724
views
0
recommends
+1 Recommend
1 collections
    1
    shares

      2023 Journal Citation Reports Journal Impact Factor is 0.9. Scopus Citescore 0.8. 

      Interested in becoming a CVIA published author?

      • Platinum Open Access with no APCs. 
      • Fast peer review/Fast publication online after article acceptance.

      Submissions should be made electronically at: https://mc04.manuscriptcentral.com/cvia-journal.

      Please refer to the Author Guidelines at https://cvia-journal.org/instructions-to-authors/ before submission.

       

      scite_
       
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      The Subcutaneous Implantable Cardioverter-Defibrillator: A Practical Review and Real-World Use and Application

      Published
      review-article
      , MD, FACC, FHRS 1 , 2 , , , MD, FACC, FHRS 2
      Cardiovascular Innovations and Applications
      Compuscript
      subcutaneous implantable cardioverter-defibrillator, sudden cardiac death
      Bookmark

            Abstract

            The subcutaneous implantable cardioverter-defibrillator (ICD) is a novel technology using a subcutaneous (extrathoracic) system for treatment of potential lethal ventricular arrhythmias. It avoids many of the risks of transvenous ICD implantation. It may be considered in patients having an ICD indication who do not have a pacing and/or cardiac resynchronization therapy indication, and who are unlikely to benefit from antitachycardia pacing therapy. We review patient selection, system components, the implantation technique, and screening considerations for subcutaneous ICD implantation. Its uses in specific patient populations, including children, patients with congenital heart disease, hypertrophic cardiomyopathy, or end-stage renal disease, and patients with preexisting pacemakers, are highlighted. Areas of future investigation are reviewed, including potential use with leadless pacing and magnetic resonance imaging.

            Main article text

            Abbreviations

            ATP

            antitachycardia pacing

            CHD

            congenital heart disease

            CIED

            cardiac implantable electronic device

            ESRD

            end-stage renal disease

            HCM

            hypertrophic cardiomyopathy

            ICD

            implantable cardioverter-defibrillator

            IDE

            investigational device exemption

            LVAD

            left ventricular assist device

            MRI

            magnetic resonance imaging

            S-ICD

            subcutaneous implantable cardioverter-defibrillator

            TV-ICD

            transvenous implantable cardioverter-defibrillator

            VF

            ventricular fibrillation

            VT

            ventricular tachycardia

            Introduction

            Sudden cardiac death remains a significant public health problem and is a leading cause of cardiovascular death, with approximately half of all cardiovascular deaths in the United States occurring suddenly [1]. Implantable cardioverter-defibrillators (ICDs) have been shown to decrease the risk of sudden cardiac death in both primary prevention trials [2, 3] and secondary prevention trials [4]. The subcutaneous ICD (S-ICD; Boston Scientific, Marlborough, MA, United States) was developed as an alternative therapy to transvenous ICD (TV-ICD) systems, as it is a fully subcutaneous system that does not require leads within (endovascular) or on the epicardial surface of the heart [5]. TV-ICD implantation carries immediate complication risks of pneumothorax, cardiac perforation, and lead dislodgement [6]; as well as long term risks of lead malfunction [7, 8], device-related infection, and venous occlusion. The S-ICD was developed to decrease periprocedural implantation risks, eliminate venous access difficulties, and reduce endovascular mechanical stress on leads. In addition, transvenous lead extraction is increasing in frequency because of the increasing number of cardiac implantable electronic devices (CIEDs), their risk of infection, and recent ICD lead recalls [79]. The S-ICD decreases the risk of future extraction-related morbidity associated with nonfunctional and infected TV-ICD systems.

            The S-ICD system may be considered for patients meeting ICD guideline criteria [10] for either primary or secondary prevention who do not have a pacing and/or cardiac resynchronization therapy indication. In addition, the S-ICD cannot deliver antitachycardia pacing (ATP) [5]. Therefore patients with known pace-terminable ventricular tachycardia (VT) and/or recurrent monomorphic VT who may benefit from ATP therapies should be considered for a TV-ICD system rather than an S-ICD system.

            S-ICD Components/Implantation Procedure

            The S-ICD system consists of a pulse generator (130 g) positioned over the sixth rib between the mid to anterior left axillary line and a single (45 cm long) lead containing two sensing electrodes positioned adjacent to either end of an 8 cm shock coil. The lead is tunneled across the chest wall and positioned parallel to and 1 to 2 cm to the left of the sternal midline, with the distal sensing electrode secured adjacent to the manubriosternal junction, and the proximal sensing electrode secured adjacent to the xiphoid process [5]. Standard implantation involves three incisions: one lateral pocket incision and two parasternal incisions. However, reports of a two-incision technique eliminating the superior parasternal incision appear safe, and may simplify future implantation [11]. A submuscular approach to S-ICD generator placement has also been described that may reduce pocket-related complications in select patients [12, 13] but it needs further clinical validation. Care should also be taken not to introduce subcutaneous air, as it can insulate the sensing electrodes, causing inadequate sensing and low-amplitude signals, leading to oversensing due to auto gain and potential inappropriate shocks [14].

