      Imaging performance of a full-ring prototype PET-MRI system based on four-layer DOI-PET detectors integrated with a RF coil

      abstract
      , , , , , ,
      EJNMMI Physics
      Springer International Publishing
      PSMR 2015: 4th Conference on PET/MR and SPECT/MR
      17-21 May 2015

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We are developing a PET system integrated with a birdcage RF-coil for PET-MRI in order to realize both high sensitivity and high spatial resolution of the PET image by using the 4-layered depth-of-interaction (DOI) PET detector. We constructed a full-ring prototype system and evaluated performances, especially imaging performance, of the prototype system in simultaneous measurement. The prototype system consists of eight four-layer DOI-PET detectors and a prototype birdcage RF-coil developed for the proposed system. The PET detectors consist of six monolithic multi-pixel photon counter array (S11064-050P), a readout circuit, fourlayer DOI scintillator arrays and a shielding box made of 35 μm thick copper foil. The crystal array consists of 2.0 mm x 2.0 mm x 5.0 mm LYSO crystals arranged in 38 x 6 x 4 layer. The RF-coil has eight coil elements and the eight PET detectors are positioned at each element gap. The diameter of the RF-coil elements is 261 mm. We conducted performance tests of the prototype system with a 3.0 T MRI (MAGNETOM Verio). Only the PET detectors, the RF-coil and the cables were in an MRI room during measurements. A data acquisition system and power supplies for the MPPCs and preamplifiers were outside the MRI room and connected to all the detectors through a penetration panel. As a result, the spatial resolutions of a Na-22 point source in the PET image were lower than 1.6 mm in whole the FOV due to the DOI capability. In addition, the influence of the simultaneous measurements on the PET performance is negligible. On the other hand, the SNR of the phantom image in the magnitude images was degraded from 259.7 to 209.4 due to noise contamination from the power supplies.

          Related collections

          Author and article information

          Conference
          EJNMMI Phys
          EJNMMI Phys
          EJNMMI Physics
          Springer International Publishing (Cham )
          2197-7364
          18 May 2015
          18 May 2015
          December 2015
          : 2
          Issue : Suppl 1 Issue sponsor : Publication of this supplement was funded by COST (European Cooperation in Science and Technology).
          : A18
          Affiliations
          [ ]National Institute of Radiological Sciences, Chiba, Japan
          [ ]Chiba University, Chiba, Japan
          Article
          128
          10.1186/2197-7364-2-S1-A18
          4798672
          26956273
          ed884d31-4bdd-4f51-97c4-93472eae4627
          © Nishikido et al; licensee Springer. 2015

          This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

          PSMR 2015: 4th Conference on PET/MR and SPECT/MR
          La Biodola, Isola d'Elba, Italy
          17-21 May 2015
          History
          Categories
          Meeting Abstract
          Custom metadata
          © The Author(s) 2015

          Comments

          Comment on this article