      Wirksamkeit der Hörgeräteversorgung bei hochgradigem Hörverlust Translated title: Effectiveness of hearing aid provision for severe hearing loss

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Hintergrund

          In der Praxis liegt das unilaterale Einsilberverstehen mit Hörgerät bei 65 dB SPL (EV 65(HG)) häufig unter dem maximalen Einsilberverstehen aus dem Sprachaudiogramm (mEV), insbesondere bei Hörgeräteträgern mit hochgradigem Hörverlust. Diese Arbeit zielte darauf ab, den Wirkungsgrad Q der Hörgeräteversorgung, den Quotienten aus EV 65(HG) und mEV, bei Hörgeräteträgern mit hochgradigem bis an Taubheit grenzendem Hörverlust zu untersuchen.

          Material und Methoden

          Es wurden Daten aus In-situ-Messungen, dem Reinton- und Sprachaudiogramm und dem Sprachverstehen mit und ohne Hörgerät von 93 Ohren von 64 Patienten ausgewertet. Die Patienten stellten sich im Jahr 2019 für eine Hörgerätekontrolle in unserem Hörzentrum vor. Es wurden die Abweichung der in-situ gemessenen frequenzabhängigen Ausgangspegelwerte von den Zielvorgaben der präskriptiven Anpassformeln NAL-NL2 und DSL v5.0 analysiert. Für die Parameter EV 65(HG) und Q wurden jeweils die Spearman-Korrelationskoeffizienten für den Sprachverständlichkeitsindex (SII) berechnet.

          Ergebnisse

          Bei mehr als 67 % der Hörgeräteeinstellungen stimmten die Ausgangspegelwerte mit den Zielkurven für NAL-NL2 oder DSL v5.0 im Bereich von ±5 dB für Frequenzen von 0,5 bis 4 kHz für 65 dB SPL überein. Trotzdem wurde das mEV mit Hörgerät bei 65 dB SPL nicht erreicht (mittlere Abweichungen: 34,4 %). EV 65(HG) und Q waren jedoch am besten, wenn Zielwerte für DSL v5.0 bei 65 dB SPL erreicht wurden, was mit einem höheren SII einhergeht.

          Schlussfolgerung

          Für Hörgeräteträger mit hochgradigem bis an Taubheit grenzendem Hörverlust führen die Anpassformeln NAL-NL2 und DSL v5.0 nicht zu einer solchen Verstärkung, dass bei Alltagssprache von 65 dB SPL das mEV erreicht wird. Es bleibt zu untersuchen, ob alternative Präskriptionen mit besserer Hörbarkeit für Eingangspegel von 50 und 65 dB SPL den Wirkungsgrad der Hörgeräteversorgung verbessern könnten.

          Translated abstract

          Background and objective

          In practice, the unilateral monosyllabic speech recognition score with hearing aid (WRS 65(HA)) is often below the maximum word recognition score with headphones (WRS max), in particular for subjects with severe hearing loss. The aim of this study was to evaluate the efficiency factor Q of hearing aid provision, the ratio WRS 65(HA)/WRS max, in patients with severe to profound hearing loss.

          Materials and methods

          Data from real-ear measurements (REM), pure tone and speech audiogram, and speech recognition with and without hearing aid of 93 ears in 64 patients were examined. The patients visited the authors’ hearing center for hearing aid evaluation in 2019. Deviations of the real-ear measured frequency-dependent output level values from the prescription targets NAL-NL2 and DSL v5.0 were analyzed. Spearman correlation coefficients for the speech intelligibility index (SII) were calculated for the parameters WRS 65(HA) and Q.

          Results

          In more than 67% of the hearing aid fittings, output level values matched the target curves of NAL-NL2 or DSL v5.0 in the range of ±5 dB for frequencies from 0.5 to 4 kHz at 65 dB SPL. Nevertheless, WRS max was not achieved with hearing aid at conversational speech levels of 65 dB SPL (mean deviations: 34.4%). However, WRS 65(HA) and Q were best when target values for DSL v5.0 were achieved at 65 dB SPL, which is associated with a higher SII.

