      Improving patient safety during introduction of novel medical devices through cumulative summation analysis

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          OBJECTIVE

          The aim of this study was to implement cumulative summation (CUSUM) analysis as an early-warning detection and quality assurance system for preclinical testing of the iSYS1 novel robotic trajectory guidance system.

          METHODS

          Anatomically accurate 3D-printed skull phantoms were created for 3 patients who underwent implantation of 21 stereoelectroencephalography electrodes by surgeons using the current standard of care (frameless technique). Implantation schema were recreated using the iSYS1 system, and paired accuracy measures were compared with the previous frameless implantations. Entry point, target point, and implantation angle accuracy were measured on postimplantation CT scans. CUSUM analysis was undertaken prospectively.

          RESULTS

          The iSYS1 trajectory guidance system significantly improved electrode entry point accuracies from 1.90 ± 0.96 mm (mean ± SD) to 0.76 ± 0.57 mm (mean ± SD) without increasing implantation risk. CUSUM analysis was successful as a continuous measure of surgical performance and acted as an early-warning detection system. The surgical learning curve, although minimal, showed improvement after insertion of the eighth electrode.

          CONCLUSIONS

          The iSYS1 trajectory guidance system did not show any increased risk during phantom preclinical testing when used by neurosurgeons who had no experience with its use. CUSUM analysis is a simple technique that can be applied to all stages of the IDEAL (idea, development, exploration, assessment) framework as an extra patient safety mechanism. Further clinical trials are required to prove the efficacy of the device.

          Related collections

          Most cited references20

          • Record: found
          • Abstract: found
          • Article: not found

          Stereoelectroencephalography: surgical methodology, safety, and stereotactic application accuracy in 500 procedures.

          Stereoelectroencephalography (SEEG) methodology, originally developed by Talairach and Bancaud, is progressively gaining popularity for the presurgical invasive evaluation of drug-resistant epilepsies.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Is SEEG safe? A systematic review and meta-analysis of stereo-electroencephalography-related complications.

            Stereo-electroencephalography (SEEG) is a procedure performed for patients with intractable epilepsy in order to anatomically define the epileptogenic zone (EZ) and the possible related functional cortical areas. By avoiding the need for large craniotomies and due to its intrinsic precision placement features, SEEG may be associated with fewer complications. Nevertheless, intracerebral electrodes have gained a reputation of excessive invasiveness, with a "relatively high morbidity" associated with their placement. A systematic literature review and meta-analysis of SEEG complications has not been previously performed. The goal of this study is to quantitatively review the incidence of various surgical complications associated with SEEG electrode implantation in the literature and to provide a summary estimate. This will allow physicians to accurately counsel their patients about the potential complications related to this method of extraoperative invasive monitoring.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Accuracy of intracranial electrode placement for stereoencephalography: A systematic review and meta-analysis.

              Stereoencephalography (SEEG) is a procedure in which electrodes are inserted into the brain to help define the epileptogenic zone. This is performed prior to definitive epilepsy surgery in patients with drug-resistant focal epilepsy when noninvasive data are inconclusive. The main risk of the procedure is hemorrhage, which occurs in 1-2% of patients. This may result from inaccurate electrode placement or a planned electrode damaging a blood vessel that was not detected on the preoperative vascular imaging. Proposed techniques include the use of a stereotactic frame, frameless image guidance systems, robotic guidance systems, and customized patient-specific fixtures.
                Bookmark

                Author and article information

                Journal
                Journal of Neurosurgery
                Journal of Neurosurgery Publishing Group (JNSPG)
                0022-3085
                1933-0693
                February 2018
                February 2018
                : 130
                : 1
                : 213-219
                Affiliations
                [1 ]1Department of Clinical and Experimental Epilepsy, Institute of Neurology, National Hospital for Neurology and Neurosurgery,
                [2 ]2Transitional Imaging Group, Centre for Medical Image Computing, and
                [3 ]3Department of Statistical Science, University College London, United Kingdom
                Article
                10.3171/2017.8.JNS17936
                5989930
                29451446
                bb7f15a3-d49d-4e70-9eb3-f3154671e7e6
                © 2018
                History

                Comments

                Comment on this article