      The performance of autoclaved high-speed dental handpieces.

      Journal of the American Dental Association (1939)
      Dental High-Speed Equipment, standards, Pressure, Steam, Sterilization, methods

      Read this article at

      ScienceOpenPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          High-speed dental handpieces, which are used so extensively in modern dental practice, may be a source of cross-contamination between patients. A standard autoclave procedure of 121 C for 15 minutes will sterilize the handpieces evaluated in this report. The performance in a standardized laboratory trial showed deterioration of speed in the simulated three-month period. Clinical evaluations for three months showed some minor mechanical problems. These dental handpieces should be used in conjunction with a formal daily and three-month maintenance program.