      High-output heart failure due to subclavian vessel fistula as a late complication following implantation of a biventricular pacemaker

      case-report

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction KEY TEACHING POINTS Key Teaching Points • Procedure-related complications can lead to deterioration of otherwise stable heart failure in cardiac resynchronization therapy patients and should be actively sought for. • High-flow arteriovenous fistulas can present as a late complication several years following pacemaker implantation. • A thorough physical examination including auscultation of the pacemaker-sided periclavicular area can help to diagnose arteriovenous fistulas. Biventricular pacing for cardiac resynchronization therapy (CRT) is a well-established treatment option for patients with inter- and/or intraventricular conduction delay with QRS width of more than 120 ms, systolic heart failure with ejection fraction (EF) <35%, and persistent symptoms of NYHA class II and higher despite optimal medical treatment. While some patients experience a large benefit from CRT, others (so-called “nonresponders”) do not. Factors that favor a beneficial response to CRT include QRS width of >150 ms and left bundle branch block morphology, resulting in a class IA recommendation for CRT in these patients according to current guidelines. 1 We present the case of a patient who met all the above criteria for a class IA indication to CRT but, following an initial improvement of NYHA class, deteriorated owing to an uncommon late complication of CRT implantation. Case report We report the case of a 74-year-old patient with ischemic cardiomyopathy due to extended 3-vessel disease with repetitive coronary interventions. Left ventricular function remained moderately impaired with EF 30%–35% despite optimal heart failure medication. As electrocardiography demonstrated left bundle branch block with a QRS width of 180 ms, the decision for implantation of a biventricular pacemaker/defibrillator device was taken in 2011. However, implantation of the coronary sinus lead via the left subclavian vein initially failed, leaving the patient with right-sided atrial and ventricular leads only. Because daily symptoms remained severe with exertional dyspnea NYHA class III, a second attempt for coronary sinus lead implantation was conducted in 2013 and succeeded without apparent complications. Two months later, both exercise tolerance and left ventricular function had significantly improved to NYHA II and EF 44%, respectively. In 2015, the patient reported a drastic decrease in exercise tolerance within a week, with dyspnea after a few steps of walking. Pacemaker control at that time revealed 99% biventricular pacing. Echocardiography demonstrated a diffuse deterioration of left ventricular function with an EF of 26%, and physical examination revealed a new left-sided continuous bruit, as well as a thrill in the middle subclavian region. Computed tomography demonstrated a large fistula between the left subclavian artery and the subclavian vein with a connecting orifice of 6 mm (Video 1, available online, and Figure 1). The shunt volume was estimated by duplex sonography to approximately 2 L/min (Figure 2). We hypothesized that the deterioration of left ventricular function and exercise tolerance was attributable to the volume overload associated with the shunt. Implantation of a covered stent graft (Be-Graft 8/27 mm) via the femoral artery was performed (Video 2, available online, and Figure 3). Both postinterventional angiography and duplex sonography demonstrated no residual shunting. The patient experienced a significant improvement of subjective functional capacity and was able to climb a flight of stairs within a few days of the intervention. A regular follow-up for reevaluation of left ventricular function was scheduled at 3 months. Unfortunately, the patient experienced non-ST-segment elevation myocardial infarction requiring stenting of the right coronary artery 2 weeks before scheduled follow-up. Echocardiography was performed following coronary intervention and demonstrated an EF of 30%. Discussion Arteriovenous fistulas (AVF) are infrequent complications in patients requiring cardiac pacemakers and are reported mostly following pacemaker lead extraction, rather than implantation. In this setting, Cronin et al 2 reported AVFs in only 8 out of 2471 patients (0.3%) that underwent pacemaker lead extraction. In only 1 of these cases, the AVF was located at the subclavian vessels. Surprisingly, most cases of AVF in this study were recognized not during the procedure, but only up to 587 days later. As in our case, the presence of a continuous bruit over the anterior chest was a consistent finding. The etiology of AVF development following lead extraction or, as in our case, lead implantation remains poorly understood. Direct trauma to the artery owing to inadvertent puncture and subsequent local arteriovenous shunting preventing closure of the puncture site might play an important role in the development of AVF early following the intervention. Because Cronin et al reported a delayed diagnosis of AVF following lead explantation in most cases, we presume that AVF size increases over time until the shunt volume is large enough to produce a significant bruit or hemodynamic symptoms. Following lead implantation, local compression to the vessel wall might contribute to late de novo formation of AVF, but clinical data to support this hypothesis are unavailable. The hemodynamic consequences of AVF are mostly studied in the context of AVF that were placed for hemodialysis. In these patients, central upper arm AVF were associated with almost 2-fold higher flow rates compared with peripheral forearm fistulas. 3 As a consequence, invasive hemodynamic measurement revealed increased cardiac output in patients with upper arm fistulas. 3 Although the general concept of high-output heart failure owing to increased AVF shunt volumes is well recognized, large controlled studies on this topic are scarce. Basile et al 4 found that a cutoff shunt volume of 2 L/min was predictive for the development of high-output heart failure. Although their calculations were based on a very limited number of patients in a subgroup with NYHA class III symptoms, it was considered a landmark study when published in 2008. More recently, an independent group also reported a higher risk for heart failure with shunt volumes of 2 L/min or more. 5 As a consequence, a shunt volume of 2 L/min or more is recommended for diagnosis of high-output heart failure. 6 Of note, these recommendations refer to patients without a history of heart failure for other reasons. Our patient presented with a calculated shunt flow of 2.1 L/min. Taking into account his underlying heart disease and severe symptoms despite otherwise optimal medical treatment, the decision was taken for an interventional approach to close the AVF by covered stenting. His functional status improved significantly over the following days, demonstrating that closure of high-flow AVF should be evaluated in this patient subset.