            S-ICD Advantages

            The S-ICD may be implanted strictly by anatomical landmarks without the use of fluoroscopy, limiting radiation exposure of the patient and providers [5]. It does not require favorable venous anatomy, which may be beneficial in patients with limited venous access such as patients undergoing dialysis and with congenital heart disease (CHD). It can preserve vascular access for future use and avoids the risks of future transvenous lead extractions. There is a potentially lower risk of systemic infection and a decreased risk of procedural complications, including pneumothorax and cardiac perforation [15]. There is also less biomechanical stress on a subcutaneous lead as it is not exposed to the dynamics of cardiac motion and/or subclavian crush forces. Procedure times are more predictable, with an average procedure time of 69±27 min in the EFFORTLESS (Boston Scientific postmarket S-ICD) registry [16].

            S-ICD Disadvantages

            The major disadvantage of the S-ICD system is its lack of pacing capabilities, including ATP therapies. This not only excludes some patients at the time of implantation, but may also be problematic in a minority of patients who develop symptomatic bradyarrhythmias after S-ICD implantation. The generator is larger than that of TV-ICDs, and the S-ICD has no impedance monitoring for surveillance of congestive heart failure patients. The S-ICD is currently not magnetic resonance imaging (MRI) compatible and it should be used with caution in patients with a history of “slow” VT, as the lowest programmed detection rate is 170 bpm. The S-ICD does not have a programmable monitor zone, which can be useful in detecting the presence of arrhythmias with rates occurring below the programmed detection intervals. The S-ICD also has a higher frequency of T-wave oversensing compared with TV-ICD systems [15, 16].

            System Characteristics

            The S-ICD can deliver only 80 J biphasic shocks (no other energy is programmable). It may deliver up to five shocks per tachycardia episode, and can reverse shock polarity if an initial shock fails. As previously noted, the S-ICD has no baseline pacing capabilities. It can deliver postshock demand pacing at 50 bpm only for up to 30 s [5]. After 30 s, no further pacing will occur regardless of the underlying rhythm. Battery longevity of first-generation devices was approximately 5 years [17]. Second-generation devices are expected to have an average longevity of 7.3 years [18]. The S-ICD can store data on more than 40 arrhythmic events (treated and untreated).

            The S-ICD detects changes in ventricular rates by using modified subsurface electrocardiography via a primary (proximal electrode to can), a secondary (distal electrode to can), or an alternate (distal electrode to proximal electrode) sensing vector (Figure 1). It automatically determines the optimal sensing vector on the basis of an R wave to T wave ratio that attempts to avoid double QRS counting or T-wave oversensing [19]. It measures rate as a rolling average of four consecutive beats, with VT/ventricular fibrillation (VF) detection indicated when 18 of 24 consecutive intervals fall within the detection zone. If this criterion is met, the S-ICD can charge capacitors and deliver a shock (Figure 2) [5, 19]. All device settings are automated except for shock therapy (on/off), postshock pacing (on/off), and conditional discrimination of supraventricular tachycardia (on/off) [5]. The device is programmable as either a single or a dual zone device. In single-zone configurations, there is a shock-only zone that relies solely on heart rate. Dual-zone programming incorporates an additional conditional zone that uses a unique morphology-based discrimination algorithm to classify rhythms as either shockable or nonshockable [20].

            The mean time to therapy in the EFFORTLESS registry for induced ventricular arrhythmia (defibrillation testing) was 15.1 s (7–37 s), and for spontaneous ventricular events was 17.5 s (6–29 s). Because of a longer time to therapy, S-ICD shocks were withheld for 46% of recorded episodes [16]. An early report raised concerns of syncope due to delays in therapy secondary to arrhythmia undersensing [21]. However, in pooled analysis of the investigational device exemption (IDE) trial and the EFFORTLESS registry, syncope was reported in only 1.7% of patients [22]. Bardy et al. [5] evaluated defibrillation threshold testing in patients in whom both an S-ICD and a TV-ICD has been implanted. The S-ICD required higher energy on average (36.6±19.8 J vs 11.1±8.5 J).

            Figure 1

            Subcutaneous Implantable Cardioverter-Defibrillator Sensing: Primary Vector B-can, Secondary Vector A-can, and Alternate Vector A-B.

            A, Distal electrode; B, proximal electrode.

            Figure 2

            Subcutaneous Implantable Cardioverter-Defibrillator (S-ICD): Conversion of Clinical Ventricular Fibrillation with a Single S-ICD Shock Back To Normal Sinus Rhythm.

            Mild undersensing of fine ventricular fibrillation is noted. C, capacitor charging; P, paced event; S, sensing of an event not classified as tachycardia; T, sensed event classified as tachycardia; lightning bolt, shock.

            S-ICD ECG Vector Screening

            All patients should undergo preprocedure S-ICD screening. This ensures that an adequate R wave to T wave ratio is present, attempting to avoid T-wave oversensing. ECG leads are positioned to mimic S-ICD sensing vectors. Evaluation of all three vectors with the patients in both the supine position and the standing position is performed for 10 s, with use of an ECG screening tool. Patients qualify for an S-ICD if all beats in any screening vector pass in both the supine position and the standing position [23]. Approximately 7–8% of patients fail the screening test [23, 24]. Reported clinical predictors of failed screening include hypertrophic cardiomyopathy (HCM), heavy weight, and prolonged QRS duration [24]. ECG predictors of failed screening include simultaneous T wave inversions in leads I, II, and aVF [23], and an R wave to T wave ratio of less than 3 in the ECG lead with the largest T wave [24]. Conversion of an ECG screening failure to a pass has been reported with a right parasternal electrode configuration [25], but safety needs to be evaluated further. If patients pass only in the alternate vector, additional testing with the patient bending/flexing forward should be considered as reports of oversensing in this position (due to low R wave amplitude), leading to triple counting of the P/QRS/T waves and inappropriate shocks, have been reported [26].

