      A No-History Multi-Formula Approach to Improve the IOL Power Calculation after Laser Refractive Surgery: Preliminary Results

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          This retrospective comparative study proposes a multi-formula approach by comparing no-history IOL power calculation methods after myopic laser-refractive-surgery (LRS). One-hundred-thirty-two eyes of 132 patients who had myopic-LRS and cataract surgery were examined. ALMA, Barrett True-K (TK), Ferrara, Jin, Kim, Latkany and Shammas methods were evaluated in order to back-calculate refractive prediction error (PE). To eliminate any systematic error, constant optimization through zeroing-out the mean error (ME) was performed for each formula. Median absolute error (MedAE) and percentage of eyes within ±0.50 and ±1.00 diopters (D) of PE were analyzed. PEs were plotted with corresponding mean keratometry (K), axial length (AL), and AL/K ratio; then, different ranges were evaluated. With optimized constants through zeroing-out ME (90 eyes), ALMA was better when K ≤ 38.00 D-AL > 28.00 mm and when 38.00 D < K ≤ 40.00 D-26.50 mm < AL ≤ 29.50 mm; Barrett-TK was better when K ≤ 38.00 D-AL ≤ 26.50 mm and when K > 40.00 D-AL ≤ 28.00 mm or AL > 29.50 mm; and both ALMA and Barrett-TK were better in other ranges. (p < 0.05) Without modified constants (132 eyes), ALMA was better when K > 38.00 D-AL ≤ 29.50 mm and when 36.00 < K ≤ 38.00 D-AL ≤ 26.50 mm; Barrett-TK was better when K ≤ 36.00 D and when K ≤ 38.00 D with AL > 29.50 mm; and both ALMA and Barrett-TK were better in other ranges (p < 0.05). A multi-formula approach, according to different ranges of K and AL, could improve refractive outcomes in post-myopic-LRS eyes.

          Related collections

          Most cited references46

          • Record: found
          • Abstract: not found
          • Article: not found

          Development of the SRK/T intraocular lens implant power calculation formula

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Protocols for studies of intraocular lens formula accuracy.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry.

              To assess how intraocular lens (IOL) formula choice affects refractive outcomes after cataract surgery using IOLMaster biometry. Department of Ophthalmology, Gloucestershire Hospitals NHS Foundation Trust, Cheltenham, United Kingdom. Database study. Hypothetical prediction errors were retrospectively calculated on prospectively collected data from electronic medical records using optimized Hoffer Q, Holladay 1, and SRK/T formulas (Sofport AO and Akreos Fit IOLs) across a range of 0.5 mm or 1.0 mm axial length (AL) subgroups. In short eyes, the Hoffer Q had the lowest mean absolute error (MAE) for ALs from 20.00 to 20.99 mm. The Hoffer Q and Holladay 1 had a lower MAE than the SRK/T for ALs from 21.00 to 21.49 mm. There were no statistically significant differences in MAE for ALs from 21.50 to 21.99 mm. In medium eyes, there were no statistically significant differences in MAE for any IOL formula for ALs from 22.00 to 23.49 mm. For ALs from 23.50 to 25.99 mm, there was a trend toward lower MAEs for the Holladay 1, with statistically significant differences in 2 subgroups. In long eyes, the SRK/T had the lowest MAE, with statistically significant differences for ALs of 27.00 mm or longer. The Hoffer Q performed best for ALs from 20.00 to 20.99 mm, the Hoffer Q and Holladay 1 for ALs from 21.00 to 21.49 mm, and the SRK/T for ALs of 27.00 mm or longer. Using optimized constants, refractive outcomes of 40%, 75%, and 95% within ±0.25 diopter (D), ±0.50 D, and ±1.00 D, respectively, were achievable. No author has a financial or proprietary interest in any material or method mentioned. Additional disclosure is found in the footnotes. Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
                Bookmark

                Author and article information

                Contributors
                (View ORCID Profile)
                (View ORCID Profile)
                (View ORCID Profile)
                (View ORCID Profile)
                (View ORCID Profile)
                (View ORCID Profile)
                Journal
                JCMOHK
                Journal of Clinical Medicine
                JCM
                MDPI AG
                2077-0383
                April 2023
                April 15 2023
                : 12
                : 8
                : 2890
                Article
                10.3390/jcm12082890
                10144756
                37109228
                a2059aa4-a5d6-4a5c-b668-4da5c08d5420
                © 2023

                https://creativecommons.org/licenses/by/4.0/

                History

                Comments

                Comment on this article