          Related collections

          Most cited references4

          • Record: found
          • Abstract: found
          • Article: not found

          The relationship between the flow of arteriovenous fistula and cardiac output in haemodialysis patients.

          Satisfactory haemodialysis (HD) vascular access flow (Qa) is necessary for dialysis adequacy. High Qa is postulated to increase cardiac output (CO) and cause high-output cardiac failure. Aim of the present prospective study was to evaluate the relationship between Qa of arteriovenous fistulas (AVFs) and CO in order to have a closer insight into this scarcely explored aspect of HD pathophysiology. Ninety-six patients bearing an AVF entered the study. All were evaluated a priori for the existence of cardiac failure according to the functional classification of the American College of Cardiology/American Heart Association task force. Qa and CO were measured by means of the ultrasound dilution Transonic Hemodialysis Monitor HD02. The mean Qa of the 65 lower arm AVFs was 0.948+/-0.428 SD l/min, whereas that of the 31 upper arm AVFs was 1.58+/-0.553 l/min. The difference was statistically significant (P or= 2.0 l/min predicted the occurrence of high-output cardiac failure more accurately than two other Qa values (sensitivity 89%, specificity 100%, curve area 0.99) and three Qa/CO ratio values (cardio-pulmonary recirculation-CPR). The better performance among the latter was that of CPR values >or= 20% (sensitivity 100%, specificity 74.7%, curve area 0.92). Our prospective study shows that the relationship between Qa of AVFs and CO is complex and a third-order polynomial regression model best fits this relationship. Furthermore, it is the first study to clearly show the high predictive power for high-output cardiac failure occurrence of Qa cut-off values >or= 2.0 l/min.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            The relation between vascular access flow and different types of vascular access with systemic hemodynamics in hemodialysis patients.

            Access flow (Qa) has an important effect on systemic hemodynamics in dialysis patients. A Qa : cardiac output (CO) ratio higher than 0.3 is considered a risk factor for high-output cardiac failure. However, the effect of different types of vascular access in hemodialysis patients has not yet been studied. The aim of the present study was to assess the relationship between Qa and systemic hemodynamics and to compare systemic hemodynamics between patients with elbow/upperarm access with forearm access types. Qa, CO, cardiac index (CI), central blood volume (CBV), and peripheral vascular resistance (PVR) were studied by the saline dilution technique in 58 hemodialysis patients (18 with elbow/upperarm access; 40 with forearm access types). This article found that Qa was significantly and positively related to CO and CI, and inversely related to PVR. CBV, Qa, and presence of cardiac failure were independent determinants of CI. Qa and the Qa : CO ratio were significantly higher, and PVR significantly lower, in patients with elbow/upperarm access compared to patients with forearm access types. When patients with cardiac failure were excluded, CO and CI were also significantly higher in patients with elbow/upperarm access types. Eleven percent of patients with elbow/upperarm fistula had a Qa : CO ratio above 0.3. In conclusion, Qa is strongly related to systemic hemodynamics in dialysis patients. In patients without cardiac failure, CO and CI are significantly higher in patients with elbow/upperarm access compared to patients with forearm access types. However, only a small percentage of patients with elbow/upperarm fistulae appeared to be in the risk zone for development of high-output cardiac failure.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              [Long-term effects of arteriovenous fistula on cardiac structure and function in non-diabetic hemodialysis patients].

              To evaluate the long-term effects of arteriovenous fistula (AVF) on heamodynamic changes and cardiac structure and function in non-diabetic hemodialysis patients.
                Bookmark

                Author and article information

                Contributors
                Journal
                HeartRhythm Case Rep
                HeartRhythm Case Rep
                HeartRhythm Case Reports
                Elsevier
                2214-0271
                11 July 2016
                September 2016
                11 July 2016
                : 2
                : 5
                : 425-427
                Affiliations
                [* ]Department of Cardiology, University Heart Center Freiburg - Bad Krozingen, Bad Krozingen, Germany
                []Department of Electrophysiology, University Heart Center Freiburg - Bad Krozingen, Bad Krozingen, Germany
                []Department of Radiology, University Heart Center Freiburg - Bad Krozingen, Bad Krozingen, Germany
                [§ ]Department of Angiology, University Heart Center Freiburg - Bad Krozingen, Bad Krozingen, Germany
                Author notes
                [* ] Address reprint requests and correspondence: Dr Björn Müller-Edenborn, University Heart Center Freiburg - Bad Krozingen, Südring 15, 79189 Bad Krozingen, Germany.University Heart Center Freiburg - Bad Krozingen, Südring 15, 79189Bad KrozingenGermany bjoern.mueller-edenborn@ 123456universitaets-herzzentrum.de
                Article
                S2214-0271(16)30058-6
                10.1016/j.hrcr.2016.06.005
                5426421
                afd4de66-a7c4-4d77-91a8-7bb733c0f896
                © 2016 Heart Rhythm Society. Published by Elsevier Inc.

                This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

                History
                Categories
                Case Report

                arrhyhtmias,catheter ablation and implantable cardioverter-defibrillator,hemodynamics,cardiomyopathy,heart failure,pacemaker,stent,vascular disease

                Comments

                Comment on this article