            S-ICD IDE Trial and the EFFORTLESS Registry (VT/VF Conversion and Infection Rates)

            The Food and Drug Administration IDE trial [15] enrolled 330 patients, 74% male with a mean age of 51.9±15.5 years and mean left ventricular ejection fraction of 36.1±15.9%. Seventy-nine percent had a primary prevention indication. Patients had to be older than 18 years, meet the ICD guideline criteria, and pass ECG vector screening. Epicardial patches, unipolar pacers, severe renal dysfunction, and patients with pace-terminable VT were excluded. The system complication-free rate at 180 days was 99%. The primary efficacy end point, defined as two consecutive defibrillations at 65 J out of a possible four attempts, was 100%. An additional substudy repeated defibrillation testing at 6 months in 78 patients, with a 96% conversion rate with one 65 J shock and a 100% conversion rate with shock of up to 80 J. One hundred nineteen episodes of VT/VF (81 during VT storms and 38 discrete episodes) had 100% spontaneous conversion or termination with an 80 J shock, with no arrhythmic deaths. There were 18 device infections (5.6%), with four (1.3%) requiring explanation. There were increased episodes of infections early in the trial that decreased with operator experience.

            The EFFORTLESS registry [16] is an observational nonrandomized multicenter standard-of-care registry of 472 patients, with interim analysis available. The mean age was 49±18 years (younger than typical TV-ICD trials), with a significant proportion of nonischemic cardiomyopathy (31%), channelopathies (13%), and CHD (7%), which differed from the IDE trial population. Sixty-three percent of devices were placed for primary prevention compared with 79% in the IDE trial. Only 29% of patients had a diagnosis of congestive heart failure, and the mean level ventricular ejection fraction was 42%. The complication-free procedure rate at 360 days was 94%. Successful defibrillation (defined as successful conversion at 80 J or less) was 99.7%. Ninety-one episodes of VT/VF (40 during VT storm and 51 discrete episodes) occurred. Ninety-six percent of discrete episodes terminated with one to five shocks, and single-shock efficacy was 88%. One episode spontaneously terminated after the fifth shock. Only one patient died of an arrhythmic death, a patient with Löffler’s syndrome receiving high-dose steroids who previously had successful defibrillation threshold testing.

            In pooled analysis (IDE trial and EFFORTLESS registry), only 17 devices needed removal/revision because of infection (1.7%) [22]. In patients in whom an S-ICD had been implanted after TV-ICD extraction, the risk of S-ICD infection is low even in patients in whom the TV-ICD was removed because of prior TV-ICD infection [27].

            S-ICD and Inappropriate Shocks

            S-ICD inappropriate shocks are more commonly caused by double counting of cardiac signals (most frequently T-wave oversensing) rather than supraventricular arrhythmias [15]. The system has shown excellent discrimination of supraventricular arrhythmias from ventricular arrhythmias [15, 28, 29]. Early clinical experience raised concern of inappropriate shocks, with rates up to 15% [30, 31]. In the IDE trial, 41 patients (13.1%) received inappropriate shocks (22 from T-wave oversensing). Of these, 32 patients underwent device reprogramming that solved the issue but nine patients required reoperation [15]. The addition of dual-zone programming significantly decreased the rate of inappropriate shocks in the IDE study [15]. In the EFFORTLESS registry the inappropriate shock rate was much improved at 7% [16], but was slightly higher than in recent TV-ICD trials [32]. In pooled analysis (IDE trial and EFFORTLESS registry), the 3-year Kaplan-Meier estimated inappropriate shock rate was 11.7% for dual-zone programming and 20.5% for single-zone programming [22]. Data support the use of dual-zone programming as a standard setting for S-ICD patients [20, 22].

            The initial step in avoiding inappropriate shocks is proper preprocedure screening. Once the S-ICD has been implanted, dual-zone programming should be considered [20]. Inappropriate shocks may be managed after implantation by reprogramming of the sensing vector and/or therapy zones of the device with use of a template acquired during exercise. Exercise-optimized programming may be considered to prevent future inappropriate shocks and should be considered in patients who are felt to be at high risk of T-wave oversensing [33] such as HCM patients. A new S-ICD discrimination algorithm has been developed to reduce T-wave oversensing but will need to be evaluated prospectively [29]. Reevaluation of ECG screening should be considered with any change in QRS morphology, such as new bundle branch blocks. Inappropriate shocks due to myopotentials [34] and electromagnetic interference [15, 35] have rarely been reported.

            Real-World Populations and Clinical Scenarios

            S-ICD and Pacemakers

            Patients with pacing indications should be considered for TV-ICD systems rather than an S-ICD. However, S-ICDs have been implanted in patients with preexisting pacemakers and vascular access difficulties. In pooled analysis (IDE trial and EFFORTLESS registry), 2.2% of patients (n=19) had preexisting pacemakers at the time of S-ICD implantation [22]. Pacemaker devices were programmed to be bipolar, as unipolar pacing is contraindicated (owing to inappropriate sensing related to the large unipolar pacing artifact) [15]. Patients have the same ECG screening as nonpaced patients. If intrinsic rhythm is present, both paced and intrinsic rhythms should be evaluated and pass in the same vector. If atrial arrhythmias are present, the stored conditional zone morphology template should reflect the native intrinsic rhythm (if present) rather than the ventricular paced rhythm. Once the devices have been implanted, interference between both devices should be evaluated, as pacing spikes could be counted independently from R waves [36]. Porterfield et al. [36] reported placement of an S-ICD in a patient with preexisting pacemaker and complete atrioventricular block who failed ECG screening. S-ICD arrhythmia detection was programmed at more than two times the upper rate limit of the pacemaker to prevent inappropriate shocks from T-wave oversensing (double counting). They also evaluated sensing at the maximum pacer output to ensure the S-ICD did not mark pacer spikes.

