          Abstract

          Ventricular assist devices (VADs) are implanted in patients with end-stage heart failure to provide both short- and long-term hemodynamic support. Unfortunately, bleeding and thromboembolic complications due to the severely disturbed, dynamic flow conditions generated within these devices require complex, long-term antiplatelet and anticoagulant therapy. While several studies have examined the effectiveness of one such agent, aspirin, under flow conditions, data comparing the efficacy of in vitro and in vivo metabolized aspirin is lacking. Two sets of studies were conducted in vitro with purified human platelets circulating for 30 min in a flow loop containing the DeBakey VAD (MicroMed Cardiovascular, Houston, TX, USA): (a) 20 μM aspirin was added exogenously in vitro to platelets isolated from aspirin-free subjects, and (b) platelets were obtained from donors 2 h (n = 14) and 20 h (n = 13) after ingestion of 1,000 mg aspirin. Near real-time platelet activation state (PAS) was measured with a modified prothrombinase-based assay. Platelets exposed to aspirin in vitro and in vivo (metabolized) showed 28.2 and 25.3 % reduction in platelet activation rate, respectively, compared to untreated controls. Our results demonstrate that in vitro treatment with antiplatelet drugs such as aspirin is as effective as in vivo metabolized aspirin in testing the effect of reducing shear-induced platelet activation in the VAD. Using the PAS assay provides a practical in vitro alternative to in vivo testing of antiplatelet efficacy, as well as for testing the thrombogenic performance of devices during their research and development.

          Related collections

          Most cited references46

          • Record: found
          • Abstract: found
          • Article: not found

          Aspirin-resistant thromboxane biosynthesis and the risk of myocardial infarction, stroke, or cardiovascular death in patients at high risk for cardiovascular events.

          We studied whether aspirin resistance, defined as failure of suppression of thromboxane generation, increases the risk of cardiovascular events in a high-risk population. Baseline urine samples were obtained from 5529 Canadian patients enrolled in the Heart Outcomes Prevention Evaluation (HOPE) Study. Using a nested case-control design, we measured urinary 11-dehydro thromboxane B2 levels, a marker of in vivo thromboxane generation, in 488 cases treated with aspirin who had myocardial infarction, stroke, or cardiovascular death during 5 years of follow-up and in 488 sex- and age-matched control subjects also receiving aspirin who did not have an event. After adjustment for baseline differences, the odds for the composite outcome of myocardial infarction, stroke, or cardiovascular death increased with each increasing quartile of 11-dehydro thromboxane B2, with patients in the upper quartile having a 1.8-times-higher risk than those in the lower quartile (OR, 1.8; 95% CI, 1.2 to 2.7; P=0.009). Those in the upper quartile had a 2-times-higher risk of myocardial infarction (OR, 2.0; 95% CI, 1.2 to 3.4; P=0.006) and a 3.5-times-higher risk of cardiovascular death (OR, 3.5; 95% CI, 1.7 to 7.4; P<0.001) than those in the lower quartile. In aspirin-treated patients, urinary concentrations of 11-dehydro thromboxane B2 predict the future risk of myocardial infarction or cardiovascular death. These findings raise the possibility that elevated urinary 11-dehydro thromboxane B2 levels identify patients who are relatively resistant to aspirin and who may benefit from additional antiplatelet therapies or treatments that more effectively block in vivo thromboxane production or activity.
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Bleeding and thrombosis in patients with continuous-flow ventricular assist devices.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Methods for the measurement of platelet function.

              This article discusses the advantages and disadvantages of methods for the measurement of platelet function. The focus is on tests that can be used to monitor antiplatelet activity in the setting of cardiovascular disease and potentially predict thrombosis and bleeding. The tests described are platelet aggregometry; impedance aggregometry; VerifyNow (Accumetrics, San Diego, CA); Plateletworks (Helena Laboratories, Beaumont, TX); platelet surface P-selectin, platelet surface-activated glycoprotein IIb/IIIa, and leukocyte-platelet aggregates; TEG Platelet Mapping system (Haemoscope, Niles, IL); Impact cone and plate(let) analyzer (DiaMed, Cressier, Switzerland); Platelet Function Analyzer-100 (Siemens Healthcare Diagnostics, Inc., Deerfield, IL); phosphorylation of vasodilator-stimulated phosphoprotein; serum thromboxane B(2); and urinary 11-dehydro thromboxane B(2). Some of the factors that differentiate these tests are sample volume requirements, the use of whole blood, the presence of shear, point-of-care status, need for a technician, and expense.
                Bookmark

                Author and article information

                Journal
                Journal of Thrombosis and Thrombolysis
                J Thromb Thrombolysis
                Springer Nature
                0929-5305
                1573-742X
                May 2014
                September 17 2013
                May 2014
                : 37
                : 4
                : 499-506
                Article
                10.1007/s11239-013-0997-6
                4160029
                24043375
                d44ad4b9-2893-4dc2-add7-83c854a13a9f
                © 2014
                History

                Comments

                Comment on this article