          Conclusion

          For patients with severe to profound hearing loss, the prescription targets of NAL-NL2 and DSL v5.0 do not provide sufficient amplification for WRS max to be achieved at a normal speech level of 65 dB SPL. It remains to be investigated whether alternative prescriptions with better audibility for input levels of 50 and 65 dB SPL might improve the effectiveness of hearing aid provision.

          Related collections

          Most cited references23

          • Record: found
          • Abstract: found
          • Article: not found

          Development and analysis of an International Speech Test Signal (ISTS).

          For analysing the processing of speech by a hearing instrument, a standard test signal is necessary which allows for reproducible measurement conditions, and which features as many of the most relevant properties of natural speech as possible, e.g. the average speech spectrum, the modulation spectrum, the variation of the fundamental frequency together with its appropriate harmonics, and the comodulation in different frequency bands. Existing artificial signals do not adequately fulfill these requirements. Moreover, recordings from natural speakers represent only one language and are therefore not internationally acceptable. For this reason, an International Speech Test Signal (ISTS) was developed. It is based on natural recordings but is largely non-intelligible because of segmentation and remixing. When using the signal for hearing aid measurements, the gain of a device can be described at different percentiles of the speech level distribution. The primary intention is to include this test signal with a new measurement method for a new hearing aid standard (IEC 60118-15).
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            The NAL-NL2 Prescription Procedure