            Some pacemakers have backup safety modes which may switch them to unipolar mode, and this feature should be disabled if possible [36]. Akin et al. [37] reported defibrillation threshold testing in a pacemaker-dependent patient in whom the pacemaker was programmed DOO. During testing the patient received a shock that failed to convert VF, but switched the pacemaker to unipolar mode. The larger unipolar pacing signals were sensed by the S-ICD, causing undersensing of VF, requiring external defibrillation. Consideration can also be given for programming postshock pacing off, as it could inhibit intrinsic pacemakers [36]. There are also reports of successful placement of S-ICDs in patients with bipolar epicardial leads [38, 39].

            S-ICD and End-Stage Renal Disease

            Because of the risk of venous occlusion and device infections with TV-ICD systems, S-ICD systems may be considered for use in patients with end-stage renal disease (ESRD). However, data in ESRD patients are limited. Patients with a glomerular filtration rate less than 29 mL/min were excluded from the Food and Drug Administration IDE trial [15] and only 9% of patients in the EFFORTLESS registry [16] had renal disease, with a pooled analysis showing only 34 patients (3.9%) with glomerular filtration rates less than 45 mL/min [22]. However, reports from two large university facilities suggest a higher percentage of S-ICD implantation in ESRD patients (28 of 92 [40] and 17 of 74 [41]) than seen in clinical trials. In ESRD an S-ICD can be used in patients with recurrent thrombosis (related to transvenous leads), in patients with preexisting stenosis preventing transvenous device placement, and in patients at risk of developing intravascular stenosis and/or infection [42].

            El-Chami et al. [40] reported a single-center experience of S-ICD implantation in 79 patients (27 with ESRD undergoing dialysis). Patients with HCM and channelopathies were excluded from the analysis. ESRD patients were older and likelier to have diabetes. The annual rate of the combined primary end point (death, heart failure, hospitalization, or appropriate S-ICD shock), although not significant, was higher in the dialysis cohort (23.8%/year vs 10.9%/year, P=0.317) driven by a higher incidence of appropriate shocks (17.0%/year vs 1.4%/year, P=0.021). There were no S-ICD-related infections in the ESRD patients. The rate of inappropriate shocks was similar (6.0%/year of dialysis vs 6.8%/year for no dialysis, P=0.509). The low inappropriate shock rate is notable, as concerns exist for T-wave oversensing in patients with fluctuating potassium levels and QRS/T wave amplitudes. Kiamanesh et al. [43] reported an episode of hyperkalemia (K+ 7.0 mmol/L) in an ESRD patient leading to T-wave oversensing and an inappropriate shock that initiated VF, requiring four additional shocks before VF termination. However, despite the potential for transient potassium level fluctuations with QRS/T wave amplitude changes, inappropriate shocks due to hyperkalemia have rarely been reported.

            S-ICD and HCM

            Pooled analysis (IDE trial and EFFORTLESS registry) included 96 patients with HCM (75% male), 88.5% with a primary prevention indication. One-year complication rates including inappropriate shocks were equivalent between HCM and non-HCM patients, with no observed deaths. Only four HCM patients experienced VT, all of which were appropriately sensed and defibrillated [44]. Careful preprocedure ECG screening is important in HCM to prevent T-wave oversensing and inappropriate shocks. HCM has been suggested as an independent predictor for ECG screening failure [24] S-ICD screening failure rates of 7% [45] to 16% [46] have been reported, with failure rates of 36% in patients considered at high-risk of sudden cardiac death [46]. Primary causes of screening failure were high T-wave voltage in 25% of vectors, T-wave inversions in more than two surface leads, and prior myectomy [46]. Exercise ECG screening should be considered as it can unmask unsuitable HCM patients formerly thought eligible at rest [45]. Postimplant exercise-based optimization should also be considered [33, 45]. This should include examination of all three sensing vectors during exercise as well as during acquisition of subcutaneous ECG templates [47]. Right parasternal chest electrode position could also be assessed if screening fails with the traditional electrode position [45].

            A single-center experience in 18 patients with HCM reported T-wave oversensing in seven patients (39%), with four patients (22%) having inappropriate shocks. Reprogramming of the sensing vector eliminated T-wave oversensing in three of the four patients. A low R wave to T wave ratio was a major risk factor for the occurrence of oversensing in this patient cohort [47].

            S-ICD with Left Ventricular Assist Devices and Other Implantable Devices

            S-ICDs have been placed in conjunction with several CIEDs and noncardiac implantable devices. Saeed et al. [48] reported placement of a HeartWare (Framingham, MA, United States) left ventricular assist device (LVAD) in a patient with preexisting S-ICD. After implantation, noise was noted in the primary and secondary sensing vectors, both of which incorporate the pulse generator close to the LVAD. No inappropriate therapies were noted; however, the device could have withheld appropriate therapy in the setting of true arrhythmia. The alternate vector was not affected. Another patient had a preexisting HeartMate II LVAD (Thoractec Corporation; Pleasanton, CA, United States), and an S-ICD was implanted in that patient. Sensing in all three vectors was found to be appropriate without any interference from the continuous-flow LVAD. The S-ICD system initially chose the secondary vector, later programmed to the primary vector. There was no interference in sensing or shock delivery from the S-ICD. LVAD readings were unchanged. The HeartMate II functions at higher speeds that appeared less likely to cause interference [49]. Reevaluation of S-ICD sensing after LVAD implantation or after LVAD speed change should be considered to exclude device-device interaction.

