      Performances of the Vitek MS matrix-assisted laser desorption ionization-time of flight mass spectrometry system for rapid identification of bacteria in routine clinical microbiology.

      Journal of Clinical Microbiology
      Bacteria, chemistry, classification, isolation & purification, Bacterial Infections, diagnosis, Bacteriological Techniques, methods, Diagnostic Errors, Humans, Sensitivity and Specificity, Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Rapid and cost-effective matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS)-based systems will replace conventional phenotypic methods for routine identification of bacteria. We report here the first evaluation of the new MALDI-TOF MS-based Vitek MS system in a large clinical microbiology laboratory. This system uses an original spectrum classifier algorithm and a specific database designed for the identification of clinically relevant species. We have tested 767 routine clinical isolates representative of 50 genera and 124 species. Vitek MS-based identifications were performed by means of a single deposit on a MALDI disposable target without any prior extraction step and compared with reference identifications obtained mainly with the VITEK2 phenotypic system; if the identifications were discordant, molecular techniques provided reference identifications. The Vitek MS system provided 96.2% correct identifications to the species level (86.7%), to the genus level (8.2%), or within a range of species belonging to different genera (1.3%). Conversely, 1.3% of isolates were misidentified and 2.5% were unidentified, partly because the species was not included in the database; a second deposit provided a successful identification for 0.8% of isolates unidentified with the first deposit. The Vitek MS system is a simple, convenient, and accurate method for routine bacterial identification with a single deposit, considering the high bacterial diversity studied and as evidenced by the low prevalence of species without correct identification. In addition to a second deposit in uncommon cases, expanding the spectral database is expected to further enhance performances.

          Related collections

          Author and article information

          Comments

          Comment on this article