tool implemented in the treatment planning system Oncentra MasterPlan ® as compared to IMRT for Elekta SynergyS ® linear accelerators.

          Materials and methods

          VMAT was implemented on a SynergyS ® linear accelerator (Elekta Ltd., Crawley, UK) with Mosaiq ® record and verify system (IMPAC Medical Systems, Sunnyvale, CA) and the treatment planning system Oncentra MasterPlan ® (Nucletron BV, Veenendaal, the Netherlands). VMAT planning was conducted for three typical target types of prostate cancer, hypopharynx/larynx cancer and vertebral metastases, and compared to standard IMRT with respect to plan quality, number of monitor units (MU), and treatment time.

          Results

          For prostate cancer and vertebral metastases single arc VMAT led to similar plan quality as compared to IMRT. For treatment of the hypopharynx/larynx cancer, a second arc was necessary to achieve sufficient plan quality. Treatment time was reduced in all cases to 35% to 43% as compared to IMRT. Times required for optimization and dose calculation, however, increased by a factor of 5.0 to 6.8.

          Conclusion

          Similar or improved plan quality can be achieved with VMAT as compared to IMRT at reduced treatment times but increased calculation times.

          Related collections

          Most cited references33

          • Record: found
          • Abstract: found
          • Article: not found

          Volumetric modulated arc therapy: IMRT in a single gantry arc.

          In this work a novel plan optimization platform is presented where treatment is delivered efficiently and accurately in a single dynamically modulated arc. Improvements in patient care achieved through image-guided positioning and plan adaptation have resulted in an increase in overall treatment times. Intensity-modulated radiation therapy (IMRT) has also increased treatment time by requiring a larger number of beam directions, increased monitor units (MU), and, in the case of tomotherapy, a slice-by-slice delivery. In order to maintain a similar level of patient throughput it will be necessary to increase the efficiency of treatment delivery. The solution proposed here is a novel aperture-based algorithm for treatment plan optimization where dose is delivered during a single gantry arc of up to 360 deg. The technique is similar to tomotherapy in that a full 360 deg of beam directions are available for optimization but is fundamentally different in that the entire dose volume is delivered in a single source rotation. The new technique is referred to as volumetric modulated arc therapy (VMAT). Multileaf collimator (MLC) leaf motion and number of MU per degree of gantry rotation is restricted during the optimization so that gantry rotation speed, leaf translation speed, and dose rate maxima do not excessively limit the delivery efficiency. During planning, investigators model continuous gantry motion by a coarse sampling of static gantry positions and fluence maps or MLC aperture shapes. The technique presented here is unique in that gantry and MLC position sampling is progressively increased throughout the optimization. Using the full gantry range will theoretically provide increased flexibility in generating highly conformal treatment plans. In practice, the additional flexibility is somewhat negated by the additional constraints placed on the amount of MLC leaf motion between gantry samples. A series of studies are performed that characterize the relationship between gantry and MLC sampling, dose modeling accuracy, and optimization time. Results show that gantry angle and MLC sample spacing as low as 1 deg and 0.5 cm, respectively, is desirable for accurate dose modeling. It is also shown that reducing the sample spacing dramatically reduces the ability of the optimization to arrive at a solution. The competing benefits of having small and large sample spacing are mutually realized using the progressive sampling technique described here. Preliminary results show that plans generated with VMAT optimization exhibit dose distributions equivalent or superior to static gantry IMRT. Timing studies have shown that the VMAT technique is well suited for on-line verification and adaptation with delivery times that are reduced to approximately 1.5-3 min for a 200 cGy fraction.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Volumetric modulated arc therapy for delivery of prostate radiotherapy: comparison with intensity-modulated radiotherapy and three-dimensional conformal radiotherapy.