            Kuschyk et al. [50] reported S-ICD implantation in six patients with cardiac contractility modulation devices and one vagus nerve stimulator. Cardiac contractility modulation devices deliver high-voltage biphasic electrical impulses during the absolute refractory period which enhance the contractile strength of failing myocardium. S-ICD implantation in patients with a cardiac contractility modulation device appeared safe during an intermediate follow-up period (mean 17 months) [50]. Intraoperative and postprocedure cross talk testing in patients with both a CIED and an S-ICD should be performed, and ergonomic testing should be considered. The sensing vector with the clearest result should be programmed. Bader et al. [51] reported S-ICD implantation in a patient with a preexisting deep brain stimulator. The deep brain stimulator should be programmed to be bipolar if possible; however, in the short term no artifact was seen with either unipolar mode or bipolar mode, likely due to S-ICD signal filtering.

            S-ICD in Children and CHD

            TV-ICDs have been effective in managing malignant arrhythmias in selected pediatric and adult CHD patients [52]. Potential transvenous lead–related problems are of particular concern in the young because of increased physical activity, ongoing growth, and longer life expectancy, allowing more time for potential lead complications [53]. Younger patients are likelier to undergo implantation for HCM and ion channelopathies, disorders likelier to present with VF, making lack of ATP capabilities less worrisome. Limited experience exists with S-ICDs in children. Patients (younger than 18 years) were excluded from the IDE trial. Early experience raised concern for wound dehiscence and threatened generator erosion in young patients [21]. An observational case-control series evaluated patients (younger than 20 years, range 10–18 years) receiving an S-ICD (n=9) and case-matched TV-ICD patients (n=8). Three patients received appropriate S-ICD shocks. In limited follow-up, the S-ICD may offer similar survival benefit to the TV-ICD but with a lower incidence of complications requiring reoperations [53]. Future evaluation of a submuscular approach may decrease concern for skin erosion in children. Alternative generator and coil configurations have been reported in small patients [54].

            Patients with CHD often have complex cardiac and venous anatomy, making TV-ICD placement challenging. S-ICD placement may therefore be considered, but limited experience exists. In the EFFORTLESS registry only 7% of patients (n=33) had CHD [16]. In pooled analysis (IDE trial and EFFORTLESS registry), the inappropriate shock rate for CHD patients was similar to that for non-CHD patients (10.5% vs 10.9%, P=0.96) and successful defibrillation threshold testing (at 65 J) was comparable in CHD versus non-CHD patients (88.2% vs 94.6%, P=0.26). No spontaneous ventricular arrhythmias were noted on follow-up [55]. Zeb et al. [56] evaluated ECG vector screening in 30 patients with CHD, with 86% having suitable screening vectors. The alternate and primary vectors were more often suitable. In addition, screening of six versus two postures did not significantly affect S-ICD eligibility in CHD patients (83% vs 87%).

            Future Directions/Investigations

            S-ICD and Leadless Pacemakers

            A significant limitation of the S-ICD is its inability to provide backup pacing and ATP therapies. Mondésert et al. [57] reported implantation of a leadless pacemaker in a patient with complete heart block and preexisting S-ICD. No oversensing was seen by the S-ICD during pacing, and the leadless pacemaker functioned normally after an S-ICD shock. Feasibility was also evaluated in a bovine model, with successful implantation of S-ICDs and leadless pacemakers without safety or performance issues in either device after multiple S-ICD and external shocks [58]. The development of a leadless pacemaker that could communicate with the S-ICD and provide both backup pacing and deliver ATP therapies would be a major advance, allowing a fully functional leadless system.

            S-ICD and MRI

            The compatibility of MRI scanning with cardiac implantable devices is an evolving field, with recent approval of MRI-compatible pacemakers and defibrillators. MRI scanning is not currently approved for use with the S-ICD system. However, Keller et al. [59] reported the safety of MRI scanning in 15 patients with S-ICDs undergoing 22 scans (5 brain, 3 cardiac, 12 lumber, and 1 knee scan) at 1.5 T. Device therapies were disabled, and a temperature probe was placed over the S-ICD pocket during scanning. As the S-ICD electrode is extravascular, device heating does not affect the myocardium, but heating of the system/generator may cause discomfort. Four patients noted discomfort due to heating over the S-ICD generator (all during lumbar spine examinations). Two of these patients were rescanned after removal of the external temperature probe, without recurrent symptoms, suggesting heating of the temperature probe may have led to their discomfort. Imaging of the lumbar spine, brain, and knee showed no device artifact. Cardiac imaging showed adequate right ventricular visualization, but S-ICD generator artifact limited visualization of parts of the left ventricle. No delayed enhancement imaging was performed, and no device malfunctions were observed. Early experience with MRI scanning is promising and will need to be confirmed in larger trials.

            Conclusions and Take-Home Message

            The S-ICD system appears safe to implant and is effective at detecting and treating induced and spontaneous VT/VF episodes. It may be considered in patients having an ICD indication who do not have a pacing and/or cardiac resynchronization therapy indication and who are unlikely to benefit from ATP therapy.

