13
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Disturbance-Estimated Adaptive Backstepping Sliding Mode Control of a Pneumatic Muscles-Driven Ankle Rehabilitation Robot

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          A rehabilitation robot plays an important role in relieving the therapists’ burden and helping patients with ankle injuries to perform more accurate and effective rehabilitation training. However, a majority of current ankle rehabilitation robots are rigid and have drawbacks in terms of complex structure, poor flexibility and lack of safety. Taking advantages of pneumatic muscles’ good flexibility and light weight, we developed a novel two degrees of freedom (2-DOF) parallel compliant ankle rehabilitation robot actuated by pneumatic muscles (PMs). To solve the PM’s nonlinear characteristics during operation and to tackle the human-robot uncertainties in rehabilitation, an adaptive backstepping sliding mode control (ABS-SMC) method is proposed in this paper. The human-robot external disturbance can be estimated by an observer, who is then used to adjust the robot output to accommodate external changes. The system stability is guaranteed by the Lyapunov stability theorem. Experimental results on the compliant ankle rehabilitation robot show that the proposed ABS-SMC is able to estimate the external disturbance online and adjust the control output in real time during operation, resulting in a higher trajectory tracking accuracy and better response performance especially in dynamic conditions.

          Related collections

          Most cited references53

          • Record: found
          • Abstract: not found
          • Article: not found

          Recent development of mechanisms and control strategies for robot-assisted lower limb rehabilitation

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Nonlinear disturbance observer design for robotic manipulators

              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              A pneumatically powered knee-ankle-foot orthosis (KAFO) with myoelectric activation and inhibition

              Background The goal of this study was to test the mechanical performance of a prototype knee-ankle-foot orthosis (KAFO) powered by artificial pneumatic muscles during human walking. We had previously built a powered ankle-foot orthosis (AFO) and used it effectively in studies on human motor adaptation, locomotion energetics, and gait rehabilitation. Extending the previous AFO to a KAFO presented additional challenges related to the force-length properties of the artificial pneumatic muscles and the presence of multiple antagonistic artificial pneumatic muscle pairs. Methods Three healthy males were fitted with custom KAFOs equipped with artificial pneumatic muscles to power ankle plantar flexion/dorsiflexion and knee extension/flexion. Subjects walked over ground at 1.25 m/s under four conditions without extensive practice: 1) without wearing the orthosis, 2) wearing the orthosis with artificial muscles turned off, 3) wearing the orthosis activated under direct proportional myoelectric control, and 4) wearing the orthosis activated under proportional myoelectric control with flexor inhibition produced by leg extensor muscle activation. We collected joint kinematics, ground reaction forces, electromyography, and orthosis kinetics. Results The KAFO produced ~22%–33% of the peak knee flexor moment, ~15%–33% of the peak extensor moment, ~42%–46% of the peak plantar flexor moment, and ~83%–129% of the peak dorsiflexor moment during normal walking. With flexor inhibition produced by leg extensor muscle activation, ankle (Pearson r-value = 0.74 ± 0.04) and knee ( r = 0.95 ± 0.04) joint kinematic profiles were more similar to the without orthosis condition compared to when there was no flexor inhibition (r = 0.49 ± 0.13 for ankle, p = 0.05, and r = 0.90 ± 0.03 for knee, p = 0.17). Conclusion The proportional myoelectric control with flexor inhibition allowed for a more normal gait than direct proportional myoelectric control. The current orthosis design provided knee torques smaller than the ankle torques due to the trade-off in torque and range of motion that occurs with artificial pneumatic muscles. Future KAFO designs could incorporate cams, gears, or different actuators to transmit greater torque to the knee.
                Bookmark

                Author and article information

                Journal
                Sensors (Basel)
                Sensors (Basel)
                sensors
                Sensors (Basel, Switzerland)
                MDPI
                1424-8220
                28 December 2017
                January 2018
                : 18
                : 1
                : 66
                Affiliations
                [1 ]School of Information Engineering, Wuhan University of Technology, Wuhan 430070, China; qingsongai@ 123456whut.edu.cn (Q.A.); zchengx0508@ 123456whut.edu.cn (C.Z.); zuojie@ 123456whut.edu.cn (J.Z.); quanliu@ 123456whut.edu.cn (Q.L.); s.q.xie@ 123456leeds.ac.uk (S.X.)
                [2 ]Key Laboratory of Fiber Optic Sensing Technology and Information Processing, Ministry of Education, Wuhan University of Technology, Wuhan 430070, China
                [3 ]School of Electronic and Electrical Engineering, University of Leeds, Leeds LS2 9JT, UK
                [4 ]Faculty of Engineering, Environment and Computing, Coventry University, Coventry CV1 5FB, UK; ab2032@ 123456coventry.ac.uk
                Author notes
                [* ]Correspondence: weimeng@ 123456whut.edu.cn ; Tel.: +86-131-6331-2360
                Author information
                https://orcid.org/0000-0003-0209-8753
                Article
                sensors-18-00066
                10.3390/s18010066
                5796385
                29283406
                cbc62440-7a2c-4a41-9777-a0474880cc58
                © 2017 by the authors.

                Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license ( http://creativecommons.org/licenses/by/4.0/).

                History
                : 01 November 2017
                : 26 December 2017
                Categories
                Article

                Biomedical engineering
                parallel robot,ankle rehabilitation,pneumatic muscles,disturbance estimation,adaptive sliding mode control

                Comments

                Comment on this article