            Volumetric modulated arc therapy (VMAT) is a novel form of intensity-modulated radiotherapy (IMRT) optimization that allows the radiation dose to be delivered in a single gantry rotation of up to 360 degrees , using either a constant dose rate (cdr-VMAT) or variable dose rate (vdr-VMAT) during rotation. The goal of this study was to compare VMAT prostate RT plans with three-dimensional conformal RT (3D-CRT) and IMRT plans. The 3D-CRT, five-field IMRT, cdr-VMAT, and vdr-VMAT RT plans were created for 10 computed tomography data sets from patients undergoing RT for prostate cancer. The parameters evaluated included the doses to organs at risk, equivalent uniform doses, dose homogeneity and conformality, and monitor units required for delivery of a 2-Gy fraction. The IMRT and both VMAT techniques resulted in lower doses to normal critical structures than 3D-CRT plans for nearly all dosimetric endpoints analyzed. The lowest doses to organs at risk and most favorable equivalent uniform doses were achieved with vdr-VMAT, which was significantly better than IMRT for the rectal and femoral head dosimetric endpoints (p < 0.05) and significantly better than cdr-VMAT for most bladder and rectal endpoints (p < 0.05). The vdr-VMAT and cdr-VMAT plans required fewer monitor units than did the IMRT plans (relative reduction of 42% and 38%, respectively; p = 0.005) but more than for the 3D-CRT plans (p = 0.005). The IMRT and VMAT techniques achieved highly conformal treatment plans. The vdr-VMAT technique resulted in more favorable dose distributions than the IMRT or cdr-VMAT techniques, and reduced the monitor units required compared with IMRT.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Feasibility of single-isocenter volumetric modulated arc radiosurgery for treatment of multiple brain metastases.

              To evaluate the relative plan quality of single-isocenter vs. multi-isocenter volumetric modulated arc therapy (VMAT) for radiosurgical treatment of multiple central nervous system metastases. VMAT plans were created using RapidArc technology for treatment of simulated patients with three brain metastases. The plans consisted of single-arc/single-isocenter, triple-arc (noncoplanar)/single-isocenter, and triple-arc (coplanar)/triple-isocenter configurations. All VMAT plans were normalized to deliver 100% of the 20-Gy prescription dose to all lesions. The plans were evaluated by calculation of Paddick and Radiation Therapy Oncology Group conformity index scores, Paddick gradient index scores, and 12-Gy isodose volumes. All plans were judged clinically acceptable, but differences were observed in the dosimetric parameters, with the use of multiple noncoplanar arcs showing small improvements in the conformity indexes compared with the single-arc/single-isocenter and triple-arc (coplanar)/triple-isocenter plans. Multiple arc plans (triple-arc [noncoplanar]/single-isocenter and triple-arc [coplanar]/triple-isocenter) showed smaller 12-Gy isodose volumes in scenarios involving three metastases spaced closely together, with only small differences noted among all plans involving lesions spaced further apart. Our initial results suggest that single-isocenter VMAT plans can be used to deliver conformity equivalent to that of multiple isocenter VMAT techniques. For targets that are closely spaced, multiple noncoplanar single-isocenter arcs might be required. VMAT radiosurgery for multiple targets using a single isocenter can be efficiently delivered, requiring less than one-half the beam time required for multiple isocenter set ups. VMAT radiosurgery will likely replace multi-isocenter techniques for linear accelerator-based treatment of multiple targets.
                Bookmark

                Author and article information

                Journal
                Radiat Oncol
                Radiation Oncology (London, England)
                BioMed Central
                1748-717X
                2010
                25 October 2010
                : 5
                : 95
                Affiliations
                [1 ]Department of Radiotherapy, Regensburg University Medical Center, D-93042 Regensburg, Germany
                Article
                1748-717X-5-95
                10.1186/1748-717X-5-95
                2987940
                20973977
                688fe345-2d80-4868-92f7-348b2ddc9685
                Copyright ©2010 Dobler et al; licensee BioMed Central Ltd.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 12 August 2010
                : 25 October 2010
                Categories
                Research

                Oncology & Radiotherapy
                Oncology & Radiotherapy

                Comments

                Comment on this article