            The ideal S-ICD candidate is still to be determined. Young patients and patients with poor venous access, indwelling catheters, channelopathies, and CHD seem the most suitable. However, limited data exist in many of these patient cohorts, and further studies and/or registry data will be needed to determine the long-term suitability of the S-ICD in these populations.

            Initial concerns of inappropriate shocks can be mitigated by appropriate preprocedure screening, dual-zone programming, and sensing optimization during exercise testing. Advances in leadless pacing may make the device more applicable to a larger population.

            Disclosure

            Dr. Panna: No disclosures.

            Dr. Miles: Consultant, Medtronic, Inc; Fellowship funding, Medtronic, St. Jude, Boston Scientific, Biosense-Webster.

            References

            1. GoldbergerJJ, BasuA, BoineauR, BuxtonAE, CainME, CantyJMJr, et al. Risk stratification for sudden cardiac death: a plan for the future. Circulation 2014;129:516–6.

            2. MossAJ, ZarebaW, HallWJ, KleinH, WilberDJ, CannomDS, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002;346:877–83.

            3. BardyGH, LeeKL, MarkDB, PooleJE, PackerDL, BoineauR, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352:225–37.

            4. The Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med 1997;337:1576–83.

            5. BardyGH, SmithWM, HoodMA, CrozierIG, MeltonIC, JordaensL, et al. An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med 2010;363:36–44.

            6. van ReesJB, de BieMK, ThijssenJ, BorleffsCJ, SchalijMJ, van ErvenL. Implantation-related complications of implantable cardioverter-defibrillators and cardiac resynchronization therapy devices: a systematic review of randomized clinical trials. J Am Coll Cardiol 2011;58:995–1000.

            7. EllenbogenKA, WoodMA, ShepardRK, ClemoHF, VaughnT, HollomanK, et al. Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications. J Am Coll Cardiol 2003;41:73–80.

            8. ZeitlerEP, PokorneySD, ZhouK, LewisRK, GreenfieldRA, DaubertJP, et al. Cable externalization and electrical failure of the Riata family of implantable cardioverter-defibrillator leads: a systematic review and meta-analysis. Heart Rhythm 2015;12:1233–40.

            9. Birgersdotter-GreenUM, PretoriusVG. Lead extractions: indications, procedural aspects, and outcomes. Cardiol Clin 2014;32:201–10.

            10. EpsteinAE, DiMarcoJP, EllenbogenKA, EstesNA3rd, FreedmanRA, GettesLS, et al. 2012 ACCF/AHA/HRS focused update incorporated into the ACCF/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol 2013;61:e6–75.

            11. KnopsRE, Olde NordkampLR, de GrootJR, WildeAA. Two-incision technique for implantation of the subcutaneous implantable cardioverter-defibrillator. Heart Rhythm 2013;10:1240–3.

            12. MiglioreF, BottioT, IlicetoS, BertagliaE. Submuscular approach for subcutaneous implantable cardioverter defibrillator: a potential alternative technique. J Cardiovasc Eletrophysiol 2015;26:905.

            13. WillnerJM, MillerMA, SinghA, SharmaD, PalaniswamyC, KukS, et al. Chronic safety and efficacy of submuscular implantation of a subcutaneous ICD. Heart Rhythm 2015;12:(5 Suppl):S336.

            14. Sing-ChienY, BhagwandienRE, Szili-TorokT, TheumsDA. Air entrapment causing early inappropriate shocks in a patient with a subcutaneous cardioverter-defibrillator. Heart Rhythm Case Rep 2015;1:156–8.

            15. WeissR, KnightBP, GoldMR, LeonAR, HerreJM, HoodM, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation 2013;128:944–53.

            16. LambiasePD, BarrC, TheunsDA, KnopsR, NeuzilP, JohansenJB, et al. Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry. Eur Heart J 2014;35:1657–65.

            17. TheunsDA, CrozierIG, BarrCS, HoodMA, CappatoR, KnopsRE, et al. Longevity of the subcutaneous implantable defibrillator: long-term follow-up of the European regulatory trial cohort. Circ Arrhythm Electrophysiol 2015;8:1159–63.

            18. Boston Scientific EMBLEMTD S-ICD pulse generator user’s manual, 359278-002 EN US 2015-02.

            19. AzizS, LeonAR, El-ChamiMF. The subcutaneous defibrillator: a review of the literature. J Am Coll Cardiol 2014;63:1473–9.

            20. GoldMR, WeissR, TheunsDA, SmithW, LeonA, KnightBP, et al. Use of a discrimination algorithm to reduce inappropriate shocks with a subcutaneous implantable cardioverter-defibrillator. Heart Rhythm 2014;11:1352–8.

            21. JarmanJW, LascellesK, WongT, MarkidesV, ClagueJR, TillJ. Clinical experience of entirely subcutaneous implantable cardioverter-defibrillators in children and adults: cause for caution. Eur Heart J 2012;33:1351–9.

            22. BurkeMC, GoldMR, KnightBP, BarrCS, TheunsDA, BoersmaLV, et al. Safety and efficacy of the totally subcutaneous implantable defibrillator: 2-year results from a pooled analysis of the IDE study and EFFORTLESS registry. J Am Coll Cardiol 2015;65:1605–15.

            23. GrohCA, SharmaS, PelchovitzDJ, BhavePD, RhynerJ, VermaN, et al. Use of an electrocardiographic screening tool to determine candidacy for a subcutaneous implantable cardioverter-defibrillator. Heart Rhythm 2014;11:1361–6.

