      Proficiency testing in immunohistochemistry—experiences from Nordic Immunohistochemical Quality Control (NordiQC)

      review-article
      ,
      Virchows Archiv
      Springer Berlin Heidelberg
      Immunohistochemistry, NordiQC, External quality assurance

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Despite extensive use of immunohistochemistry (IHC) for decades, lack of standardization remains a major problem, even aggravated in the era of targeted therapy. Nordic Immunohistochemical Quality Control (NordiQC) is an international academic proficiency testing (PT) program established in 2003 primarily aimed at assessing the analytical phases of the laboratory IHC quality. About 700 laboratories from 80 countries are currently participating. More than 30,000 IHC slides have been evaluated during 2003–2015. Overall, about 20 % of the staining results in the breast cancer IHC module and about 30 % in the general module have been assessed as insufficient for diagnostic use. The most common causes for insufficient results are less successful antibodies (poor and less robust antibodies, poorly calibrated ready-to-use (RTU) products, and stainer platform-dependent antibodies; 17 %), insufficiently calibrated antibody dilutions (20 %), insufficient or erroneous epitope retrieval (27 %), less sensitive visualization systems (19 %), and other (heat- and proteolysis-induced impaired morphology, endogenous biotin reaction, drying out phenomena, stainer platform-dependant protocol issues; 17 %). Approximately, 90 % of the insufficient results are characterized by either a too weak or false negative staining, whereas in the remaining 10 %, a poor signal-to-noise ratio or false positive staining is seen. Individually tailored recommendations for protocol optimization and identification of best tissue controls to ensure appropriate calibration of the IHC assay have for many markers improved IHC staining as well as inter-laboratory consistency of the IHC results. RTUs will not always provide an optimal result and data sheets frequently misguide the laboratories hampering the improvement in IHC quality. The overall data generated by NordiQC during 12 years indicates that continuous PT is valuable and necessary. Detailed description of the results of the NordiQC programme is available on www.nordiqc.org and summarized in this paper.

          Related collections

          Most cited references3

          • Record: found
          • Abstract: found
          • Article: not found

          Recommendations for improved standardization of immunohistochemistry.

          Immunohistochemistry (IHC) continues to suffer from variable consistency, poor reproducibility, quality assurance disparities, and the lack of standardization resulting in poor concordance, validation, and verification. This document lists the recommendations made by the Ad-Hoc Committee on Immunohistochemistry Standardization to address these deficiencies. Contributing factors were established to be underfixation and irregular fixation, use of nonformalin fixatives and ancillary fixation procedures divested from a deep and full understanding of the IHC assay parameters, minimal or absent IHC assay optimization and validation procedures, and lack of a standard system of interpretation and reporting. Definitions and detailed guidelines pertaining to these areas are provided.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            The total test approach to standardization of immunohistochemistry.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Demonstration of CDX2 is highly antibody dependant.

              CDX2 is a widely used immunohistochemical marker for intestinal differentiation in neoplasms. In the Nordic Immunohistochemical Quality Control external quality assessment scheme, only 45% of the laboratories participating in the CDX2 challenge in 2009 produced sufficient staining. A major cause of insufficient staining results appeared to be less successful primary antibody (Ab) clones. To evaluate the Ab performance in a standardized way, a comparative study was carried out.
                Bookmark

                Author and article information

                Contributors
                +45-2712-5248 , mv@rn.dk
                Journal
                Virchows Arch
                Virchows Arch
                Virchows Archiv
                Springer Berlin Heidelberg (Berlin/Heidelberg )
                0945-6317
                1432-2307
                26 August 2015
                26 August 2015
                2016
                : 468
                : 19-29
                Affiliations
                [ ]NordiQC, Institute of Pathology, Aalborg University Hospital, Aalborg, Denmark
                [ ]Department of Clinical Medicine, Aalborg University, Aalborg, Denmark
                Article
                1829
                10.1007/s00428-015-1829-1
                4751198
                26306713
                71491b86-f314-49ef-9735-9f8dc23f7b49
                © The Author(s) 2015

                Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

                History
                : 10 August 2015
                : 11 August 2015
                Categories
                Annual Review Issue
                Custom metadata
                © Springer-Verlag Berlin Heidelberg 2016

                Pathology
                immunohistochemistry,nordiqc,external quality assurance
                Pathology
                immunohistochemistry, nordiqc, external quality assurance

                Comments

                Comment on this article