          PEEK biomaterials in trauma, orthopedic, and spinal implants.

          Since the 1980s, polyaryletherketones (PAEKs) have been increasingly employed as biomaterials for trauma, orthopedic, and spinal implants. We have synthesized the extensive polymer science literature as it relates to structure, mechanical properties, and chemical resistance of PAEK biomaterials. With this foundation, one can more readily appreciate why this family of polymers will be inherently strong, inert, and biocompatible. Due to its relative inertness, PEEK biomaterials are an attractive platform upon which to develop novel bioactive materials, and some steps have already been taken in that direction, with the blending of HA and TCP into sintered PEEK. However, to date, blended HA-PEEK composites have involved a trade-off in mechanical properties in exchange for their increased bioactivity. PEEK has had the greatest clinical impact in the field of spine implant design, and PEEK is now broadly accepted as a radiolucent alternative to metallic biomaterials in the spine community. For mature fields, such as total joint replacements and fracture fixation implants, radiolucency is an attractive but not necessarily critical material feature.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Cervical radiographical alignment: comprehensive assessment techniques and potential importance in cervical myelopathy.

            Narrative review.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Subsidence of stand-alone cervical cages in anterior interbody fusion: warning.

              Anterior cervical decompression and fusion with anterior plating of the cervical spine is a well-accepted treatment for cervical radiculopathy. Recently, to minimise the extent of surgery, anterior interbody fusion with cages has become more common. While there are numerous reports on the primary stabilising effects of the different cervical cages, little is known about the subsidence behaviour of such cages in vivo. We retrospectively reviewed eight patients with cervical radiculopathy operated upon with anterior discectomy and fusion with a stand-alone titanium cervical cage. During surgery, only the cartilage portion of the end plate was removed and the cages were filled with autologous cancellous bone graft from the iliac crest. To assess possible subsidence or migration, three different radiographic measurements in the sagittal plane were taken for each case, postoperatively and at the latest follow-up. Subsidence was defined as any change in at least one of our parameters of at least 3 mm. Follow-up time was 12-18 months (average 15 months). Five of the nine fused levels had radiological signs of cage subsidence. No posterior or anterior migration was observed. However, subsidence did not correlate with clinical symptoms in four of the five patients. The remaining patient with signs of subsidence, whose neck pain and neurologic symptoms had regressed in the early postoperative course, suffered recurrence of radiculopathy 6 months after the surgery. Her symptoms were explained by the subsidence of the cage and the subsequent foraminal stenosis observed on the magnetic resonance imaging (MRI) scan. At 15 months' follow-up, her cage was broken. Our preliminary results, so far limited in number, represent a serious warning to the proponents of stand-alone cervical cages
                Bookmark

                Author and article information

                Contributors
                (View ORCID Profile)
                (View ORCID Profile)
                (View ORCID Profile)
                Journal
                Acta Neurochirurgica
                Acta Neurochir
                Springer Science and Business Media LLC
                0942-0940
                June 2022
                April 26 2022
                June 2022
                : 164
                : 6
                : 1501-1507
                Article
                10.1007/s00701-022-05217-7
                35471708
                2b643294-4e70-444d-b0a6-269c5f578903
                © 2022

                https://www.springer.com/tdm

                https://www.springer.com/tdm

                History

                Comments

                Comment on this article