            24. Olde NordkampLR, WarnaarsJL, KooimanKM, de GrootJR, RosenmöllerBR, WildeAA, et al. Which patients are not suitable for a subcutaneous ICD: incidence and predictors of failed QRS-T-wave morphology screening. J Cardiovasc Electrophysiol 2014;25:494–9.

            25. ChanNY, YuenHC, MokNS. Right parasternal electrode configuration converts a failed electrocardiographic screening to a pass for subcutaneous implantable cardioverter-defibrillator implantation. Heart Lung Circ 2015;24:e203–5.

            26. SharmaD, SharmaPS, MillerMA, SinghSM, KalahastyG, EllenbogenKA. Position and sensing vector-related triple counting and inappropriate shocks in the subcutaneous implantable cardioverter-defibrillator system. Heart Rhythm 2015;12:2458–60.

            27. BoersmaL, BurkeMC, NeuzilP, LambiaseP, FriehlingT, TheunsDA, et al. Infection and mortality after implantation of a subcutaneous ICD after transvenous extraction. Heart Rhythm 2016;13:157–64.

            28. GoldMR, TheunsDA, KnightBP, SturdivantJL, SangheraR, EllenbogenKA, et al. Head-to-head comparison of arrhythmia discrimination performance of subcutaneous and tranvenous ICD arrhythmia detection algorithms: the START study. J Cardiovasc Electrophysiol 2012;23:359–66.

            29. BrisbenAJ, BurkeMC, KnightBP, HahnSJ, HerrmannKL, AllavatamV, et al. A new algorithm to reduce inappropriate therapy in the S-ICD system. J Cardiovasc Electrophysiol 2015;26:417–23.

            30. JarmanJW, ToddDM. United Kingdom national experience of entirely subcutaneous implantable cardioverter-defibrillator technology: important lessons to learn. Europace 2013;15:1158–65.

            31. Olde NordkampLR, Dabiri AbkenariL, BoersmaLV, MaassAH, de GrootJR, van OostromAJ, et al. The entirely subcutaneous implantable cardioverter-defibrillator: initial clinical experience in a large Dutch cohort. J Am Coll Cardiol 2012;60:1933–9.

            32. MossAJ, SchugerC, BeckCA, BrownMW, CannomDS, DaubertJP, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med 2012;367:2275–83.

            33. KooimanKM, KnopsRE, Olde NordkampL, WildeAA, de GrootJR. Inappropriate subcutaneous implantable cardioverter-defibrillator shocks due to T-wave oversensing can be prevented: implications for management. Heart Rhythm 2014;11:426–34.

            34. CorzaniA, ZiacchiM, BiffiM, DiembergerI, MartignaniC, BorianiG. Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD. Indian Heart J 2015;67:56–9.

            35. FrommeyerG, ReinkeF, EckardtL, WasmerK. Inappropriate shock in a subcutaneous ICD due to interference with a street lantern. Int J Cardiol 2015;198:6–8.

            36. PorterfieldC, DiMarcoJP, MasonPK. Effectiveness of implantation of a subcutaneous implantable cardioverter-defibrillator in a patient with complete heart block and a pacemaker. Am J Cardiol 2015;115(2):276–8.

            37. AkinI, RögerS, BorggrefeM, KuschykJ. Impeding sudden cardiac death despite subcutaneous implantable defibrillator due to fatal crosstalk. J Cardiovasc Electrophysiol 2015. doi:10.1111/jce.12871.

            38. HongPS, CallinanP, AmitG, HealeyJS. Successful implant of a subcutaneous ICD system in a patient with an ipsilateral epicardial pacemaker. Indian Pacing Electrophysiol J 2015;15:62–4.

            39. GemeinC, HajM, ChasanR, WeipertK, AbaciG, HelmigIM, et al. Combining a S-ICD and a pacemaker with abdominal device location and bipolar epicardial LV-lead: first-in-man approach. Heart Rhythm 2015;12(5 Suppl):S534.

            40. El-ChamiMF, LevyM, KelliHM, CaseyM, HoskinsMH, GoyalA, et al. Outcome of subcutaneous implantable cardioverter defibrillator implantation in patients with end-stage renal disease on dialysis. J Cardiovasc Electrophysiol 2015;26:900–4.

            41. KomanE, GuptaA, SubzposhF, TiwariA, FnataineJ, KusmirekSL, et al. Safety and efficacy of subcutaneous implantable cardioverter defibrillator implantation in patients on hemodialysis. Heart Rhythm 2015;12(5 Supp):S249.

            42. DhamijaRK, TanH, PhilbinE, MathewRO, SidhuMS, WangJ, et al. Subcutaneous implantable cardioverter defibrillator for dialysis patients: a strategy to reduce central vein stenoses and infections. Am J Kidney Dis 2015;66:154–8.

            43. KiamaneshO, O’NeillD, SivakumaranS, KimberS, KimberS. Inappropriate shocks by subcutaneous implantable cardioverter-defibrillator due to T-wave oversensing in hyperkalemia leading to ventricular fibrillation. Heart Rhythm Case Rep 2015;1:257–9.

            44. LambiasePD, HoodM, BoersmaL, TheunsD, BurkeM, WeissR, et al. Combined analysis of subcutaneous ICD outcomes in hypertrophic cardiomyopathy patients from the Effortless registry and IDE study. Heart Rhythm 2015;12 (5 Suppl):S15.