            NAL-NL2 is the second generation of prescription procedures from The National Acoustic Laboratories (NAL) for fitting wide dynamic range compression (WDRC) instruments. Like its predecessor NAL-NL1 (Dillon, 1999), NAL-NL2 aims at making speech intelligible and overall loudness comfortable. This aim is mainly driven by a belief that these factors are most important for hearing aid users, but is also driven by the fact that less information is available about how to adjust gain to optimise other parameters that affect prescription such as localisation, tonal quality, detection of environmental sounds, and naturalness. In both formulas, the objective is achieved by combining a speech intelligibility model and a loudness model in an adaptive computer-controlled optimisation process. Adjustments have further been made to the theoretical component of NAL-NL2 that are directed by empirical data collected during the past decade with NAL-NL1. In this paper, the data underlying NAL-NL2 and the derivation procedure are presented, and the main differences from NAL-NL1 are outlined. The optimisation procedure A schematic overview of the adaptive optimisation procedure is shown in Figure 1. The two inputs to the process are the input speech spectrum and level, and the audiogram for which a prescription is required. The output is the prescription expressed as a gain-frequency response. The two input parameters are entered into two feedback loops, which operate in tandem to optimise the gain-frequency response. One loop uses an intelligibility model to find the gain-frequency response that maximises speech intelligibility. If left unchecked, this loop would produce the same output level irrespective of the input level of speech, a result that would not give the hearing aid wearer an acceptable representation of the auditory world. The second loop therefore uses a loudness model (Moore & Glasberg, 1997; 2004) to calculate the loudness that would be perceived by the hearing-impaired person with the selected gain-frequency response. The calculated loudness is compared to the loudness that would be perceived by a normal-hearing person listening to the same input speech spectrum and level. If loudness calculated for the audiogram exceeds the normal-hearing loudness, the overall gain is decreased. The adaptive process was used to derive the optimal gain-frequency responses for 240 audiograms, covering a wide range of severity and slopes, each at seven speech input levels from 40 to 100 dB SPL. Using a neural network, the optimised gain values from all the audiograms and all the input levels were drawn together into a single composite prescription formula. The theoretical derivation of NAL-NL2 differed from that of NAL-NL1 on two points. First, the intelligibility model, which is a revised version of the speech intelligibility index (SII) formula (ANSI, 1997), was updated. The difference between the original SII formula and the speech intelligibility model used to derive NAL-NL1 and NAL-NL2 is in the audibility factor. In the original SII formula the audibility factor assumes that, irrespective of the degree of hearing loss, speech is fully understood when all speech components are audible. In the speech intelligibility model used in the optimisation procedure, an effective audibility factor has been introduced, which takes into account that as the hearing loss gets more severe, less information is extracted from the speech signal, even when it is audible above threshold. The effective audibility factor was based on data collected at NAL prior to deriving NAL-NL1 (Ching, et al., 1998). The further revision made to the speech intelligibility model before deriving NAL-NL2 was based on more extensive data collected on 70 adults on how much information people with hearing loss can extract from speech once it has been made audible. Data were collected both in quiet and in noise and resulted in a new effective audibility factor. Second, constraints to the selected gain were applied such that no compression was introduced for speech presented below 50 dB SPL, and no gain was prescribed at frequencies below 50 Hz and above 16 kHz. The latter constraints ensured a smoother gain-frequency response between the two anchor points. These changes have resulted in NAL-NL2 prescribing a different gain-frequency response slope to NAL-NL1. Generally, NAL-NL2 prescribes relatively more gain across low and high frequencies and less gain across mid frequencies than NAL-NL1, see Figure 2. In the speech intelligibility model, an importance function is used to ensure that sufficient gain is applied at the frequencies that are most important for speech understanding. Low frequencies are more important in tonal languages, which are most common across Asia and Africa, than in non-tonal languages. Therefore, when deriving NAL-NL2, the optimisation procedure was run twice using different importance functions in the speech intelligibility model to derive gain for the two types of languages. As a result there are two versions of NAL-NL2, and slightly more gain is prescribed across the low frequencies for tonal than for non-tonal languages. Adjustments to optimised data Before deriving the theoretical NAL-NL2 formula, constraints were further applied to the optimised gain values such that the compression ratio for a given frequency and degree of hearing loss could not exceed a maximum value, selected to avoid any detrimental effect on speech understanding. Data have suggested that hearing aid users with severe or profound hearing loss prefer lower compression ratios than prescribed by NAL-NL1, when fitted with fast-acting compression (Keidser, et al., 2007). As demonstrated in Figure 3, the study participants selected lower compression ratios across the low than high frequencies, presumably to obtain a better preservation of both the speech envelope and the prosodic cues. There is, however, no reasons to believe that this population could not benefit from higher compression ratios, which would provide audibility of a wider range of input levels, when listening with slow-acting compression. Consequently, two sets of limits were applied to the optimised gain data, and as a result NAL-NL1 will prescribe higher compression ratios to people with a severe or profound hearing loss if they are fitted with slow-acting compression than when fitted with fast-acting compression, see Figure 4. Compression speed has no real effect on the compression ratio prescribed to hearing aid users with milder hearing loss. Adjustments to overall gain Many experiments in which NAL-NL1 has been used as baseline response have provided empirical data that suggest how much gain hearing aid users with different profiles prefer. For example, the realear insertion gain measurements of the fine-tuned, or preferred, response for a 65 dB input level by 187 adults who have participated in various research projects were analysed. The analysis revealed that female hearing aid users, irrespective of degree of hearing loss and experience with amplification preferred less gain (2 dB, on average) than male hearing aid users (Keidser & Dillon, 2006). The difference in preferred gain was statistically significant (P<0.05). Consequently, NAL-NL2 prescribes gender specific gain. The same data set showed that when the hearing loss was mild there was no difference in overall gain preferred by new and experienced hearing aid users. However, when the hearing loss became moderate, new hearing aid users preferred significantly less gain than experienced hearing aid users. In fact, progressively less gain was preferred by new users as the degree of hearing loss increased (Keidser, et al., 2008). Data collected for a small sample suggested that, on average, new hearing aid users with moderate hearing loss adapted to gain levels preferred by experienced hearing aid users with a similar degree of hearing loss over a period of about two years. On this background, NAL-NL2 recommends gain adaptation for new hearing aid users with more than a mild hearing loss. From the same data set, it could also be concluded that adults with mild and moderate hearing loss generally preferred less overall gain (3 dB, on average) than prescribed by NAL-NL1 for a 65 dB SPL input. At least two studies have further demonstrated that hearing aid users with mild or moderate hearing loss preferred a relatively higher gain reduction for higher input levels (80 dB SPL) but a relatively smaller gain reduction for lower input levels (50 dB SPL), which means that the adults preferred a slightly higher compression ratio than prescribed by NAL-NL1 (Smeds, et al., 2006; Zakis, et al., 2007), see Figure 5. In contrast, a study on children suggested that the younger population preferred higher gain than adults (Scollie, et al., 2010). An increase in gain is more likely to lead to greater speech intelligibility at low input levels where speech is most limited by audibility, and is less likely to cause noise-induced hearing loss for low input levels than for high input levels. Therefore, gain is for children increased for low input levels with a progressive decrease in increased gain with increased input level (Figure 5). That is, NAL-NL2 also prescribes a relatively higher compression ratio for children with mild or moderate hearing loss than do NAL-NL1. Conclusions NAL-NL2 is a revised version of NAL-NL1. The revisions are based on extensive empirical data. In comparison to NAL-NL1, NAL-NL2 prescribes a different gain-frequency response shape, and slightly higher compression ratios are prescribed for those with mild or moderate hearing loss. NAL-NL2 further takes the profile of the hearing aid user (age, gender, and experience), language, and compressor speed into consideration.
              Bookmark
              • Record: found
              • Abstract: not found
              • Book: not found

