      Patient-prosthesis mismatch in patients treated with transcatheter aortic valve implantation – predictors, incidence and impact on clinical efficacy. A preliminary study

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction

          Patient-prosthesis mismatch (PPM) is relatively frequent after surgical aortic valve replacement (SAVR) and negatively impacts prognosis.

          Aim

          We sought to determine the frequency and clinical effects of PPM after transcatheter aortic valve implantation (TAVI).

          Material and methods

          Overall, 238 patients who underwent TAVI were screened. Moderate PPM was defined as indexed effective orifice area (EOAi) between 0.65 and 0.85 cm 2/m 2, and severe PPM as < 0.65 cm 2/m 2. All-cause mortality and the Valve Academic Research Consortium 2 (VARC-2) defined composite of clinical efficacy at 1 year were the primary endpoints.

          Results

          Finally, 201 patients were included (mean age: 79.6 ±7.4 years, 52% females). The femoral artery served as the delivery route in 79% and most of the prostheses were self-expanding (68%). Any PPM was present in 48 (24%) subjects, and only 7 (3.5%) had severe PPM. Body surface area (BSA) independently predicted any PPM (OR = 16.9, p < 0.001) whereas post-dilation tended to protect against PPM (OR = 0.46, p = 0.09). All-cause mortality was similar in patients with moderate or severe PPM as compared to those with no PPM (14.6% vs. 14.3% vs. 13.1%, respectively, log-rank p = 0.98). Neither moderate (OR = 1.6, 95% CI: 0.8–3.2, p = 0.16) nor severe PPM (OR = 1.67, 95% CI: 0.36–7.7, p = 0.51) had a significant impact on composite endpoint, or its elements, with the exception of transvalvular pressure gradient > 20 mm Hg.

          Conclusions

          Severe PPM after TAVI is rare, can be predicted by larger BSA and does not seem to affect mid-term mortality or composite clinical outcome. Larger studies are needed to find different independent predictors of PPM and elucidate its impact in terms of device durability and long-term clinical efficacy.

          Related collections

          Most cited references22

          • Record: found
          • Abstract: found
          • Article: not found

          Two-year outcomes after transcatheter or surgical aortic-valve replacement.

          The Placement of Aortic Transcatheter Valves (PARTNER) trial showed that among high-risk patients with aortic stenosis, the 1-year survival rates are similar with transcatheter aortic-valve replacement (TAVR) and surgical replacement. However, longer-term follow-up is necessary to determine whether TAVR has prolonged benefits. At 25 centers, we randomly assigned 699 high-risk patients with severe aortic stenosis to undergo either surgical aortic-valve replacement or TAVR. All patients were followed for at least 2 years, with assessment of clinical outcomes and echocardiographic evaluation. The rates of death from any cause were similar in the TAVR and surgery groups (hazard ratio with TAVR, 0.90; 95% confidence interval [CI], 0.71 to 1.15; P=0.41) and at 2 years (Kaplan-Meier analysis) were 33.9% in the TAVR group and 35.0% in the surgery group (P=0.78). The frequency of all strokes during follow-up did not differ significantly between the two groups (hazard ratio, 1.22; 95% CI, 0.67 to 2.23; P=0.52). At 30 days, strokes were more frequent with TAVR than with surgical replacement (4.6% vs. 2.4%, P=0.12); subsequently, there were 8 additional strokes in the TAVR group and 12 in the surgery group. Improvement in valve areas was similar with TAVR and surgical replacement and was maintained for 2 years. Paravalvular regurgitation was more frequent after TAVR (P<0.001), and even mild paravalvular regurgitation was associated with increased late mortality (P<0.001). A 2-year follow-up of patients in the PARTNER trial supports TAVR as an alternative to surgery in high-risk patients. The two treatments were similar with respect to mortality, reduction in symptoms, and improved valve hemodynamics, but paravalvular regurgitation was more frequent after TAVR and was associated with increased late mortality. (Funded by Edwards Lifesciences; ClinicalTrials.gov number, NCT00530894.).
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Prosthesis-patient mismatch: definition, clinical impact, and prevention.

            Prosthesis-patient mismatch (PPM) is present when the effective orifice area of the inserted prosthetic valve is too small in relation to body size. Its main haemodynamic consequence is to generate higher than expected gradients through normally functioning prosthetic valves. This review updates the present knowledge about the impact of PPM on clinical outcomes. PPM is common (20-70% of aortic valve replacements) and has been shown to be associated with worse haemodynamic function, less regression of left ventricular hypertrophy, more cardiac events, and lower survival. Moreover, as opposed to most other risk factors, PPM can largely be prevented by using a prospective strategy at the time of operation.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Incidence and sequelae of prosthesis-patient mismatch in transcatheter versus surgical valve replacement in high-risk patients with severe aortic stenosis: a PARTNER trial cohort--a analysis.

              Little is known about the incidence of prosthesis-patient mismatch (PPM) and its impact on outcomes after transcatheter aortic valve replacement (TAVR).
                Bookmark

                Author and article information

                Journal
                Postepy Kardiol Interwencyjnej
                Postepy Kardiol Interwencyjnej
                PWKI
                Postępy w Kardiologii Interwencyjnej = Advances in Interventional Cardiology
                Termedia Publishing House
                1734-9338
                1897-4295
                29 November 2017
                2017
                : 13
                : 4
                : 281-287
                Affiliations
                [1 ]First Chair and Department of Cardiology, Medical University of Warsaw, Warsaw, Poland
                [2 ]Department of Cardiac Surgery, Medical University of Warsaw, Warsaw, Poland
                Author notes
                [*]

                Authors equally contributed to the writing of the manuscript

                Corresponding author: Piotr Scisło MD, PhD, First Department of Cardiology, Medical University of Warsaw, 1a Banacha St, 02-097 Warsaw, Poland. phone: +48 22 599 19 58. fax: +48 22 599 19 57. e-mail: scislo@ 123456wum.edu.pl
                Article
                31039
                10.5114/aic.2017.71608
                5770857
                75fd8e10-0420-4687-9224-ada494f5ee74
                Copyright: © 2017 Termedia Sp. z o. o.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License, allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.

                History
                : 16 October 2017
                : 17 November 2017
                Categories
                Original Paper

                transcatheter aortic valve implantation,patient-prosthesis mismatch,effective orifice area

                Comments

                Comment on this article