            45. FranciaP, AdduciC, PalanoF, SempriniL, SerdozA, MontesantiD, et al. Eligibility for the subcutaneous implantable cardioverter-defibrillator in patients with hypertrophic cardiomyopathy. J Cardiovasc Electrophysiol 2015;26:893–9.

            46. MauriziN, OlivottoI, Olde NordkampLR, BaldiniK, FumagalliC, BrouwerTF, et al. Prevalence of subcutaneous implantable cardioverter-defibrillator candidacy based on template ECG screening in patients with hypertrophic cardiomyopathy. Heart Rhythm 2016. Doi:10.1016/j.hrthm.2015.09.007.

            47. FrommeyerG, DecheringDG, ZumhagenS, LöherA, KöbeJ, EckardtL, et al. Long-term follow-up of subcutaneous ICD systems in patients with hypertrophic cardiomyopathy: a single-center experience. Clin Res Cardiol 2016;105:89–93.

            48. SaeedD, AlbertA, WestenfeldR, MaxheraB, Gramsch-ZabelH, O’ConnorS, et al. Left ventricular assist device in a patient with a concomitant subcutaneous implantable cardioverter defibrillator. Circ Arrhythm Electrophysiol 2013;6:e32–3.

            49. GuptaA, SubzposhF, HankinsSR, KutalekSP. Subcutaneous implantable cardioverter-defibrillator implantation in a patient with a left ventricular assist device already in place. Tex Heart Inst J 2015;42:140–3.

            50. KuschykJ, StachK, TülümenE, RudicB, LiebeV, SchimpfR, et al. Subcutaneous implantable cardioverter-defibrillator: first single-center experience with other cardiac implantable electronic devices. Heart Rhythm 2015;12:2230–8.

            51. BaderY, WeinstockJ. Successful implantation of a subcutaneous cardiac defibrillator in a patient with a preexisting deep brain stimulator. Heart Rhythm Case Rep 2015;1:241–4.

            52. AlexanderME, CecchinF, WalshEP, TriedmanJK, BevilacquaLM, BerulCI. Implications of implantable cardioverter defibrillator therapy in congenital heart disease and pediatrics. J Cardiovasc Electrophysiol 2004;15(1):72–6.

            53. PettitSJ, McLeanA, ColquhounI, ConnellyD, McLeodK. Clinical experience of subcutaneous and transvenous implantable cardioverter defibrillators in children and teenagers. Pacing Clin Electrophysiol 2013;36(12):1532–8.

            54. ReevesJ, KimJ, MitchellM, McCantaA. Implantation of the subcutaneous implantable cardioverter-defibrillator with retroperitoneal generator placement in a child with hyoplastic left heart syndrome. Heart Rhythm Case Rep 2015;1:176–9.

            55. D’SouzaBA, KimYY, BurkeM, MorganJM, LeonAR, PattonKK, et al. Outcome in patients with congenital heart disease receiving the subcutaneous implantable defibrillator (S-ICD): results from a pooled analysis from IDE study and EFFORTLESS registry. Heart Rhythm 2015;12(5 Suppl):S225.

            56. ZebM, CurzenN, VeldtmanG, YueA, RobertsP, WilsonD, et al. Potential eligibility of congenital heart disease patients for subcutaneous implantable cardioverter-defibrillator based on surface electrocardiogram mapping. Europace 2015;17(7):1059–67.

            57. MondésertB, DubucM, KhairyP, GuerraP, GosselinG, ThibaultB. Combination of a leadless pacemaker and subcutaneous defibrillation: first in-human report. Heart Rhythm Case Rep 2015;1(6):469–71.

            58. TjongF, SmedingL, KooimanK, OldeNorkamp, De GrootJ, WildeA. Combined implantation of a subcutaneous cardioverter-defibrillator (S-ICD) and a leadless pacemaker: evaluation of feasibility, safety, and performance in an ovine model. Heart Rhythm 2015;12(5 Suppl):S64.

            59. KellerJ, NeužilP, VymazalJ, JanotkaM, BradaJ, ŽáčekR, et al. Magnetic resonance imaging in patients with a subcutaneous implantable cardioverter-defibrillator. Europace 2015;17(5):761–6.

            Author and article information

            Journal
            CVIA
            Cardiovascular Innovations and Applications
            CVIA
            Compuscript (Ireland )
            2009-8618
            2009-8618
            February 2016
            February 2016
            : 1
            : 2
            : 199-209
            Affiliations
            [1] 1Division of Cardiology, Department of Medicine, North Florida/South Georgia Veterans Affairs Hospital, Gainesville, FL, USA
            [2] 2Section of Electrophysiology, Division of Cardiology, Department of Medicine, University of Florida, FL, USA
            Author notes
            Correspondence: Mark E. Panna Jr, MD, FACC, FHRS, Division of Cardiology, Department of Medicine, University of Florida,1600 SW Archer Road, PO Box 100277, Gainesville, FL 32610, USA, Tel.: +1-325-2739082, E-mail: mark.panna@ 123456medicine.ufl.edu
            Article
            cvia20150018
            10.15212/CVIA.2015.0018
            2e80181f-5c9b-4910-8764-573761720d58
            Copyright © 2016 Cardiovascular Innovations and Applications

            This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 Unported License (CC BY-NC 4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc/4.0/.

            History
            Categories
            REVIEWS

            General medicine,Medicine,Geriatric medicine,Transplantation,Cardiovascular Medicine,Anesthesiology & Pain management
            sudden cardiac death,subcutaneous implantable cardioverter-defibrillator

            Comments

            Comment on this article