              World Report on Hearing

              WHO, WHO Team (2021)
                Bookmark

                Author and article information

                Contributors
                max.engler@uk-erlangen.de
                Journal
                HNO
                HNO
                Hno
                Springer Medizin (Heidelberg )
                0017-6192
                1433-0458
                21 January 2022
                21 January 2022
                2022
                : 70
                : 7
                : 520-532
                Affiliations
                GRID grid.411668.c, ISNI 0000 0000 9935 6525, Audiologische Abteilung, Hals-Nasen-Ohrenklinik, Kopf- und Halschirurgie, , Universitätsklinikum Erlangen, ; Waldstr. 1, 91054 Erlangen, Deutschland
                Article
                1139
                10.1007/s00106-021-01139-5
                9242947
                35061063
                c251a65b-0b10-4223-b74f-c8a031048e2b
                © The Author(s) 2022

                Open Access Dieser Artikel wird unter der Creative Commons Namensnennung 4.0 International Lizenz veröffentlicht, welche die Nutzung, Vervielfältigung, Bearbeitung, Verbreitung und Wiedergabe in jeglichem Medium und Format erlaubt, sofern Sie den/die ursprünglichen Autor(en) und die Quelle ordnungsgemäß nennen, einen Link zur Creative Commons Lizenz beifügen und angeben, ob Änderungen vorgenommen wurden.

                Die in diesem Artikel enthaltenen Bilder und sonstiges Drittmaterial unterliegen ebenfalls der genannten Creative Commons Lizenz, sofern sich aus der Abbildungslegende nichts anderes ergibt. Sofern das betreffende Material nicht unter der genannten Creative Commons Lizenz steht und die betreffende Handlung nicht nach gesetzlichen Vorschriften erlaubt ist, ist für die oben aufgeführten Weiterverwendungen des Materials die Einwilligung des jeweiligen Rechteinhabers einzuholen.

                Weitere Details zur Lizenz entnehmen Sie bitte der Lizenzinformation auf http://creativecommons.org/licenses/by/4.0/deed.de.

                History
                : 22 November 2021
                Funding
                Funded by: Universitätsklinikum Erlangen (8546)
                Categories
                Originalien
                Custom metadata
                © Springer Medizin Verlag GmbH, ein Teil von Springer Nature 2022

                freiburger sprachtest,maximales einsilberverstehen,sprachaudiometrie,in-situ-messung,hochgradiger bis an taubheit grenzender hörverlust,freiburg speech test,maximum monosyllabic word recognition score,speech audiometry,real-ear measurements,severe to profound hearing loss