      Haemolysis as a first sign of thromboembolic event and acute pump thrombosis in patients with the continuous-flow left ventricular assist device HeartMate II

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Despite advances in pump technology, thromboembolic events/acute pump thrombosis remain potentially life-threatening complications in patients with continuous-flow left ventricular assist devices (CF-LVAD). We sought to determine early signs of thromboembolic event/pump thrombosis in patients with CF-LVAD, which could lead to earlier intervention.

          Methods

          We analysed all HeartMate II recipients ( n = 40) in our centre between December 2006 and July 2013. Thromboembolic event/pump thrombosis was defined as a transient ischaemic attack (TIA), ischaemic cerebrovascular accident (CVA), or pump thrombosis.

          Results

          During median LVAD support of 336 days [IQR: 182–808], 8 (20 %) patients developed a thromboembolic event/pump thrombosis (six TIA/CVA, two pump thromboses). At the time of the thromboembolic event/pump thrombosis, significantly higher pump power was seen compared with the no-thrombosis group (8.2 ± 3.0 vs. 6.4 ± 1.4 W, p = 0.02), as well as a trend towards a lower pulse index (4.1 ± 1.5 vs. 5.0 ± 1.0, p = 0.05) and a trend towards higher pump flow (5.7 ± 1.0 vs. 4.9 ± 1.9 L m, p = 0.06).

          The thrombosis group had a more than fourfold higher lactate dehydrogenase (LDH) median 1548 [IQR: 754–2379] vs. 363 [IQR: 325–443] U/L, p = 0.0001). Bacterial ( n = 4) or viral ( n = 1) infection was present in 5 out of 8 patients. LDH > 735 U/L predicted thromboembolic events/pump thrombosis with a positive predictive value of 88 %.

          Conclusions

          In patients with a CF-LVAD (HeartMate II), thromboembolic events and/or pump thrombosis are associated with symptoms and signs of acute haemolysis as manifested by a high LDH, elevated pump power and decreased pulse index, especially in the context of an infection.

          Related collections

          Most cited references31

          • Record: found
          • Abstract: not found
          • Article: not found

          ESC guidelines for the diagnosis and treatment of acute and chronic heart failure 2012: The Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure 2012 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association (HFA) of the ESC.

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            The 2013 International Society for Heart and Lung Transplantation Guidelines for mechanical circulatory support: executive summary.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Unexpected abrupt increase in left ventricular assist device thrombosis.

              We observed an apparent increase in the rate of device thrombosis among patients who received the HeartMate II left ventricular assist device, as compared with preapproval clinical-trial results and initial experience. We investigated the occurrence of pump thrombosis and elevated lactate dehydrogenase (LDH) levels, LDH levels presaging thrombosis (and associated hemolysis), and outcomes of different management strategies in a multi-institutional study. We obtained data from 837 patients at three institutions, where 895 devices were implanted from 2004 through mid-2013; the mean (±SD) age of the patients was 55±14 years. The primary end point was confirmed pump thrombosis. Secondary end points were confirmed and suspected thrombosis, longitudinal LDH levels, and outcomes after pump thrombosis. A total of 72 pump thromboses were confirmed in 66 patients; an additional 36 thromboses in unique devices were suspected. Starting in approximately March 2011, the occurrence of confirmed pump thrombosis at 3 months after implantation increased from 2.2% (95% confidence interval [CI], 1.5 to 3.4) to 8.4% (95% CI, 5.0 to 13.9) by January 1, 2013. Before March 1, 2011, the median time from implantation to thrombosis was 18.6 months (95% CI, 0.5 to 52.7), and from March 2011 onward, it was 2.7 months (95% CI, 0.0 to 18.6). The occurrence of elevated LDH levels within 3 months after implantation mirrored that of thrombosis. Thrombosis was presaged by LDH levels that more than doubled, from 540 IU per liter to 1490 IU per liter, within the weeks before diagnosis. Thrombosis was managed by heart transplantation in 11 patients (1 patient died 31 days after transplantation) and by pump replacement in 21, with mortality equivalent to that among patients without thrombosis; among 40 thromboses in 40 patients who did not undergo transplantation or pump replacement, actuarial mortality was 48.2% (95% CI, 31.6 to 65.2) in the ensuing 6 months after pump thrombosis. The rate of pump thrombosis related to the use of the HeartMate II has been increasing at our centers and is associated with substantial morbidity and mortality.
                Bookmark

                Author and article information

                Contributors
                k.caliskan@erasmusmc.nl
                Journal
                Neth Heart J
                Neth Heart J
                Netherlands Heart Journal
                Bohn Stafleu van Loghum (Houten )
                1568-5888
                1876-6250
                21 December 2015
                21 December 2015
                February 2016
                : 24
                : 2
                : 134-142
                Affiliations
                [ ]Thoraxcenter, Department of Cardiology, Unit Heart Failure, Heart Transplantation & Mechanical Circulatory Support, Erasmus Medical Center, ’s-Gravendijkwal 230, 3015 CE Rotterdam, The Netherlands
                [ ]Department of Intensive Care Unit, Erasmus Medical Center, Rotterdam, The Netherlands
                [ ]Cardialysis, Clinical Research Management & Core Laboratories, Rotterdam, The Netherlands
                [ ]Department of Cardiothoracic Surgery, Erasmus Medical Center, Rotterdam, The Netherlands
                Article
                786
                10.1007/s12471-015-0786-2
                4722006
                26689927
                a91286b8-06ce-418c-9062-18aaebfda4f6
                © The Author(s) 2015

                Open Access This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

                History
                Categories
                Original Article
                Custom metadata
                © The Author(s) 2016

                Cardiovascular Medicine
                thromboembolic event,pump thrombosis,haemolysis,left ventricular assist device (lvad),heartmate ii

                Comments

                Comment on this article