      Whole shaft visibility and mechanical performance for active MR catheters using copper-nitinol braided polymer tubes

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Catheter visualization and tracking remains a challenge in interventional MR.

          Active guidewires can be made conspicuous in "profile" along their whole shaft exploiting metallic core wire and hypotube components that are intrinsic to their mechanical performance. Polymer-based catheters, on the other hand, offer no conductive medium to carry radio frequency waves. We developed a new "active" catheter design for interventional MR with mechanical performance resembling braided X-ray devices. Our 75 cm long hybrid catheter shaft incorporates a wire lattice in a polymer matrix, and contains three distal loop coils in a flexible and torquable 7Fr device. We explored the impact of braid material designs on radiofrequency and mechanical performance.

          Results

          The incorporation of copper wire into in a superelastic nitinol braided loopless antenna allowed good visualization of the whole shaft (70 cm) in vitro and in vivo in swine during real-time MR with 1.5 T scanner. Additional distal tip coils enhanced tip visibility. Increasing the copper:nitinol ratio in braiding configurations improved flexibility at the expense of torquability. We found a 16-wire braid of 1:1 copper:nitinol to have the optimum balance of mechanical (trackability, flexibility, torquability) and antenna (signal attenuation) properties. With this configuration, the temperature increase remained less than 2°C during real-time MR within 10 cm horizontal from the isocenter. The design was conspicuous in vitro and in vivo.

          Conclusion

          We have engineered a new loopless antenna configuration that imparts interventional MR catheters with satisfactory mechanical and imaging characteristics. This compact loopless antenna design can be generalized to visualize the whole shaft of any general-purpose polymer catheter to perform safe interventional procedures.

          Related collections

          Most cited references33

          • Record: found
          • Abstract: found
          • Article: not found

          MR-guided intravascular interventions: techniques and applications.

          Magnetic resonance imaging (MRI) offers several advantages over other imaging modalities that make it an attractive imaging tool for diagnostic and therapeutic procedures. This tremendous potential of MRI has provided the rationale for increased attention toward MR-guided endovascular interventions. MR guidance has been used recently to navigate endovascular catheters and deliver stents, vena cava filters, embolization materials, and septum closure devices. However, its potential goes beyond just copying existing procedures toward the development of new minimally invasive techniques that cannot be performed with conventional guiding techniques. Because of technical limitations and safety issues associated with some of the currently available devices, a limited number of clinical studies have been performed so far. The overall success for this developing field requires considerable interdisciplinary research within both the interventional and the MR community. Only through a combined effort can this complex technology find its way into clinical practice. This review discusses the hardware and software improvements that have helped to advance endovascular interventions under MR imaging guidance from a pure research tool to become a clinical reality. In addition, technical and safety issues specific to endovascular MR image guidance will be described and practical applications will be shown that take advantage of the benefits of MR for endovascular interventions. (Copyright) 2008 Wiley-Liss, Inc.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            RF heating due to conductive wires during MRI depends on the phase distribution of the transmit field.

            In many studies concerning wire heating during MR imaging, a "resonant wire length" that maximizes RF heating is determined. This may lead to the nonintuitive conclusion that adding more wire, so as to avoid this resonant length, will actually improve heating safety. Through a theoretical analysis using the method of moments, we show that this behavior depends on the phase distribution of the RF transmit field. If the RF transmit field has linear phase, with slope equal to the real part of the wavenumber in the tissue, long wires always heat more than short wires. In order to characterize the intrinsic safety of a device without reference to a specific body coil design, this maximum-tip heating phase distribution must be considered. Finally, adjusting the phase distribution of the electric field generated by an RF transmit coil may lead to an "implant-friendly" coil design. Copyright 2002 Wiley-Liss, Inc.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Interventional magnetic resonance angiography with no strings attached: wireless active catheter visualization.

              Active instrument visualization strategies for interventional MR angiography (MRA) require vascular instruments to be equipped with some type of radiofrequency (RF) coil or dipole RF antenna for MR signal detection. Such visualization strategies traditionally necessitate a connection to the scanner with either coaxial cable or laser fibers. In order to eliminate any wire connection, RF resonators that inductively couple their signal to MR surface coils were implemented into catheters to enable wireless active instrument visualization. Instrument background to contrast-to-noise ratio was systematically investigated as a function of the excitation flip angle. Signal coupling between the catheter RF coil and surface RF coils was evaluated qualitatively and quantitatively as a function of the catheter position and orientation with regard to the static magnetic field B0 and to the surface coils. In vivo evaluation of the instruments was performed in interventional MRA procedures on five pigs under MR guidance. Cartesian and projection reconstruction TrueFISP imaging enabled simultaneous visualization of the instruments and vascular morphology in real time. The implementation of RF resonators enabled robust visualization of the catheter curvature to the very tip. Additionally, the active visualization strategy does not require any wire connection to the scanner and thus does not hamper the interventionalist during the course of an intervention.
                Bookmark

                Author and article information

                Journal
                J Cardiovasc Magn Reson
                Journal of Cardiovascular Magnetic Resonance
                BioMed Central
                1097-6647
                1532-429X
                2009
                12 August 2009
                : 11
                : 1
                : 29
                Affiliations
                [1 ]Translational Medicine Branch, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, MD, USA
                [2 ]Institute of Biomedical Engineering, Bogazici University, Istanbul, Turkey
                Article
                1532-429X-11-29
                10.1186/1532-429X-11-29
                2743675
                19674464
                a4beb3f1-e877-4f12-9e69-c03f60c98e49
                Copyright © 2009 Kocaturk et al; licensee BioMed Central Ltd.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 16 March 2009
                : 12 August 2009
                Categories
                Research

                Cardiovascular Medicine
                Cardiovascular Medicine

                Comments

                Comment on this article