      Optimizing Wavelet ECG Watermarking to Maintain Measurement Performance According to Industrial Standard

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Watermarking is currently investigated as an efficient and safe method of embedding additional patient or environment-related data into the electrocardiogram. This paper presents experimental work on the assessment of the loss of ECG (electrocardiogram signal) diagnostic quality from the industrial standard EN60601-2-25:2015 point of view. We implemented an original time-frequency watermarking technique with an adaptive beat-to-beat lead-independent data container design. We tested six wavelets, six coding bit depth values (including the automatic noise-dependent one) and two types of watermark content to find the conditions that are necessary for watermarked ECG to maintain the compliance with International Electrotechnical Commission (IEC) requirements for interpretation performance. Unlike other authors, we did not assess the differences of signal values, but errors in ECG wave delineation results. The results of a total of 7300 original and watermarked 10 s ECGs were statistically processed to reveal possible interpretation quality degradation due to watermarking. Finally we found (1) the Symlet of 11-th order as the best of the wavelets that were tested; (2) the important role of ECG wave delineation and noise tracking procedures; (3) the high influence of the watermark-to-noise similarity of amplitude and values distribution and (4) the stability of the watermarking capacity for different heart rates in atrial rhythms.

          Related collections

          Most cited references33

          • Record: found
          • Abstract: not found
          • Article: not found

          Techniques for data hiding

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Compressed sensing for real-time energy-efficient ECG compression on wireless body sensor nodes.

            Wireless body sensor networks (WBSN) hold the promise to be a key enabling information and communications technology for next-generation patient-centric telecardiology or mobile cardiology solutions. Through enabling continuous remote cardiac monitoring, they have the potential to achieve improved personalization and quality of care, increased ability of prevention and early diagnosis, and enhanced patient autonomy, mobility, and safety. However, state-of-the-art WBSN-enabled ECG monitors still fall short of the required functionality, miniaturization, and energy efficiency. Among others, energy efficiency can be improved through embedded ECG compression, in order to reduce airtime over energy-hungry wireless links. In this paper, we quantify the potential of the emerging compressed sensing (CS) signal acquisition/compression paradigm for low-complexity energy-efficient ECG compression on the state-of-the-art Shimmer WBSN mote. Interestingly, our results show that CS represents a competitive alternative to state-of-the-art digital wavelet transform (DWT)-based ECG compression solutions in the context of WBSN-based ECG monitoring systems. More specifically, while expectedly exhibiting inferior compression performance than its DWT-based counterpart for a given reconstructed signal quality, its substantially lower complexity and CPU execution time enables it to ultimately outperform DWT-based ECG compression in terms of overall energy efficiency. CS-based ECG compression is accordingly shown to achieve a 37.1% extension in node lifetime relative to its DWT-based counterpart for "good" reconstruction quality.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Application of the phasor transform for automatic delineation of single-lead ECG fiducial points.

              This work introduces a new single-lead ECG delineator based on phasor transform. The method is characterized by its robustness, low computational cost and mathematical simplicity. It converts each instantaneous ECG sample into a phasor, and can precisely manage P and T waves, which are of notably lower amplitude than the QRS complex. The method has been validated making use of synthesized and real ECG sets, including the MIT-BIH arrhythmia, QT, European ST-T and TWA Challenge 2008 databases. Experiments with the synthesized recordings reported precise detection and delineation performances in a wide variety of ECGs, with signal-to-noise ratios of 10 dB and above. For real ECGs, the QRS detection was characterized by an average sensitivity of 99.81% and positive predictivity of 99.89%, for all the analyzed databases (more than one million beats). Regarding delineation, the maximum localization error between automatic and manual annotations was lower than 6 ms and its standard deviation was in agreement with the accepted tolerances for expert physicians in the onset and offset identification for QRS, P and T waves. Furthermore, after revising and reannotating some ECG recordings by expert cardiologists, the delineation error decreased notably, becoming lower than 3.5 ms, on average, and reducing by a half its standard deviation. This new proposed strategy outperforms the results provided by other well-known delineation algorithms and, moreover, presents a notably lower computational cost.
                Bookmark

                Author and article information

                Journal
                Sensors (Basel)
                Sensors (Basel)
                sensors
                Sensors (Basel, Switzerland)
                MDPI
                1424-8220
                11 October 2018
                October 2018
                : 18
                : 10
                : 3401
                Affiliations
                Department of Biocybernetics and Biomedical Engineering, AGH University of Science and Technology, 30 Mickiewicz Ave, 30-059 Krakow, Poland; aswierk@ 123456agh.edu.pl
                Author notes
                [* ]Correspondence: august@ 123456agh.edu.pl ; Tel.: +48-12-617-4712
                Author information
                https://orcid.org/0000-0001-5986-3247
                Article
                sensors-18-03401
                10.3390/s18103401
                6210888
                30314291
                dd56c59f-b08d-44c0-9aa8-9ff26b6a5d04
                © 2018 by the authors.

                Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license ( http://creativecommons.org/licenses/by/4.0/).

                History
                : 11 August 2018
                : 09 October 2018
                Categories
                Article

                Biomedical engineering
                time-frequency steganography,ecg watermarking,wavelets,interpretation performance standard,cse database

                Comments

                Comment on this article