      The Impact of Flip Angle and TR on the Enhancement Ratio of Dynamic Gadobutrol-enhanced MR Imaging: In Vivo VX2 Tumor Model and Computer Simulation.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) is widely used to diagnose cancer and monitor therapy. The maximum enhancement ratio (ERmax) obtained from the curve of signal intensity over time could be a biomarker to distinguish cancer from normal tissue or benign tumors. We evaluated the impact of flip angle (FA) and repetition time (TR) on the ERmax values of dynamic gadobutrol-enhanced MR imaging, obtaining T1-weighted (T1W) MR imaging of VX2 tumors using 2-dimensional fast spoiled gradient echo (2D FSPGR) with various FAs (30°, 60° and 90°) at 1.5 tesla before and after injection of 0.1 mmol/kg gadobutrol. In vivo study indicated significant differences between ERmax values and area under the ER-time curve (AUC100) of VX2 tumors and muscle tissue, with the highest ERmax and AUC100 at FA 90°. Computer simulation also demonstrated the ER as a strictly increasing monotonic function in the closed interval [0°, 90°] for a given TR when using T1W FSPGR, and the highest ER value always occurred at FA 90°. The FA for the highest ER differed from that for the highest signal-to-noise or contrast-to-noise ratio. For long TR, the ER value increases gradually. However, for short TR, the ER value increases rapidly and plateaus so that the ER value changes little beyond a certain FA value. Therefore, we suggest use of a higher FA, near 90°, to obtain a higher ERmax for long TR in 2D SPGR or FSPGR and a smaller FA, much less than 90°, to reach an appropriate ERmax for short TR in 3D SPGR or FSPGR. This information could be helpful in setting the optimal parameters for DCE-MRI.

          Related collections

          Author and article information

          Journal
          Magn Reson Med Sci
          Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine
          Japanese Society for Magnetic Resonance in Medicine
          1880-2206
          1347-3182
          2015
          : 14
          : 3
          Affiliations
          [1 ] Department of Biomedical Engineering, I-SHOU University.
          Article
          10.2463/mrms.2014-0048
          25833269
          10114001-305c-45d1-b6cb-ee1181c4f7b7
          History

          Comments

          Comment on this article