      Noninvasive glucose sensing by transcutaneous Raman spectroscopy.

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          We present the development of a transcutaneous Raman spectroscopy system and analysis algorithm for noninvasive glucose sensing. The instrument and algorithm were tested in a preclinical study in which a dog model was used. To achieve a robust glucose test system, the blood levels were clamped for periods of up to 45 min. Glucose clamping and rise/fall patterns have been achieved by injecting glucose and insulin into the ear veins of the dog. Venous blood samples were drawn every 5 min and a plasma glucose concentration was obtained and used to maintain the clamps, to build the calibration model, and to evaluate the performance of the system. We evaluated the utility of the simultaneously acquired Raman spectra to be used to determine the plasma glucose values during the 8-h experiment. We obtained prediction errors in the range of ~1.5-2  mM. These were in-line with a best-case theoretical estimate considering the limitations of the signal-to-noise ratio estimates. As expected, the transition regions of the clamp study produced larger predictive errors than the stable regions. This is related to the divergence of the interstitial fluid (ISF) and plasma glucose values during those periods. Two key contributors to error beside the ISF/plasma difference were photobleaching and detector drift. The study demonstrated the potential of Raman spectroscopy in noninvasive applications and provides areas where the technology can be improved in future studies.

          Related collections

          Author and article information

          Journal
          J Biomed Opt
          Journal of biomedical optics
          1560-2281
          1083-3668
          May 2015
          : 20
          : 5
          Affiliations
          [1 ] University of Houston, Department of Electrical and Computer Engineering, 4800 Calhoun Road, Houston, Texas 77204, United StatesbUniversity of Houston, Department of Biomedical Engineering, 4800 Calhoun Road, Houston, Texas 77204, United StatescSpectrosco.
          [2 ] Spectroscopy Laboratory, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139, United States.
          [3 ] iSense CGM, 27700SW 95th Avenue, Wilsonville, Oregon 97070, United States.
          Article
          2136574
          10.1117/1.JBO.20.5.051036
          4330710
          25688542
          1080c0de-c9ff-49a0-b2ae-e4baf00c040c
          History

          Comments

          Comment on this article